In the manual tuning mode amplitude-frequency dependence is obtained
continuously and user should choose resonance peak needed by placing it between two
cross-like pointers (see pic. 2.9) and press Zoom In button.
Then the program sets new frequency range and changes modulation amplitude in a way
that maximum amplitude signal in this range being equal to value in the Defl. field.
Note that zero-level of resonance amplitude is set by the program using
first amplitude-frequency dependence scan. So if the tuning has been started
with very narrow frequency range it can lead to errors in zero amplitude estimation. Due to the
wrong zero level wrong modulation amplitudes will be calculated during
tuning.