To recalibrate NICMOS data, you will need the calnica software (soon to be included in the STSDAS distribution) and the necessary calibration reference files (available from the HST Data Archive using StarView).
The data processing flow chart for normal imaging and spectroscopic images is shown in Figure 13.2.
Figure 13.2: Calibration Steps of the calnica Pipeline
Contemporaneous Observations-calnicb
While previously it has been possible to execute multiple exposures from a single proposal logsheet line (e.g., WFPC2, CR-SPLIT, and NEXP=n constructs), this capability has been significantly expanded to support new requirements of the second generation science instruments. Typical examples include the removal of cosmic rays, the construction of a mosaic image, and the subtraction of the sky background from a sequence of on-target and off-target observations. These observations are distinguished by the fact that their calibration and processing depends upon other observations obtained at the same time.
Figure 13.3: Conceptual calnicb Pipeline
We refer to these sets of exposures as associations. The calnicb task operates on an entire association, and produces one or more products from that set (Figure 13.3). In the case of dither patterns, calnicb reads from the headers of the individual exposure files what the telescope offsets were. It then identifies sources in the images, and, starting with the telescope pointing information from the headers as an initial guess, determines what the pointings actually were. It then combines the images into a final mosaic, rejecting from the output any Cosmic Rays that had not been detected in the individual exposures when they were processed by calnica. The calnicb code uses all the data quality information generated by calnica to avoid propagating identified Cosmic Rays, bad pixels, or saturated pixels, into the output mosaic. In the case of chopped images, if multiple images were obtained at each chop position, calnicb generates a mosaic for each background position and outputs each of those; it then combines each of the background images to generate a background for the target position, and removes this background from the mosaic it has generated for the target position, and then outputs the result of this operation as the final, background subtracted mosaiced image of the target. When calnicb is calculating the offsets between images, it starts with the telescope pointing information as its first guess. If it is unable to match the various images by adjusting this pointing information by more than some limit, the code reverts to using the pointing information alone, on the assumption that there are no sources bright enough to detect in the individual images, or that there are instrumental artifacts which are confusing the offset calculation, or that the telescope suffered a loss of guide star lock during one or more of the exposures resulting in a potentially very large offset being introduced. To date, we have found that all of these things happen very rarely indeed.