ACOUSTIC MITIGATION STATUS FOR S3

Robert Schofield, University of Oregon Doug Cook, Akiteru Takamori, Daniel Sigg, LHO Jonathan Kern, Joe Kovalik, LLO Peter Fritschel, Ken Mason, MIT AND MANY OTHERS


PROBLEM
Red shows coherence between H1 and H2 AS_Q over 12h during S2
Coherence
0.3 H2:LSC-AS_Q / H1:LSC-AS_Q H0:PEM-BSC7_MIC / H2:LSC-AS_Q

0.25

0.2

Coherence

0.15

0.1

0.05

0

0

100

200

300

400

500

600

700

800

900

Frequency (Hz)
T0=09/03/2003 02:15:00 Avg=400 BW=0.0117178

Blue shows coherence between microphone near 4k dark port and 2k AS_Q 2


At some frequencies, the sensitivities of all 3 interferometers were limited by acoustic coupling during S2.
Power spectrum
10 10 10
3 H0:PEM-BSC7_MIC(REF0) H0:PEM-HAM1_MIC(REF1) H1:LSC-AS_Q(REF2) H0:PEM-BSC7_MIC(REF3) H0:PEM-HAM1_MIC(REF4) H1:LSC-AS_Q(REF5)

2

Magnitude

1 10 10 10 10
-1

-2

-3

-4

10 *T0=12/03/2003 05:29:46 *Avg=1

2

10

3

Frequency (Hz)
BW=0.187499

Black: AS_Q with noise; Red: normal; Orange: mic with noise; Yellow: normal We would like to reduce acoustic-seismic contribution to noise by 100 to 1000 3


S2 INVESTIGATION OF ACOUSTIC COUPLING
COUPLING SITES confirmed using propagation delays; if a sound shows up on AS_Q before it shows up on a certain mic, there is a coupling site between the sound source and the microphone. Localized injections showed that the coupling was at least 10x worse at the dark port tables than at other coupling sites. MAIN COUPLING MECHANISM was shown to be clipping modulated by acoustic excitation; experiments included manipulation of clipping with irises and experiments to asses backscattering. ACOUSTIC SOURCES: HVAC below 100 Hz, electronics cabinets above 100 Hz, and transients (e.g. building relaxation creeks), about as loud as continuous sources. 7 PROPAGATION FROM SOURCE TO COUPLING SITE; investigations with LLO PSL enclosure showed that acoustic isolation reduced motions on tables, indicating that propagation was mainly through the air. 4


MITIGATION PLANS

(red indicates completed)

I. REDUCE COUPLING (factor of 10 to 20) A. Clipping 1) Eliminate some clipping sites (e.g. EO shutter) 2) Larger optics where needed; lighter mounts for higher resonant frequencies 3) Damp mounts and dumps etc. 4) New periscopes with higher frequency resonances and damping 5) Reduce table resonances around 100 Hz B. Backscattering from table (out of prudence - we haven't seen coupling) 1) Rigid legs or float table II. ACOUSTICALLY ISOLATE WORST COUPLING SITES (factor of 10 to 20) A. Dark port enclosures with internal absorption kits III. REDUCE CONTINUOUS SOURCES (factor of 3 to 5) A. Remove most electronics cabinets from LVEA B. Absorption and damping kits for vacuum electronics cabinets C. Damp single walled sections of ducts D. Insulate pipe feed through from mechanical room E. Insulate PSL chillers

5


OPTICS UPGRADE SINCE S2 TO REDUCE MODULATED CLIPPING
Dark and reflected ports on all three interferometers 1) 2) 3) 4) Removed EO shutters and associated polarizers Replaced several 1" optics with 2" optics Replaced heavy optics mounts with light mounts Simplified beam path Old Mount New Mount EO shutter removed


NEW PERISCOPE DESIGN WITH DAMPING
Accelerometer on periscope top; Blue: no damping Red: damping
Power spectrum
H1:GDS-TEST_34_1_20(REF5)

Damping

10

2
H1:GDS-TEST_34_1_20(REF8)

Magnitude

10

1

100

200

300

400

500

600

700

800

900

Frequency (Hz) *T0=01/08/2003 02:31:49 Avg=1/Bin=8 BW=0.187499

7


AT MOST LOCATIONS WE ARE CURRENTLY USING OLD PERISCOPES WITH DAMPING Steel plate and damping foam attached to back of periscope

Latest version of new periscope design should be compared to this modified periscope 8


TABLE LEG TESTING
Compared 5 leg designs using the following figures of merit: 1) lower rms amplitude of table motion above 70 Hz (to reduce in-band modulation of clipping) 2) lower rms velocity (to reduce potential back-scattering noise - this is less important because we did not find back-scattering noise [next slide], and table motion amplitudes are less than 1/10 laser wavelength).
current tall tripod small tripod (damped) pneumatic minus-k (spring)


No Evidence for Parasitic Interferometry Noise From Backscattering p
10 10 10
4 3

H0:PEM-HAM4_ACCX(REF4) H1:LSC-AS_Q(REF7) H0:PEM-HAM4_ACCX(REF28) H1:LSC-AS_Q(REF31)

2

Magnitude

10 1 10 10 10 10
-1

-2

-3

-4

10

2

Frequency (Hz)
*T0=21/05/2003 08:11:03 Avg=1 BW=0.187499

Red: Accelerometer on Dark Port Table Blue: AS_Q Solid: Normal; Dashed: Shaker shaking table at 14 Hz (sway resonance) 10


COMPARISON OF RIGID AND FLOATING TABLE LEGS ON ISCT3 Red: current leg; Orange: tall tripod; Blue: minus-k; Black: pneumatic
10
-7

Rigid Legs
10
-8

Floating Legs

normal leg 34" tripod leg minus-k leg air leg

Displacement [m/Hz ]

1/2

10

-9

10

-10

10

-11

10

-12

10

-13 2 3 4 5 6 7 8 9 2 3 4 5 6 7 8 9

10

100

Frequency [Hz]

Sum in quadrature of 3 accelerometer axes, converted to displacement

11


TABLE LEG CONCLUSIONS

1) The rigid tripod legs have not proven to be substantially better than our current legs, either for rms velocity or rms amplitude above 70 Hz. However, we have not yet tried bonding the tripod legs to the floor.

