Notice
Hewlett-Packard to Agilent Technologies Transition
This documentation supports a product that previously shipped under the HewlettPackard company brand name. The brand name has now been changed to Agilent Technologies. The two products are functionally identical, only our name has changed. The document still includes references to Hewlett-Packard products, some of which have been transitioned to Agilent Technologies.

Printed in USA

March 2000


Contacting Agilent

By internet, phone, or fax, get assistance with all your test and measurement needs. Table 1-1 Contacting Agilent
Online assistance: www.agilent.com/find/assist United States (tel) 1 800 452 4844 New Zealand (tel) 0 800 738 378 (fax) (+64) 4 495 8950 Latin America (tel) (305) 269 7500 (fax) (305) 269 7599 Japan (tel) (+81) 426 56 7832 (fax) (+81) 426 56 7840 Canada (tel) 1 877 894 4414 (fax) (905) 282-6495 Australia (tel) 1 800 629 485 (fax) (+61) 3 9210 5947 Europe (tel) (+31) 20 547 2323 (fax) (+31) 20 547 2390

Asia Call Center Numbers
Country Singapore Malaysia Philippines Phone Number 1-800-375-8100 1-800-828-848 (632) 8426802 1-800-16510170 (PLDT Subscriber Only) (088) 226-008 (outside Bangkok) (662) 661-3999 (within Bangkok) 800-930-871 0800-047-866 800-810-0189 (preferred) 10800-650-0021 1-600-11-2929 Fax Number (65) 836-0252 1-800-801664 (632) 8426809 1-800-16510288 (PLDT Subscriber Only) (66) 1-661-3714 (852) 2506 9233 (886) 2 25456723 10800-650-0121 000-800-650-1101

Thailand Hong Kong Taiwan People's Republic of China India

2

Chapter 1


User's Guide

HP 83711A/12A and HP 83711B/12B Synthesized CW Generators


HP part number: 83711-90131 Printed in USA July 1997 Supersedes April 1995

Notice.
The information contained in this document is subject to change without notice. Hewlett-Packard makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and tness for a particular purpose. Hewlett-Packard shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.

Copyright Hewlett-Packard Company 1995, 1997 All Rights Reserved. Reproduction, adaptation, or translation without prior written permission is prohibited, except as allowed under the copyright laws. 1400 Fountaingrove Parkway, Santa Rosa, CA 95403-1799, USA

c


The HP 83711A/12A and HP 83711B/12B Synthesized CW Generators
The HP 83711A/12A and HP 83711B/12B synthesized CW generators are referred to as \synthesizers" throughout this manual. The HP 83711A/11B has a carrier frequency range of 1 GHz to 20 GHz and the HP 83712A/12B has a carrier frequency range of 10 MHz to 20 GHz. Speci cation information can be found in Chapter 4, \Speci cations and Options." The HP 83711A/12A and HP 83711B/12B Synthesized CW Generators User's Guide is written to accommodate the novice and the expert user. If you are unfamiliar with the synthesizer, Chapter 2, \Performing Fundamental Synthesizer Operations," is useful. If you are familiar with the synthesizer, Chapter 3, \Generating Signals with the Synthesizer" section is helpful. If you are looking for speci c, detailed information about the synthesizer, refer to the remaining chapters as needed.

Notes
1. This manual applies to instruments with rmware revision 10.0 or greater. 2. If you have an HP 83711A/12A instrument with rmware revision number (5960-7088). 3. To view rmware revision, press 4SPCL5, 415, 4HZ5 (ENTER).
HP 83711A/12A Synthesized Signal Generator User's Guide

<

10.0, refer to the

iii


The Synthesizer at a Glance

The following gure and accompanying text explains some features of the HP 83711A and HP 83711B. The HP 83712A andHP 83712B are nearly identical.

HP 83711B Synthesized CW Generator

iv


5 key allows you to display 1. The 4 any error messages on the front panel display. Error messages are generated when you perform a keystroke sequence that is not valid, try to operate the synthesizer in a mode that is not allowed, etc. 5 key changes the function 2. The 4 of some of the keys. When you press the 4 5 key and then press another key, the synthesizer performs the function printed in blue above the key.

MSG

These keys set the carrier frequency and RF output power level of the synthesizer.
5.

SHIFT

SHIFT

The display shows the current values of synthesizer parameters as well as the status of many of the synthesizer functions. The display shows the current carrier frequency and output power level. The annunciators that appear below the parameters are only visible when their associated function is active.
6.

5 / SAVE key is used to The 4 save most of the synthesizer operating parameters in one of nine nonvolatile register locations so that they can be recalled and used at a later time.
3.

RECALL

The 4 5 and 4 5 keys move the cursor that is over one of the digits in the display either to the right or left when pressed. The digit that is under the cursor will be modi ed when the knob is rotated. If no cursor appears in the display, parameter entry or modi cation has been inhibited.
7.

(

)

The data entry keys are used to enter and modify various synthesizer parameters. 5 key cancels all or The 4 part of an erroneous parameter entry before it has been terminated. The terminator keys (the right-most column of keys) are used to choose the units for the entered parameter as well as to terminate the parameter entry. 5 keys The 4 5, 4 5, and 4 are used to increase or decrease a parameter in predetermined steps.
10.

BACKSPACE

*+

STEP SIZE

The Automatic Level Control voltage input (ALC IN) connector is used as the feedback path to the synthesizer when its RF output power level is being leveled externally.
11. 12. The 4 5 key is used to initiate the activation of several special functions available in the synthesizer. Special functions are additional functions that are not activated by pressing a front panel key or shifted key.

SPCL

The Automatic Level Control keys select the method used to regulate the synthesizer output power level. Either internal leveling, external power meter leveling, or external diode detector leveling can be selected. Additional external equipment is required when either external power meter leveling or external diode detector leveling is used.
4.

The RF OUTPUT connector mates with a Type-N male connector on non-Option 1E9 instruments. The connector mates with a female APC-3.5 mm precision connector on instruments with Option 1E9 installed.
8.

13.

The knob is used to increase or decrease the digit under the cursor ( 9 ) in the display in steps of one.
9.

The POWER switch (LINE on an HP 83711A/12A) turns the synthesizer either on or o.

Synthesizer rear panel features are depicted and described in detail in Chapter 5, \Front/Rear Panel," in this book.

v


In This Book

This book is divided into the following chapters: Chapter 1, \Installing and Verifying the Synthesizer," contains procedures for installing the synthesizer and verifying its operation. Chapter 2, \Performing Fundamental Synthesizer Operations," familiarizes you with the fundamental operation of the synthesizer. Chapter 3, \Generating Signals with the Synthesizer," explains how to generate CW signals and level signals. Chapter 4, \Speci cations and Options," contains a list of the synthesizer performance speci cations as well as the various mechanical, electrical, warranty, and documentation options that are available. Chapter 5, \Front/Rear Panel," contains entries that explain dierent aspects of the synthesizer front and rear panel. (For example, you turn to this chapter for information on the RF connectors). Chapter 6, \Keys/Shifted Functions," contains entries on the function of each key on the synthesizer front panel as well as the shifted or alternate function of certain keys. Chapter 7, \Special Functions," contains entries on the special functions available in the synthesizer. Special functions are hidden during normal instrument operation and can only be invoked by typing a speci c key sequence on the synthesizer front panel. Chapter 8, \Error Messages," contains a table that lists all of the error messages that might be generated during use of the instrument. Each table entry contains a sequence that can be followed to recover from the error condition. Chapter 9, \Legal and Regulatory Information," contains information related to safety and SCPI conformance information. The product warranty is also contained in this chapter.

vi


Contents

1. Installing and Verifying the Synthesizer
Installing the Synthesizer . . . . To Unpack the Synthesizer . . To Install the Synthesizer . . Verifying Synthesizer Functionality If You Encounter a Problem . . Mechanical or Electrical Damage Power-up Problems . . . . . Self Test Failures . . . . . .

.... .... .... ... .... ... .... .... . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

1-3 1-3 1-5 1-7 1-8 1-8 1-9 1-11 2-3 2-4 2-6 2-8 2-9 2-10 2-11 2-12 2-12 2-13 2-13 2-15 3-3 3-4 3-7 3-10 3-10 3-11 3-13 3-13 3-14 3-19 3-20

2. Performing Fundamental Synthesizer Operations
To Enter Data with the Numeric Keypad To Mo dify Data with the Knob . . . . To Mo dify Data with the Arrow Keys . To Save and Recall Synthesizer States . Programming Example . . . . . . . To Read the Contents of the Error Queue Programming Example . . . . . . . To Set the HP-IB Address . . . . . . Programming Example . . . . . . . If You Encounter a Problem . . . . . Data Entry Problems . . . . . . . Programming Problems . . . . . .

