Документ взят из кэша поисковой машины. Адрес оригинального документа : http://www.stsci.edu/hst/ghrs/documents/advisories/ghrs_cdbs_badval.html
Дата изменения: Unknown
Дата индексирования: Mon Apr 11 17:57:40 2016
Кодировка:

Поисковые слова: reflection nebula
Unintended Values in Your CDBS Tables
STScI Logo

Hubble Space Telescope
Unintended Values in Your CDBS Tables

GHRS INSTRUMENT REPORT   

"Unintended Values in Your CDBS Tables"  August 21,1991
by: Doug Duncan

Abstract supplied by author:

  Results of SV tests on the accuracy of routine spectral calibration lamp 
  wavelength calibration of GHRS Side 2 data are reported. The Dispersion 
  Constants reference table used for all pipeline calibration to date was 
  found to contain some bad values which can lead to offsets of up to 1 
  Angstrom on the wavelength scale. Using a correct, though still pre-flight, 
  Dispersion Constants reference table, routine accuracy of 0.02 Angstroms. or better
  is produced. For first order gratings, recalibration of data can be expected 
  to lead to wavelength accuracy of around 0.006 Angstroms.