2) The two sets of legs that float the table (pneumatic, minus-k spring) were comparable (pneumatic, slightly better) and resulted in a factor of about ten lower rms velocity and nearly a factor of ten lower rms amplitude above 70 Hz.

We are installing the minus-k legs at ISCT3 to study drift of floating tables.

12


DARK PORT ENCLOSURES WITH ABSORPTION KITS

Installed ISCT4 enclosure is giving a factor of about 20 above 100 Hz; red: before; blue: after
10
3

ISCT4_MIC_B4-enclosure H0:PEM-ISCT4_MIC

10

2

Magnitude

10

1

10

2

10

3

Frequency (Hz)


MITIGATION OF CONTINUOUS SOURCES:
Damped visibly vibrating sections of single walled ducts at LHO.

Completed just before S3, so havent had a chance to test completely, but looks like, at most, a factor of 2 reduction in LVEA levels around 100 Hz. 14


Plan to remove 4k electronics cabinets from LVEA, place 2k cabinets in acoustic enclosures, internal absorption for vacuum cabinets and PSL chiller enclosures.

15


REDUCTION IN ACOUSTIC COUPLING SINCE S2
Red: AS_Q normal; Black: AS_Q with noise; Yellow & Orange: BSC7 mic March 12: S2
10 10 10
-12 H0:PEM-BSC7_MIC(REF0) H0:PEM-HAM1_MIC(REF1) H1:LSC-AS_Q(REF2) H0:PEM-BSC7_MIC(REF3) H0:PEM-HAM1_MIC(REF4) H1:LSC-AS_Q(REF5) -13

m/sqrt(Hz)

10 10 10 10 10 10

-14

-15

-16

-17

-18

-19

10

2

10

3

Frequency (Hz)
*T0=12/03/2003 05:29:46 *Avg=1 BW=0.187499

August 9: EO shutter out, periscope mount replaced
10 10
-11 -12

)

Magnitude (m/Hz

1/2

10 10 10 10 10 10 10

-13

H1:LSC-AS_Q(REF35) H1:LSC-AS_Q(REF39) H0:PEM-BSC7_MIC(REF38) H0:PEM-BSC7_MIC(REF42) H0:PEM-HAM1_MIC(REF36) H0:PEM-HAM1_MIC(REF40)

-14

-15

-16

-17

-18

-19

10

2

10

3

Frequency (Hz)
*T0=09/08/2003 04:10:03 Avg=1 BW=0.187499


Sept 13: after enclosure, 2 inch optics; Blue: injection at REFL port (now limiting)

Oct 4: after REFL port work - removal of EO shutter, 2" optics, damped periscope
10 10
-11 -12

)

Magnitude (m/Hz

1/2

10 10 10 10 10 10 10

-13

H1:LSC-AS_Q(REF21) H1:LSC-AS_Q(REF27) H1:LSC-AS_Q(REF29) H0:PEM-BSC7_MIC(REF89) H0:PEM-BSC7_MIC(REF30)

-14

-15

-16

-17

-18

-19

10

2

10

3

Frequency (Hz)
*T0=04/10/2003 07:05:07 Avg=1 BW=0.187499


Red shows coherence between H1 and H2 AS_Q over about 12h Blue shows coherence between microphone near 4k dark port and 2k AS_Q S2:
0.3 H2:LSC-AS_Q / H1:LSC-AS_Q H0:PEM-BSC7_MIC / H2:LSC-AS_Q 0.25

0.2

Coherence

0.15

0.1

0.05

0

0

100

200

300

400

500

600

700

800

900

Frequency (Hz)

S3:

T0 09/03/2003 02 15 00

A

400

BW 0 0117178

0.3 H2:LSC-AS_Q / H1:LSC-AS_Q H0:PEM-BSC7_MIC / H2:LSC-AS_Q H0:PEM-BSC7_MIC / H1:LSC-AS_Q

0.25

Coherence

0.2

0.15

0.1

0.05

0

0

100

200

300

400

500

600

700

800

900

Frequency (Hz)
T0=08/11/2003 06:30:00 Avg=400/Bin=11 BW=0.0117178


PEM UPGRADE IS PUTTING MICROPHONES AND ACCELEROMETERS ON ALL TABLES Microphone on ISCT1, (the worst 4k coupling site) indicates that we have to increase the ambient sound level by a factor of about 100 to produce acoustic peaks reaching 10-17 m/ sqrt(Hz), suggesting that ambient levels will produce peaks reaching 10-19m/sqrt(Hz).

CONCLUSIONS:
1) WE HAVE REDUCED H1 ACOUSTIC COUPLING BY A FACTOR OF OVER 100 SINCE S2 2) AMBIENT SOUND LEVELS ON H1 NOW LIMITING AT ABOUT 10-19 m/sqrt(Hz) AT 200 Hz 3) A LITTLE LESS THAN A FACTOR OF TEN TO GO TO REACH DESIRED INSENSITIVITY 4)FURTHER IMPROVEMENTS WILL NOT BE VISIBLE IN TYPICAL SPECTRA 5) MAKING SURE THESE BENEFITS ARE FULLY EXTENDED TO H2 AND L1

18