3. Generating Signals with the Synthesizer

To Generate a CW Signal . . . . . . . . To Generate Millimeter Signals . . . . . . To Use External Dio de Detector Leveling . . Programming Command . . . . . . . . Related Tasks . . . . . . . . . . . . . To Use External Power Meter Leveling . . . Programming Commands . . . . . . . . Related Tasks . . . . . . . . . . . . . To Use the Level Correct Routine . . . . . Related Tasks . . . . . . . . . . . . . To Use Previously Stored Level Correct Data

Contents-1


Related Tasks . . . . . If You Encounter a Problem Annunciators Turned On RF Output Problems .

. . . .

. . . .

. .. . .

.. . .. ..

. .. . .

. . . .

. . . .

. . . .

. . . .

. . . .

. . . .

.. .. .. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3-21 3-22 3-22 3-23 4-3 4-3 4-4 4-6 4-9 4-9 4-10 4-13 4-17 4-18 4-19 4-20 4-20 4-20 4-20 4-20 4-21 4-21 4-21 4-21 4-21 4-22 4-22 4-22 4-22 4-22 4-23 4-23 4-23 4-23

4. Speci cations and Options

HP 83711A/11B Speci cations . . . . . . . . . . . Frequency . . . . . . . . . . . . . . . . . . . RF Output . . . . . . . . . . . . . . . . . . . Spectral Purity . . . . . . . . . . . . . . . . . HP 83712A/12B Speci cations . . . . . . . . . . . Frequency . . . . . . . . . . . . . . . . . . . RF Output . . . . . . . . . . . . . . . . . . . Spectral Purity . . . . . . . . . . . . . . . . . General . . . . . . . . . . . . . . . . . . . . Front Panel Connectors . . . . . . . . . . . . . Rear Panel Connectors . . . . . . . . . . . . . . Options . . . . . . . . . . . . . . . . . . . . . Electrical Options . . . . . . . . . . . . . . . . Option 1E1 - Add Output Step Attenuator . . . . Option 1E5 - Add High Stability Timebase . . . . Option 1E8 - 1 Hz Frequency Resolution . . . . . Option 1E9 - 3.5 mm RF Output Connector . . . Mechanical Options . . . . . . . . . . . . . . . Option 1CM - Rack Mount Kit . . . . . . . . . Option 1CP - Rack Mount and Handle Kit . . . . Option 1CR - Rack Slide Kit . . . . . . . . . . Warranty Options . . . . . . . . . . . . . . . . Option W30 - Two Additional Years Return-to-HP Service . . . . . . . . . . . . . . . . . . Option W32 - Three Year Return-to-HP Calibration Service . . . . . . . . . . . . . . . . . . Do cumentation Options . . . . . . . . . . . . . Option OB0 - Delete Operating Do cumentation . . Option OB1 - Extra Operating Do cumentation . . Option OBV - Component-Level Information Packet (CLIP) . . . . . . . . . . . . . . . . . . Option OBW - Service Do cumentation . . . . . . Option OBX - Component-Level Information Packet (CLIP)/Service Do cumentation . . . . . . .

Contents-2


5. Front/Rear Panel

Annunciators . . . . . . . See Also . . . . . . . . . Connectors . . . . . . . . Coaxial Connectors . . . HP-IB Connector . . . . Display . . . . . . . . . See Also . . . . . . . . Knob . . . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . POWER Switch . . . . . . See Also . . . . . . . . Power Cables . . . . . . .

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.. . .. .. .. .. .. .. .. .. .. .. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5-3 5-5 5-6 5-7 5-8 5-11 5-12 5-13 5-13 5-14 5-15 5-15 5-16 6-3 6-4 6-4 6-5 6-6 6-6 6-7 6-7 6-8 6-9 6-9 6-10 6-11 6-11 6-11 6-12 6-12 6-12 6-13 6-13 6-14 6-15 6-15 6-15 6-16

6. Keys/Shifted Functions
....... Equivalent SCPI See Also . . . (....... Equivalent SCPI See Also . . . * ....... Equivalent SCPI See Also . . . + ....... Equivalent SCPI See Also . . . ADDRESS . . . Equivalent SCPI See Also . . . BACK SPACE . Equivalent SCPI See Also . . . ENTER (Hz) . . Equivalent SCPI See Also . . . ENTRY OFF . . Equivalent SCPI See Also . . . EXT DIODE . .

)

..... Command ..... ..... Command ..... ..... Commands ..... ..... Commands ..... ..... Command ..... ..... Command ..... ..... Command ..... ..... Command ..... .....

Contents-3


External Dio de Detector Leveling . . . Applications . . . . . . . . . . . . . . Advantages of Dio de Detector Leveling . Disadvantages of Dio de Detector Leveling Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . EXT METER . . . . . . . . . . . . . External Power Meter Leveling . . . . Applications . . . . . . . . . . . . External Equipment Limitations . . . Power Meter . . . . . . . . . . . Power Sensor . . . . . . . . . . . Advantages of Power Meter Leveling . . Disadvantages of Power Meter Leveling . Equivalent SCPI Commands . . . . . See Also . . . . . . . . . . . . . . FREQ . . . . . . . . . . . . . . . . Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . INT LEVEL . . . . . . . . . . . . . Advantages of Internal Leveling . . . . Disadvantages of Internal Leveling . . . Equivalent SCPI Commands . . . . . See Also . . . . . . . . . . . . . . LOCAL . . . . . . . . . . . . . . . Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . LVL CR . . . . . . . . . . . . . . . Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . LVL CR ON/OFF . . . . . . . . . . . Equivalent SCPI Commands . . . . . See Also . . . . . . . . . . . . . . MSG . . . . . . . . . . . . . . . . Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . MULTIPLIER . . . . . . . . . . . . Equivalent SCPI Command . . . . . . See Also . . . . . . . . . . . . . . Numeric Keypad . . . . . . . . . . . See Also . . . . . . . . . . . . . .
Contents-4

.... .... .... ... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... .... ....

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-16 6-17 6-18 6-18 6-19 6-19 6-20 6-20 6-22 6-22 6-22 6-22 6-23 6-23 6-24 6-24 6-25 6-26 6-26 6-27 6-28 6-28 6-29 6-29 6-30 6-30 6-30 6-31 6-34 6-34 6-35 6-36 6-36 6-37 6-37 6-37 6-38 6-39 6-39 6-40 6-40


POWER LEVEL . . . . . Equivalent SCPI Command . See Also . . . . . . . . PRESET . . . . . . . . . Equivalent SCPI Commands See Also . . . . . . . . RCL CR . . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . RECALL . . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . RF ON/OFF . . . . . . . Equivalent SCPI Command See Also . . . . . . . . SAVE . . . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . SAVE CR . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . SHIFT . . . . . . . . . . Equivalent SCPI Command See Also . . . . . . . . SPCL . . . . . . . . . . Equivalent SCPI Command SPCL OFF (kHz) . . . . . Equivalent SCPI Command See Also . . . . . . . . SPCL ON (MHz) . . . . . Equivalent SCPI Command See Also . . . . . . . . STEP SIZE . . . . . . . . Equivalent SCPI Commands See Also . . . . . . . .

.. . .. .. . .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. . ..

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. .. ..

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6-41 6-42 6-42 6-43 6-44 6-44 6-45 6-46 6-46 6-47 6-47 6-47 6-48 6-48 6-48 6-49 6-49 6-49 6-50 6-51 6-51 6-52 6-52 6-52 6-53 6-53 6-54 6-54 6-55 6-56 6-56 6-56 6-57 6-57 6-58

Contents-5


7. Special Functions

ATTEN HOLD ON/OFF . . . Applications . . . . . . . . Advantages . . . . . . . . . Disadvantages . . . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . . CLEAR DISPLAY . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . ERASE MEMORY . . . . . . Application . . . . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . PWR MTR ADDRS . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . PWR MTR SELECT . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . REMOTE LANGUAGE . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . SELF TEST . . . . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . . Service-Related Special Functions See Also . . . . . . . . . . SHOW INFO . . . . . . . . . Equivalent SCPI Command . . See Also . . . . . . . . . .

.. . . . . . . . . . . . . . . . . . . . . . . . .

. . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.. . . . . . . . . . . . . . . . . . . . . . . . .

. . . . .

. . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7-3 7-4 7-4 7-5 7-5 7-5 7-6 7-7 7-7 7-8 7-8 7-8 7-9 7-10 7-11 7-11 7-12 7-12 7-13 7-14 7-14 7-14 7-15 7-16 7-16 7-17 7-17 7-18 7-18 7-18 8-3 8-5

8. Error Messages

Error Messages List . . . . . . . . . . . . . . . . . Messages . . . . . . . . . . . . . . . . . . . . . .

