17.5 Calibration Goals

The Cycle 7 calibration goals for NICMOS are listed in Table 17.1 below. Since NICMOS is expected to run out of cryogens between the end of 1998 and the beginning of 1999, these represent the final calibration goals for the instrument. The main limiting factor for attaining such goals will be the thermal stability of the instrument. The Table summarizes the accuracy we expect to achieve for reference files, photometry, astrometry, etc. (see the NICMOS Instrument Handbook).


NICMOS Calibration Goals

Attribute

Accuracy

Comments/Limiting Factors

MULTIACCUM and ACCUM darks

10 ADU

All three detectors. All the supported MULTIACCUM sequences and a subset of ACCUM exposure times will be calibrated to this accuracy. Other ACCUM exposure times will require interpolation.

Flatfields

2% global uncertainty
<1% pixel-to-pixel

NIC1 and NIC2 (reduced program for NIC3). Flatfields will be obtained from a combination of internal lamp flats (1% accuracy, polarizers, broad and medium band filters) and Earth flats (2% acccuracy, narrow band filters). Color dependence may limit flatfields in some cases.

Photometry

5-10% absolute

2% relative over field-of-view
2% stability

Main limiting factors: accuracy of standards; filter leaks on red sources

Focus position

maintained within 1 mm

Instrument focus will be monitored throughout the entire Cycle 7 in all three cameras

Polarizers

1% relative photometry
3-5% absolute polarization

Instrumental polarization and zero position angle of NIC1 and NIC2 will be measured. Limiting factors: dark and flatfield uncertainties; polarizing efficiency.

Grisms

0.01 µm (wavelength calibration)

20%-30% (absolute and relative photometric sensitivity)

Expected accuracy over central 80% of spectral range. Grism C flat may not achieve the declared accuracy.

Coronagraph

0.019" positioning

Hole and detector are not at common focus

NICMOS to FGS astrometry

0.1"

Plate Scale Calibration

0.2%



[Top] [Prev] [Next] [Bottom]