Contents-6


9. Legal and Regulatory Information
Safety Symbols . . . . Miscellaneous Symbols Safety Considerations . Before Applying Power Safety Earth Ground . Cleaning the Product . Certi cation . . . . . Warranty . . . . . . . Limitation of Warranty Exclusive Remedies . Assistance . . . . . . . . . . . . . . . . . . . . . . . .. .. . . . . . . . . . . . . . . . .

.. . . . . .

. . . . . . . . . .

.. . . . . .

. . . . .

. . . . . . . . . .

.

. . . . . . . . . . .

. . . . . . . . . . .

. . . . . . . . . . .

. . . . . . . . . . .

. . . . . . . . . . .

. . . . . . . . . . .

.. . . . . .

. . . . .

. . . . . . . . . .

9-3 9-4 9-5 9-5 9-5 9-6 9-7 9-8 9-8 9-9 9-10

Index

Contents-7


Figures

1-1. 3-1. 3-2. 3-3. 3-4. 4-1. 4-2. 4-3. 4-4. 4-5. 4-6. 4-7. 4-8. 4-9. 4-10. 4-11. 5-1. 5-2. 5-3. 5-4. 5-5. 5-6.

Line Fuse Removal and Replacement . . . . . . . . . . . Millimeter-Wave Equipment Setup . . . . . . . . . . . . . External Diode Detector Leveling Setup . . . . . . . . . . External Power Meter Leveling Setup . . . . . . . . . . . Level Correct Routine Equipment Setup . . . . . . . . . . Typical Maximum Available Output Power from 1 to 20 GHz, at 25 C with Output Step Attenuator (Option 1E1) Installed . . . . . . . . . . . . . . . . . . . . . . Typical Output Level Accuracy and Flatness at +10 and 085 dBm . . . . . . . . . . . . . . . . . . . . . . . . Typical single-sideband phase noise at 1 GHz, 10 GHz and 20 GHz, 25 C, CW mode. Osets less than 100 Hz require the high stability timebase, Option 1E5. . . . . . . . . Typical 2nd Harmonic Levels Measured at Output Power of +6 dBm . . . . . . . . . . . . . . . . . . . . . . . . Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth, CW Mode, with High Stability Timebase (Option 1E5) . . Typical Maximum Available Output Power from 1 to 20 GHz, at 25 C with Output Step Attenuator (Option 1E1) Installed . . . . . . . . . . . . . . . . . . . . . . Typical Maximum Available Output Power from 0.01 to 1 GHz at 25 C. . . . . . . . . . . . . . . . . . . . . . . Typical Output Level Accuracy and Flatness at +10 and 085 dBm . . . . . . . . . . . . . . . . . . . . . . . . Typical single-sideband phase noise at 50 MHz, 1 GHz, 10 GHz and 20 GHz, 25 C, CW mode. Osets less than 100 Hz require the high stability timebase, Option 1E5. . . . . . Typical 2nd Harmonic Levels Measured at Output Power of +6 dBm . . . . . . . . . . . . . . . . . . . . . . . . Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth, CW mode, with High Stability Timebase (Option 1E5) . . Synthesizer Display Annunciators . . . . . . . . . . . . . Synthesizer Connectors - Front Panel . . . . . . . . . . . Synthesizer Connectors - Rear Panel . . . . . . . . . . . . HP-IB Connector and Cable . . . . . . . . . . . . . . . . Synthesizer Display . . . . . . . . . . . . . . . . . . . Power Cable and Line (Mains) Plug Part Numbers . . . . . .

1-10 3-4 3-8 3-11 3-15 4-4 4-5 4-6 4-7 4-8 4-10 4-11 4-12 4-14 4-15 4-16 5-3 5-6 5-6 5-8 5-11 5-17

Contents-8


Contents

6-1. External Diode Detector Leveling Setup . . . . . . . . . . 6-2. External Power Meter Leveling Setup . . . . . . . . . . .

6-17 6-21

Contents-9


T ables

4-1. 4-2. 4-3. 4-4. 5-1. 6-1. 7-1. 9-1.

Osets (HP 83711A only) . . . . Osets (HP 83711B only) . . . . Osets (HP 83712A only) . . . . Osets (HP 83712B only) . . . . HP-IB Interface Cables Available PRESET Conditions . . . . . . Service-Related Special Functions Hewlett-Packard Sales and Service

.... .... .... .... .... .... .... Oces

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

. . . . . . . .

4-6 4-6 4-13 4-13 5-8 6-43 7-17 9-11

Contents-10


1

Installing and Verifying the Synthesizer


Installing and Verifying the Synthesizer

This chapter contains procedures for properly installing your HP 83711A/12A or HP 83711B/12B synthesizer and procedures for functional veri cation of the instrument.

1-2


Installing the Synthesizer

This procedure explains how to inspect, install, and power-up the synthesizer.

T Unpack the Synthesizer o

1. Inspect the shipping container for damage. Look for signs of damage such as a dented or torn shipping container or cushioning material that shows signs of unusual stress or compacting. 2. Carefully remove the contents from the shipping container and inspect

each item for damage.

If the instrument or any accessories appear to be damaged, refer to \Mechanical or Electrical Damage" at the end of this chapter. The following items should have been received in the shipment.

1-3


Installing and Verifying the Synthesizer

Installing the Synthesizer

Item

Model/Part Number

Instrument Power Cable User's Guide Quick Start Guide Programming Guide Calibration Guide Calibration Kit Calibration Guide Calibration Software
Also includes one of the following:

HP 83711A/12A

or

HP 83711B/12B

Depends on shipment destination. Refer to \Power Cables"in Chapter 5 of this manual. 83711-90131 83711-90133 83711-90132 83731-90125 5063-1635 83731-90125 5010-7723 5062-3989 (standard) 5062-3977 (Option 1CM) 5062-3983 (Option 1CP) 1494-0059 (Option 1CR)

Handle Kit Rack Mount Kit Rack Mount Kit with Handles Rack Slide Kit

3. Keep the shipping materials for future use. If undamaged, shipping materials are useful for shipment or storage of the instrument. If damaged, shipping materials should be kept for the carrier's inspection.

1-4


Installing and Verifying the Synthesizer

Installing the Synthesizer

T Install the Synthesizer o

CAUTION

The following provides a general procedure for installation and initial power up of the HP 83711A/12A or HP 83711B/12B synthesizer. 1. Check to make sure that the power cable is undamaged. Do not use the power cable if the plug contacts are bent or broken or if the wire insulation is damaged or if wire is exposed. Never use a power cable if the grounding contact has been removed. Always use the three-prong AC power cord supplied with this product. Failure to ensure adequate earth grounding by not using this cord may cause product damage. 2. Set the POWER switch (LINE on an HP 83711A/12A) on the

synthesizer to standby ( ). 3. Set the line voltage selector switch to match the mains voltage.
The line voltage selector switch is located on the rear panel of the synthesizer to the right of the power module. Use a small at blade screwdriver to set the switch up for mains voltages in the range of 90 V to 132 V; 50, 60, or 400 Hz or down for mains voltages in the range of 198 V to 264 V; 50 or 60 Hz.

1-5


Installing and Verifying the Synthesizer

Installing the Synthesizer

WARNING

This is a Safety Class I product (provided with a protective earthing ground incorporated in the power cord). The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited.
4. Push the module end of the power cable into the power module on the

rear panel of the synthesizer until it is rmly seated. 5. Connect the plug end of the power cable to a suitable mains power receptacle. 6. Set the POWER switch (LINE on an HP 83711A/12A) to on ( ).

When you turn the synthesizer on, the displays and annunciators light momentarily and the ventilation fan starts. In addition, it is normal for the LED annunciators (including the MSG LED) to turn on and then o during a power-up test. (The MSG LED should be o when the power-up test is complete.)

NOTE
Error message 940 TIMEBASE OVEN COLD will remain on after the power-up test if the instrument has the Option 1E5 installed, and has been disconnected from the mains power. This error should go away after a short warm-up period.

Once the power-up test is complete, frequency and power level will be displayed in the right-most display and pulse parameters will be shown in the display if pulse modulation is enabled.

1-6


Verifying Synthesizer Functionality

The veri cation procedure is suitable for incoming inspection; however, you can refer to the HP 83711A/12A and 83711B/12B Synthesized CW Generators Service Guide or the HP 83711A/12A, HP 83711B/12B, HP 83731A/32A, and HP 83731B/32B Synthesized CW Generators Calibration Guide for procedures that test all warranted speci cations. 1. Activate the SELF TEST special function. To activate the SELF TEST special function, perform the following procedure. 5 key. a. Press the 4 b. Press 4 5 on the synthesizer numeric keypad. c. Terminate the special function entry by pressing the 4 5 (ENTER) key. The display will read SELF TEST?, PRESS ENTER. 2. Press the ENTER key again to initiate the synthesizer self test routine. When the self test routine is running, the display will alternately read SELF TESTING and PRESETTING INSTRUMENT. After the test completes, the display momentarily reads SELF TEST PASSED. 3. If the self test indicates an error condition, refer to the section entitled \If You Encounter a Problem" at the end of this chapter. If an error condition exists, the display will momentarily read TEST XX = YY where XX and YY are numbers indicating the error condition.
SPCL 5 Hz

1-7


If You Encounter a Problem

If you have a problem while installing or verifying the synthesizer, check the following list of commonly encountered problems and troubleshooting procedures. If the problem that you encounter is not in the following list, contact the nearest Hewlett-Packard oce for assistance.

NOTE
When transporting the product, use original packaging or comparable only.

Mechanical or Electrical Damage

If the instrument is mechanically or electrically damaged:
Contact the nearest Hewlett-Packard oce.

If the shipping materials are damaged and the instrument is mechanically or electrically damaged:
Contact the carrier as well as the nearest Hewlett-Packard oce. Keep the shipping materials for the carrier's inspection.

1-8


Installing and Verifying the Synthesizer

If You Encounter a Problem

Power-up Problems

If the synthesizer has no power:
Check that the power cord is fully seated in both the mains power receptacle and the synthesizer power module. Check that the synthesizer line fuse is not open.
WA RN I N G

For continued protection against re hazard, replace line fuse only with same type and ratings (type nA/nV). The use of other fuses or materials is prohibited.
Refer to Figure 1-1 to remove the fuse from the power module. You can use a continuity light or an ohmmeter to check the fuse. An ohmmeter should read very close to zero ohms if the fuse is good. The 6.3A, 250 V fuse is HP part number 2110-0703.

1-9


Installing and Verifying the Synthesizer

If You Encounter a Problem

Figure 1-1. Line Fuse Removal and Replacement

Contact the nearest Hewlett-Packard oce for service, if necessary.

If the display lights, but the ventilation fan does not start:
Check that the fan is not stuck. To check the fan, follow these steps: 1. Set the POWER switch (LINE on an HP 83711A/12A) to standby ( ). 2. Check that the fan blades are not jammed. Contact the nearest Hewlett-Packard oce for service, if necessary.

1-10


Installing and Verifying the Synthesizer

If the synthesizer MSG LED annunciator remains on after the power-up test is complete:
NOTE
Error message 940 TIMEBASE OVEN COLD will remain on after the power-up test if the instrument has the Option 1E5 installed, and has been disconnected from the mains power. This error should go away after a short warm-up period.

If the synthesizer MSG annunciator is on, there is a problem with the synthesizer. To determine the error and turn o the MSG annunciator, refer to \To Read the Contents of the Error Queue" in Chapter 2 and the listing of error messages in Chapter 8.

Self T est Failures

If the Self Test Fails:
Check the contents of the error queue for error messages. To determine errors and clear them, refer to \To Read the Contents of the Error Queue" in Chapter 2 and the listing of error messages in Chapter 8.

1-11


Installing and Verifying the Synthesizer


2

Performing Fundamental Synthesizer Operations


Performing Fundamental Synthesizer Operations
This chapter describes fundamental synthesizer operations. The purpose of this chapter is to familiarize you with the fundamental operation of the synthesizer. Procedures cover how to enter or modify data, how to set the HP-IB address, how to save and recall instrument states, etc.

NOTE
This product has been designed and tested in accordance with IEC Publication 1010, Safety Requirements for Electronic Measuring Apparatus, and has been supplied in a safe condition. The instruction documentation contains information and warnings which must be followed by the user to ensure safe operation and to maintain the product in a safe condition.

WARNING

If this product is not used as speci ed, the protection provided by the equipment could be impaired. This product must be used in a normal condition (in which all means for protection are intact) only.

2-2


To Enter Data with the Numeric Keypad

The synthesizer numeric keypad and units terminator keys provide one way to enter function parameters. 1. Select the desired function key or shifted function key. The function must have a numeric value (parameter) associated with 5 and it in order to enter a new value with the numeric keypad. 4 4 5 are examples of functions with parameters. For example, to select power level so that its parameter is active, press the 5 key. 4 Notice that a cursor ( 9 ) appears over one of the digits in the power level display. The cursor will always appear over one of the digits of the active parameter.
FREQ POWER LEVEL POWER LEVEL

NOTE
Two cursors will appear ( 9 9 ) when the cursor is positioned o the display.

2. Enter the desired value of the parameter by pressing the numeric keys

and, if necessary, the negative and decimal keys.

Notice that, as you press the keys, the display shows the numbers that are entered. 3. Press the appropriate units terminator key to enter the value. The units terminator keys appear to the right side of the numeric keypad. The following steps show how to enter a value of 09.5 dBm for power level: 5 key. a. Press the 4 b. Press 40 5 on the numeric keypad. c. Press the dBm (4 5) key to terminate the entry. Once you terminate the entry, the synthesizer updates the power level value to 09.5 dBm.
POWER LEVEL 9.5 GHz

2-3


To Modify Data with the Knob

The knob on the synthesizer front panel is used to modify data. You turn the knob in order to increase or decrease the parameter value of the currently active function. Additionally, you can modify the position of the cursor ( 9 ) that is over the active parameter in order to increase or decrease the rate at which the function parameter changes. 1. Select the function key of the parameter to be modi ed. When the function key is selected, the cursor appears over one of the digits of the selected parameter. 5 key. The For example, to select the frequency function, press the 4 frequency parameter is selected when the cursor appears over one of the digits of the frequency display. 2. Turn the knob clockwise to increase the parameter or
FREQ

counterclockwise to decrease the parameter.

When you turn the knob, the digit under the cursor increases or decreases in steps of one. For example, when frequency is in its preset state and is the active parameter, the frequency display shows the following: 3.000000000 GHz where the cursor appears over the third \0" to the right of the decimal. When you turn the knob in this case, the output frequency changes in 1 MHz steps. 3. If you wish to move the position of the cursor one position to the right,

press the 4)5 key.

Pressing 4)5 once moves the position of the cursor to the right one digit in order to decrease the knob resolution by a factor of ten. 4. If you wish to move the position of the cursor one position to the left,

press the 4(5 key.
4

(5 moves the position of the cursor to the left one digit in order to increase the knob resolution by a factor of ten.

2-4


Performing Fundamental Synthesizer Operations

To Modify Data with the Knob

5. Continue to press the

the desired digit.

4

)5 or 4(5 key until the cursor is positioned over

For example, suppose that frequency is the current active parameter, and you wish to change the cursor position from 1 MHz to 1 GHz. The steps that follow illustrate how to accomplish this: 5 key. a. Press the 4 When the instrument has been set to the preset state, the frequency display indicates the following: 3.000000000 GHz where the cursor appears over the third \0" to the right of the decimal. b. Press the 4(5 key three times. This moves the cursor from the position corresponding to MHz to the position corresponding to GHz. (The cursor is now over the \3".) c. Rotate the knob slowly so that you can see the frequency change in 1 GHz steps.
PRESET

NOTE

The Knob and 4 5, 4 5 keys operate independently of each other. The 4 5, 4 5 keys increase or decrease parameters in de ned steps.

*+

*+

2-5


To Modify Data with the Arrow Keys

The synthesizer 4*5 and 4+5 (arrow) keys increase or decrease the value of the 5 key. The currently active parameter by an amount set with the 4 5 key is referred to as the increment value. amount set with the 4 1. Select the function key of the parameter to be modi ed. When the function key is selected, a cursor ( 9 ) appears over one of the digits of the selected parameter. 5 key. The For example, to select the frequency function, press the 4 frequency parameter is selected when the cursor appears over one of the digits of the frequency display. 2. Press the 4*5 key to increase the parameter by the current increment
STEP SIZE STEP SIZE FREQ

value or press the 4+5 key to decrease the parameter by the current increment value.

For example, when frequency is in its preset state, the frequency display shows the following: 3.000000000 GHz where the cursor appears over the third \0" to the right of the decimal. The preset frequency increment value is 100 MHz. When you press the 4*5 key, the value of frequency increases to 3.100000000 GHz. 3. If you wish to change the increment value, you can do so using the
4STEP
SIZE

5

key.

The 4 5 key allows you to change the increment value of the currently active function. For example, suppose that frequency is the current active parameter, and you wish to change the increment value from its preset value of 100 MHz to 250 MHz. The steps that follow illustrate how to accomplish this: a. Press the 4 5 key. b. Type 4 5 on the synthesizer numeric keypad.
STEP SIZE STEP SIZE 250

2-6


Performing Fundamental Synthesizer Operations

To Modify Data with the Arrow Keys

c. Terminate the frequency increment value entry by pressing the 4 5 key. When either the 4*5 or 4+5 key is pressed, the frequency will be either increased or decreased by 250 MHz instead of 100 MHz.
MHz

NO T E

The knob and 4 5, 4 5 keys operate independently of each other. The knob increases or decreases the digit in the display that is under the cursor.

*+

2-7


To Save and Recall Synthesizer States

When you use the synthesizer for a speci c application, you can save and then recall the instrument state for future use. You can save up to ten dierent instrument states. 1. Press the 4 5 key and then the 4 5 (SA VE) key. When 4 5, 4 5 (SA VE) is pressed, the text SAVE STATE IN REG XXX is shown on the synthesizer display where XXX is the last register number entered. 2. Use the numeric keypad to enter the desired register number. Valid register numbers are 0 through 9. 3. Press the 4 5 (ENTER) key to terminate the entry. 5 key. 4. To recall the instrument state from memory, press the 4 When 4 5 is pressed, the text RECALL STATE FROM REG XX is shown on the synthesizer display where XX is the last register recalled. 5. Use the numeric keypad to enter the desired register number. 6. Press the 4 5 (ENTER) key to terminate the entry.
SHIFT RECALL SHIFT RECALL Hz RECALL RECALL Hz

Notes
1. When an instrument state is saved to an instrument state register, it will write over any instrument state previously stored to that register. 2. If an instrument state has not been previously stored to an instrument state register, the synthesizer will be set to the preset state if you attempt to recall the instrument state from that register.

2-8


Performing Fundamental Synthesizer Operations

To Save and Recall Synthesizer States

Programming Example

Use the following commands to store the instrument state to register 9 and then recall it from register 9:
OUTPUT 719; "*SAV 9" OUTPUT 719; "*RCL 9"

Saves the current instrument state to register #9. Recalls the previously stored instrument state from register #9.

2-9


To Read the Contents of the Error Queue

When one or more error messages are stored in the synthesizer error queue, the front panel MSG LED annunciator will light. Once all error messages have been read and all error conditions have been corrected, the MSG annunciator will turn o. 1. Press the 4 5 key. When the 4 5 key is pressed, the most recent uncleared manual error number and the front panel error message will appear on the display. 2. Look up the manual error number in the table in Chapter 8. Chapter 8 is organized in ascending manual error number order. 3. Perform the instructions following the error message in the list to
MSG MSG

correct or clear the error condition.

After you have completed the procedure in the list, return to this procedure to continue. 4. If the MSG LED annunciator is still lit, perform steps 1 through 3 again

until the MSG annunciator turns o. If the MSG annunciator is turned o, continue with the next step. 5. Press the 4 5 key one more time.
MSG

Pressing the

4MSG5

key again returns the display to normal operation.

NOTE
In the case of unleveled power or unlocked frequency, the MSG LED annunciator may remain lit continuously until the problem is corrected.

2-10


Performing Fundamental Synthesizer Operations

To Read the Contents of the Error Queue

Programming Example

To read the entire contents of the error queue, run the following program:
10 DIM B$[160] 20 OUTPUT 719; "SYST:ERR?" 30 40 ENTER 719; A,B$ PRINT A,B$

50 IF A<>0 THEN 20 60 END

Dimensions array B$ to accept 160 characters. Queries the oldest uncleared error number and message in the HP-IB error queue. Enters the error number into variable A and the HP-IB error message into variable B$. Prints the error number and HP-IB error message to the controller screen. Returns to line 20 if there are more errors in the queue.

2-11


To Set the HP-IB Address

The synthesizer default HP-IB address is preset to 19. You can, however, change the HP-IB address of the synthesizer from the front panel. 1. Press the 4 5 key on the synthesizer front panel. 2. Press the ADDRESS (4 5) key. 5 key is pressed The display indicates HPIB ADDRESS XX when the 4 where XX is the current HP-IB address. 3. Enter the desired HP-IB address using the numeric keypad. For example, if you want to set the synthesizer HP-IB address to 12, type 4 5 on the numeric keypad. The display indicates HPIB ADDRESS 12. 4. Terminate the HP-IB address entry by pressing the 4 5 (ENTER) key. Note that when 4 5 (ENTER) is pressed, the cursor appears over the address parameter in the display. The synthesizer HP-IB address is now set to the new value (12 in the example), but you can still change it with the numeric keypad, knob, or arrow keys at this point. Pressing 4 5 (ENTER) again returns the display to normal operation.
SHIFT LOCAL LOCAL 12 Hz Hz Hz

Programming Example

To set the synthesizer HP-IB address to 12, send the following command:
OUTPUT 719; "SYST:COMM:GPIB:ADDR 12"

Sets the synthesizer HP-IB address to 12.

2-12


If You Encounter a Problem

If you have a problem operating the synthesizer, check the following list of commonly encountered problems and troubleshooting procedures. If the problem that you encounter is not in the following list, contact the nearest Hewlett-Packard oce for assistance.

NOTE
When transporting the product, use original packaging or comparable only.

Data Entry Problems

If the data entry controls (keypad, knob, 4*5, 4+5, 4(5, 4)5 keys) do not respond: Check that the ENTRY OFF function is not enabled. The ENTRY OFF function is not enabled when the cursor ( 9 ) appears
FREQ

over any of the parameters in the display. To return to normal entry mode, press the desired function key which has a numeric parameter associated 5 if you want to enter frequency). with it (for example, press 4 Check that the function key which is selected accepts data. 5 accepts data, but, 4 5 does not. For instance, 4 Check that the synthesizer is in the local (not remote) operating mode. If the synthesizer is in the remote operating mode, the RMT annunciator will be lit. Press the 4 5 key on the front panel to return the synthesizer to local operating mode.
FREQ LOCAL LOCAL

2-13


Performing Fundamental Synthesizer Operations

If You Encounter a Problem

If no cursor ( 9 ) appears over a parameter in the display: Check that the ENTRY OFF function is not enabled. The ENTRY OFF function is enabled when the cursor ( 9 ) doesn't appear
FREQ

over any of the parameters in the display. To return to normal entry mode, press the desired function key which has a numeric parameter associated 5 if you want to enter frequency). with it (for example, press 4

If the parameter you are trying to enter is not accepted by the synthesizer:
Ensure that you are not trying to set the parameter greater than or less than its limit. Refer to the speci cation table in this manual for the parameter limits. Check that the MSG LED annunciator is o. If the synthesizer MSG annunciator is on, there is a problem with the synthesizer. To determine the error and turn o the MSG annunciator, refer to \To Read the Contents of the Error Queue" in this chapter and the listing of error messages in Chapter 8.

If the synthesizer does not display/output the carrier frequency entered:
Check that the frequency multiplier value entered is the expected value. When a multiplier value other than one is entered, the frequency resolution of the signal before multiplication must be obeyed. For example, for a desired frequency of 40 GHz using a multiplier value of three, the synthesizer would have to output a frequency of 13.333333333 GHz. Since frequency resolution at that frequency is 1 kHz, the closest the synthesizer can set the frequency is 13.333333000 GHz. This yields 39.999999000 GHz after multiplication, not 40 GHz.

2-14


Performing Fundamental Synthesizer Operations

Programming Problems

If the synthesizer does not respond to programming commands:
Refer to the procedure, \To Set the HP-IB Address," in this chapter to check and, if necessary, change the HP-IB address.

2-15


Performing Fundamental Synthesizer Operations


3

Generating Signals with the Synthesizer


Generating Signals with the Synthesizer

This section provides procedures for generating signals with the HP 83711A/12A or 83711B/12B. The steps in the procedures assume that you are familiar with the fundamental synthesizer operations. Refer to Chapter 2, \Performing Fundamental Synthesizer Operations," if you are not familiar with these. Procedures in this chapter include how to generate CW signals, how to use external automatic level control, and how to use the level correct routine. The procedures in this chapter are, in general, listed in order of increasing complexity.

3-2


To Generate a CW Signal

You can generate a CW (continuous wave) signal with no modulation characteristics. 1. Press 4 5 to set the synthesizer to the default state. 2. Set the desired frequency. For example, perform the following procedure to set the frequency to 2.000203 GHz. 5 key. a. Press the 4 5 on the synthesizer's numeric keypad. b. Type 4 c. Terminate frequency entry by pressing the 4 5 key. 3. Set the desired RF output power level. For example, perform the following procedure to set the output power level to 2.5 dBm. 5 key. a. Press the 4 b. Type 4 5 on the synthesizer's numeric keypad. c. Terminate the power level entry by pressing the 4 5 (dBm) key.
PRESET FREQ 2.000203 GHz POWER LEVEL 2.5 GHz

3-3


To Generate Millimeter Signals

By using external equipment with the synthesizer, you can generate millimeter-wave signals. Generating millimeter-wave signals uses the following external equipment.
Equipment Requirements

Microwave Ampli er HP 8355X Source Module

Must be compatible with the HP 8355X Source Module used. Must be capable of generating the desired frequency.

CAUTION

In addition, you must supply any cables and adapters necessary to connect the equipment. 1. Connect the equipment as shown in Figure 3-1: Turn o the AC power to the microwave ampli er prior to connecting or disconnecting the source module interface cable.

Figure 3-1. Millimeter-Wave Equipment Setup

2. Enter the proper multiplier value for the HP 8355X-series source

module that you are using into the synthesizer.

Perform the following procedure to select and enter the proper multiplier value. a. Press the 4 5 key. b. Press the 4 5 key.
SHIFT FREQ

3-4


Generating Signals with the Synthesizer

To Generate Millimeter Signals

c. Press the numeric key on the synthesizer numeric keypad that corresponds to the proper multiplier value from the following table:
Source Module Model Number Frequency Band Multiplier Value

HP 83554A HP 83555A HP 83556A HP 83557A HP 83558A

26.5 - 40 GHz 33 - 50 GHz 40 - 60 GHz 50 - 75 GHz 75 - 110 GHz

2 3 3 4 6

d. Terminate the multiplier value entry by pressing the

4Hz5

(ENTER) key.

3. Set the frequency for the desired output signal. The synthesizer frequency display will show the frequency at the output of the millimeter source module, not the synthesizer RF OUTPUT connector. For example, if you want to generate a 30 GHz CW only signal, perform the following steps: 5 key. a. Press the 4 b. Type 4 5 on the numeric keypad. c. Press the 4 5 key to terminate the entry. 5 key to 4. If the RF output is currently turned o, press the 4
FREQ 30 GHz

turn it on.

RF ON/OFF

If the RF output is o, the word OFF appears in the power level portion of the right-most display.
5 key to enable internal leveling. 5. Press the 4 6. Set the approximate desired RF output power at the output of the
INT LEVEL

millimeter source module using the display on the microwave ampli er.
POWER LEVEL

7. Press the

5 and rotate For example, to set the level to 0 dBm, press 4 the synthesizer knob until \0 dBm" is shown on the microwave ampli er display. 4EXT
DIODE

5

key.

3-5


Generating Signals with the Synthesizer

To Generate Millimeter Signals

8. Set the RF output power level desired at the output of the millimeter

source module using the display on the microwave ampli er.

The display on the microwave ampli er shows the power level at the output of the source module to within 62 dB. You should use the display on the microwave ampli er, not the synthesizer, when adjusting the RF output power. For example, to set the output power level to 0 dBm, press 5 key and rotate the knob until \0 dBm" is shown on the the 4 microwave ampli er display.
POWER LEVEL

NOTE

The knob resolution can be changed using the 4 5 and 4 5 keys. However, the multiplied signal frequency resolution is further limited due to the multiplier value used. For example, if the multiplier value is set to 3 and the synthesizer baseband resolution is 1 kHz, the resulting resolution is 3 kHz.

)

(

3-6


To Use External Diode Detector Leveling

External diode detector leveling is useful when you desire leveled RF output power from the synthesizer at a point other than the RF OUTPUT connector. External diode detector leveling uses the following external equipment.
Equipment Requirements

Diode Detector

Must be speci ed for use at the desired synthesizer output frequency. Must produce greater than 1 mv of ALC voltage for the power levels present at the sampling point. Detector can be either positive or negative. Must be speci ed for use at the desired synthesizer output frequency. None. Must be capable of measuring power at the frequency and level present at the sampling point in the leveling loop. You must supply the cables and adapters necessary to connect the equipment.

Power Splitter
or

Directional Coupler Power Meter
(optional)

Power Sensor
(optional)

Cables and Adapters

NOTE
The power meter and power sensor are not required, but are helpful when adjusting the synthesizer output power for the desired power level at the output.

3-7


Generating Signals with the Synthesizer

To Use External Diode Detector Leveling

In addition, you must supply the cables and adapters necessary to connect the equipment. 1. Connect the equipment as shown in Figure 3-2:

Figure 3-2. External Diode Detector Leveling Setup

2. Press the
NOTE

4EXT

DIODE

5

key to enable external leveling.

When the 4EXT DIODE5 key is pressed, the synthesizer enters the external diode detector leveling mode. Power is then held at a constant level at the sampling point, regardless of gain changes in the signal path between the synthesizer RF OUTPUT connector and the output.

3. Set the desired output signal frequency using the

modulation. 4. If the RF output is currently turned o, press the turn it on.

4FREQ5 4RF

key and any
5

ON/OFF

key to

(If the RF output is o, the word OFF appears in the power level portion of the right-most display.) 3-8


To Use External Diode Detector Leveling

Generating Signals with the Synthesizer

5. Adjust the synthesizer output power so that the desired power at the For example, assume that you are using a power splitter and you want 05 dBm at the output of the splitter. In this case, you must set the power at the output to 05 dBm. Perform the following procedure to set the power at the output to 05 dBm. a. Temporarily connect the power meter and sensor at the output. 5 key. b. Press the 4 c. Rotate the knob until the power meter reads 05 dBm (disregard the synthesizer power level reading). d. Disconnect the power meter and sensor from the output and connect the device under test (DUT).
POWER LEVEL

output is attained.

Notes
1. When the 4EXT DIODE5 key is pressed, the synthesizer output power might change. Therefore, you might want to check the output power and readjust if necessary. 2. External diode detector leveling does not provide temperature compensation, thus, output level recalibration might be required in environments that are not temperature stabilized. 3. External diode detector leveling does not provide proper compensation from square law to linear regions of the detector. Therefore, power level changes may require output level recalibration.

3-9


Generating Signals with the Synthesizer

To Use External Diode Detector Leveling

Programming Command

The following command is related to external diode detector leveling.
POW:ALC:SOUR DIOD

Enables external diode detector leveling.

Related T asks



To Use External Power Meter Leveling

3-10


To Use External Power Meter Leveling

External power meter leveling is useful when you desire leveled RF output power from the synthesizer at a point other than the RF OUTPUT connector. External power meter leveling requires the following external equipment.
Equipment Requirements

Power Meter Power Sensor Power Splitter
or

Must have a recorder output and range hold capability. Must be capable of measuring power at the frequency and level present at the sampling point in the leveling loop. Must be speci ed for use at the desired synthesizer output frequency.

Directional Coupler

In addition, you must supply the cables and adapters necessary to connect the equipment. 1. Connect the equipment as shown in Figure 3-3:

Figure 3-3. External Power Meter Leveling Setup

2. Press the 4 5 key to enable internal leveling. 3. Set the desired output signal frequency using the 4
INT LEVEL

FREQ

5

key.
3-11


Generating Signals with the Synthesizer

To Use External Power Meter Leveling

4. Set the desired output power level using the 4 5. If the RF output is currently turned o, press the

POWER LEVEL

5

turn it on.

4RF

ON/OFF

key. 5 key to

If the RF output is o, the word OFF appears in the power level portion of the display. 6. Select auto range on the power meter. 7. Modify the synthesizer output power so that the power meter display

indicates the power desired at the sampling point.

8. Select range hold on the power meter. 5 key on the synthesizer. 9. Press the 4 5 key is pressed, the text RNG-HOLD,POWER When the 4 XXXdBm appears in the display, where XXX is the last range hold meter value that was entered. 10. Enter the power shown on the power meter display into the
EXT METER EXT METER

For example, assume that you are using a directional coupler that has a coupling factor of 22 dB and you want +5 dBm at the output of the coupler. In this case, you must set the power at the sampling point (the coupled output) to 017 dBm. Perform the following procedure to set the power at the sampling point to 017 dBm. 5 key. a. Press the 4 b. Rotate the knob on the synthesizer until the power level displayed on the power meter is 017 dBm. Disregard the power level shown on the synthesizer display.
POWER LEVEL

synthesizer using the numeric keypad.

The value entered into the synthesizer is called the range hold meter value. For example, if the power meter currently reads 017 dBm, perform the following procedure to enter 017 dBm into the synthesizer: a. Type 40 5 on the synthesizer numeric keypad. b. Terminate the range hold meter value entry by pressing the 4 5 (dBm) key.
17 GHz

3-12


To Use External Power Meter Leveling

Generating Signals with the Synthesizer

When the range hold meter value is terminated, the synthesizer enters the external power meter leveling mode. Power is then held at a constant level at the sampling point, regardless of gain changes in the signal path between the synthesizer RF OUTPUT connector and the sampling point.

Programming Commands

The following commands are related to external power meter leveling.
POW:ALC:PMET reading POW:ALC:SOUR PMET

Reads the initial power meter reading into the synthesizer. The initial power meter reading is de ned by the \reading" parameter. Enables external power meter leveling.

Related T asks



To Use External Diode Detector Leveling

3-13


To Use the Level Correct Routine

When activated, the level correct function adjusts the synthesizer output power (to compensate for losses or gains) to provide constant, leveled power at output of the external signal path. When the level correct routine is used, the synthesizer performs a calibration process that corrects for external losses or power variations in an external signal path. When the level correct routine is run, a power meter that is under control of the synthesizer will measure power variations at the output of the external signal path over a user-de ned frequency range. The synthesizer then reads back the power level data from the power meter and creates a table of correction values for each frequency point (the level correction table). The level correction table resides in non-volatile memory and up to four level correction tables can be stored for later use. If the power meter is in a favorable range, the level correct routine takes approximately two minutes to run.

NOTE
When using the level correct routine, the external path to the synthesizer must be linear (must have a 1 dB per 1 dB transfer function).

Using the level correct routine requires the following external equipment.
Equipment Requirements

HP 437B, 438A, 70100A, or SCPI compatible power meter Power Sensor

No substitute. Must be compatible with the power meter used. Must be capable of measuring power at the frequencies and levels present at the output of the signal path being leveled.

3-14


Generating Signals with the Synthesizer

To Use the Level Correct Routine

In addition, you must supply the cables and adapters necessary to connect the equipment. 1. Connect the equipment as shown in Figure 3-4:

Figure 3-4. Level Correct Routine Equipment Setup

2. Set the address that the synthesizer will use to address the power

meter during the level correct routine.

For example, use the following procedure to set the power meter address to 23. 5 key. a. Press the 4 b. Type 4 5 on the synthesizer numeric keypad. c. Press the 4 5 (ENTER) key. You will be prompted to enter the power meter address. d. Type 4 5 on the synthesizer numeric keypad. e. Terminate the power meter address by pressing the 4 5 (ENTER) key.
SPCL 20 Hz 23 Hz

3. Choose the programming language that the synthesizer will use to

communicate with the power meter over the HP-IB.

For example, use the following procedure to choose HP 437B compatible programming language. a. Press the 4 5 key. b. Type 4 5 on the synthesizer numeric keypad. c. Press the 4 5 (ENTER) key.
SPCL 21 Hz

3-15


Generating Signals with the Synthesizer

To Use the Level Correct Routine

d. Press the HP437B.

45

*

or

45

+

key until the display reads POWER METER IS

5 key. 4. Press the 4 5. Adjust the power level to alevel near the center of the power sensor
POWER LEVEL

range. 6. Press the

4SHIFT5

key and then the

4INT

LEVEL

5

(LVL CR) key.

When 4 5, 4 5 (LVL CR) is pressed, the words COR START XXX appear in the left-most display where XXX is the current level correct start frequency.
SHIFT INT LEVEL

7. Enter the desired level correct start frequency. If the start frequency

shown in the display is the desired start frequency, continue with step 8.
For example, use the following procedure to set the level correct start frequency to 3.5 GHz. a. Type 4 5 on the synthesizer numeric keypad. b. Terminate the level correct start frequency by pressing the 4 5 key. Note that the level correct start frequency remains on the display after it is terminated so that it can be modi ed using the knob or 4*5, 4+5 keys if desired.
3.5 GHz

8. Press the

entry.

4Hz5

(ENTER) key to begin level correct stop frequency

When the 4 5 (ENTER) key is pressed, the words COR STOP XXX appear in the left-most display where XXX is the current level correct stop frequency.
Hz

9. Enter the desired level correct stop frequency. If the stop frequency

shown in the display is the desired stop frequency, continue with step 10.
For example, use the following procedure to set the level correct stop frequency to 9.75 GHz. a. Type 4 5 on the synthesizer numeric keypad. b. Terminate the level correct stop frequency by pressing the 4 5 key.
9.75 GHz

3-16


Generating Signals with the Synthesizer

To Use the Level Correct Routine

Note that the level correct stop frequency remains on the display after it is terminated so that it can be modi ed using the knob or 4*5, 4+5 keys if desired. 10. Press the 4 5 (ENTER) key to begin the number of points entry. When the 4 5 (ENTER) key is pressed, the words SET NUMBER OF POINTS XXX appear in the left-most display where XXX is the current number of points. A maximum of 401 points can be entered. 11. Enter the desired number of points. If the number of points shown in
Hz Hz

the display is the desired number of points, continue with step 12.

12. Press the 4 5 (ENTER) key. When the 4 5 (ENTER) key is pressed, the words RUN CORRECTION?, HIT ENTER appear in the left-most display 13. Press 4 5 (ENTER) to run the level correct routine. The level correct routine can be aborted before it begins running by pressing the 4 5 key. It can also be aborted any time while it is 5 key. running by pressing the 4 As the level correct routine is running, the right-most display will be blanked and the left-most display will show each frequency point and measured power level as each measurement occurs. When the level correct routine has nished running, the left-most display will blank and the frequency and output power level will return to the right-most display.
Hz Hz Hz LOCAL LOCAL

Note that the start frequency and stop frequency are included in the number of points. Use the following procedure to set the number of points to 100, for example. a. Type 4 5 on the synthesizer numeric keypad. b. Terminate the number of points entry by pressing the 4 5 (ENTER) key. Note that the number of points entry remains on the display after it is terminated so that it can be modi ed using the knob or 4*5, 4+5 keys if desired.
100 Hz

3-17


Generating Signals with the Synthesizer

To Use the Level Correct Routine

NOTE
As the routine is running, the synthesizer sends the power meter each frequency point so that the correct calibration factor at each frequency can be used. For best accuracy, the correct calibration factors for the power sensor being used should be entered into the power meter prior to running the level correct routine.

14. When the level correct routine has nished running, press the
POWER LEVEL

5 (LVL CR ON/OFF) key to enable the key and then the 4 synthesizer to use the level correct data when determining the output power level. When the level correct data has been enabled, the LVL CR annunciator

4SHIFT5

15. Disconnect the power meter and power sensor from the level correct

will be lit.

setup.

The output of the external signal path can now be connected to the device under test. The data stored in non-volatile memory will only be valid for the current external signal path. If the external signal path is changed, the level correct routine will have to be rerun.

N OT E
The table of level correct values should be stored in one of four level correction register locations for later use.

3-18


Generating Signals with the Synthesizer

To Use the Level Correct Routine

To store the table of level correct values in a level correction register location, perform the following procedure. 5 (SA 5 key and then the 4 1. Press the 4 VE CR) key. The display will show SAVE IN TABLE XXX where XXX is the current level correction register number. 2. Enter the desired level correction register number. For example, use the following procedure to choose level correction register number 2. a. Type 4 5 on the synthesizer numeric keypad. b. Terminate the level correction register number entry by pressing the 4 5 (ENTER) key.
SHIFT INT LEVEL 2 Hz

NOTE
Once the level correction register number entry has been terminated, the current table of level correct values will write over any old level correct data stored at that register location.

Related T asks



To Use Previously Stored Level Correct Data

3-19


To Use Previously Stored Level Correct Data

When the level correct routine is run, the synthesizer creates a table of correction values for each frequency point in the table. Up to four level correction tables can be stored in the synthesizer memory. If the current signal path at the synthesizer output (the external signal path) is identical to the external signal path on which a level correct routine had been previously run, the level correction table can be recalled from memory and be used at a later time. 1. Verify that the current external signal path is correct for the level

correct table you want to recall from memory and use.

The current external signal path must be identical to the external signal path that was calibrated and stored in the level correct table that you wish to recall. If the current external signal path is dierent, the data can cause inaccurate output power levels at the output of the external signal path.
5 (RCL CR) key. 5 key and then the 4 2. Press the 4 The display will show RECALL FRM TABLE XXX where XXX is the last level correction register number recalled. 3. Enter the level correction register number of the register that contains
SHIFT EXT METER

the level correct data you want to use.

4. Press the

For example, use the following procedure to choose level correction register number 2. a. Type 4 5 on the synthesizer numeric keypad. b. Terminate the level correction register number entry by pressing the 4 5 (ENTER) key.
2 Hz SHIFT POWER LEVEL

5 (LVL CR ON/OFF) key 5 key and then the 4 4 to enable the synthesizer to use of level correct data when determining the output power level. When the use the level correct data has been enabled, the LVL CR

annunciator will be lit.

3-20


To Use Previously Stored Level Correct Data

Generating Signals with the Synthesizer

Related T asks



To Use the Level Correct Routine

3-21


If You Encounter a Problem

If you have a problem generating signals with the synthesizer, check the following list of commonly encountered problems and troubleshooting procedures. If the problem involves data entry, check the section in Chapter 2 entitled, \If You Encounter a Problem." If the problem that you encounter is not in the following list or in Chapter 2, contact the nearest Hewlett-Packard oce for assistance.

Annunciators Turned On

If the UNLVL LED annunciator is on:
Check the leveled power speci cation to make sure that you have not exceeded the speci cation. The maximum power speci cation is frequency dependent. Typically less maximum power is available as the synthesizer output frequency is increased. If external diode detector leveling or external power meter leveling is being used, ensure that the leveling loop is not opened. If external diode detector leveling or external power meter leveling is being used, ensure that the synthesizer output is not being forced into an unleveled condition by the leveling circuitry. If the circuit path before the sampling point has a lot of attenuation, the synthesizer will have to source high power even though the required power at the sampling point seems low. For example, if the circuit before the sampling point has 23 dB of attenuation and the power required at the sampling point is 5 dBm, the synthesizer will need to source 28 dBm, which is more that its maximum speci ed power level.

3-22


Generating Signals with the Synthesizer

If the MSG LED annunciator is on: If the synthesizer MSG annunciator is on, there is a problem with the synthesizer. To determine the error and turn o the MSG annunciator, refer

to \To Read the Contents of the Error Queue" in Chapter 2 and the listing of error messages in Chapter 8.

RF Output Problems

If there is no signal at the RF Output connector: Check that the signal at the RF OUTPUT connector is turned on.
If the RF OUTPUT connector is turned o, the text OFF will be displayed in place of dBm in the right-most display. If the signal at the RF OUTPUT connector is turned o, press the 5 key once to turn it on. 4
RF ON/OFF

If the signal at the RF OUTPUT connector does not appear to be phase locked to the external reference:
Check that the external reference is within synthesizer speci cations. Refer to the Speci cation table in Chapter 4 for the external reference input speci cations.

3-23


Generating Signals with the Synthesizer


4

Speci cations and Options


Speci cations and Options

This chapter contains listings of the synthesizer performance speci cations and the mechanical, electrical, warranty, and documentation options that are available. Speci cations for each instrument appear under the instrument number, further on in this chapter. The general speci cations given here apply equally to the HP 83711A/12A and HP 83711B/12B.

4-2


HP 83711A/11B Speci cations

Speci cations describe the instrument's warranted performance over the 0 to 55 temperature range unless otherwise noted. Supplemental characteristics, indicated by italics, are intended to provide information useful in estimating instrument capability in your application by describing typical, but not warranted, performance.

Frequency

Range: 1.0 to 20.0 GHz Resolution: 1 kHz (1 Hz with Option 1E8) Stability (with high stability timebase, Option 1E5): Aging Rate: <1.5210 /day after 24-hour warm up.
-9

Stability (without high stability timebase):

Temperature Eects: <1210-7 over 0 to 55 C, nominally <1.4210-9 / C Line Voltage Eects: <5210-10 for 10% change in line voltage Aging Rate: <1.0210-8 /day after 72-hours at 25 C610 C Temperature Eects: <5210-6 over 0 to 55 C referenced to 25 C

Stability (with external 10 MHz reference):

Same as external reference. Frequency Switching Time: <50 ms to within 1 kHz For <1 GHz steps, not across 10 GHz band switch point: (HP 83711B only)

<35 ms

4-3


Speci cations and Options

HP 83711A/11B Speci cations

RF Output

Maximum Leveled Output Power (0 Cto +35 C):
Frequency

1 20 GHz

0

Standard

with Option 1E1

+11 dBm

+10 dBm

Figure 4-1. Typical Maximum Available Output Power from 1 to 20 GHz, at 25 C with Output Step Attenuator (Option 1E1) Installed

Minimum Leveled Output Power (without Option 1E1):
0 dBm

Minimum Leveled Output Power (with Option 1E1): Display Resolution:
0.01 dB

0110

dBm (090 dBm on HP 83711A)

4-4


HP 83711A/11B Speci cations

Speci cations and Options

Accuracy:

61.0

dB (0 dBm 62.0 dB 62.5 dB

(over all power levels) to speci ed maximum leveled output power) (power 090 dBm) (power < 090 dBm

The use of Type-N RF connectors above 18.0 GHz degrades speci cation typically by 0.2 dB.

Flatness:

60.5 60.7

dB (power dB (power

090 <090

dBm) dBm)

The use of Type-N RF connectors above 18.0 GHz degrades speci cation typically by 0.2 dB. Level Switching Time: <17 ms (without step attenuator range change)

HP 83711B, HP 83712B 01 dBm, 011 dBm, 021 dBm, etc.) Output SWR: <2.0 : 1 nominal

Attenuator range changes occur at:

Figure 4-2. Typical Output Level Accuracy and Flatness at +10 and 085 dBm

4-5


Speci cations and Options

HP 83711A/11B Speci cations

Spectral Purity

SSB Phase Noise (dBc/Hz):
Table 4-1. Osets (HP 83711A only)
Carrier Freq. 100 Hz 1 kHz 10 kHz 100 kHz

2 GHz 10 GHz 18 GHz

0 0 0

66 69 63

0 0 0

74 75 70

0 0 0

91 79 73

0107 0101 099

Table 4-2. Osets (HP 83711B only)
Carrier Freq.

1 GHz to 2 GHz to 5 GHz to

10 GHz to 20 GHz

<2 <5 <10

100 Hz

GHz GHz GHz

0 0 0 0

1 kHz

73 70 69 65

0 0 0 0

10 kHz

83 78 78 73

0 0 0 0

100 kHz

92 83 82 76

0 0 0 0

107 100 100 100

Figure 4-3. Typical single-sideband phase noise at 1 GHz, 10 GHz and 20 GHz, 25 C, CW mode. Osets less than 100 Hz require the high stability timebase, Option 1E5.

4-6


HP 83711A/11B Speci cations

Speci cations and Options

Harmonics: <050 dBc at output levels <+6 dBm, 1 to 20 GHz

Figure 4-4. Typical 2nd Harmonic Levels Measured at Output Power of +6 dBm

Non-Harmonic Spurious (3 kHz): <060 dBc (includes power supply and
frequency synthesis spurious) Non-Harmonic Spurious (<3 kHz): Sub-Harmonics: None

< 050 dBc

4-7


Speci cations and Options

HP 83711A/11B Speci cations

Residual FM: At 1 GHz, <15 Hz in 50 Hz015 kHz bandwidth.

Figure 4-5. Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth, CW Mode, with High Stability Timebase (Option 1E5)

<0150 dBm/Hz

AM Noise Floor: (at 0 dBm and osets greater than 5 MHz from carrier)

4-8


HP 83712A/12B Speci cations

Speci cations describe the instrument's warranted performance over the 0 to 55 temperature range unless otherwise noted. Supplemental characteristics, indicated by italics, are intended to provide information useful in estimating instrument capability in your application by describing typical, but not warranted, performance.

Frequency

Range: 0.01 to 20.0 GHz Resolution: 1 kHz (1 Hz with Option 1E8) Stability (with high stability timebase, Option 1E5): Aging Rate: <1.5210 /day after 24-hour warm up.
-9

Stability (without high stability timebase):

Temperature Eects: <1210-7 over 0 to 55 C, nominally <1.4210-9 / C Line Voltage Eects: <5210-10 for 10% change in line voltage Aging Rate: <1.0210-8 /day after 72-hours at 25 C610 C Temperature Eects: <5210-6 over 0 to 55 C referenced to 25 C

Stability (with external 10 MHz reference):

Same as external reference. Frequency Switching Time: <50 ms to within 1 kHz For <1 GHz steps, not across 10 GHz band switch point: (HP 83712B only)

<35 ms

4-9


Speci cations and Options

HP 83712A/12B Speci cations

RF Output

Maximum Leveled Output Power (0 Cto +35 C):
Frequency

0.01 1 GHz 1 20 GHz

0 0

Standard

with Option 1E1

+13 dBm +11 dBm

+13 dBm +10 dBm

Figure 4-6. Typical Maximum Available Output Power from 1 to 20 GHz, at 25 C with Output Step Attenuator (Option 1E1) Installed

4-10


HP 83712A/12B Speci cations

Speci cations and Options

Figure 4-7. Typical Maximum Available Output Power from 0.01 to 1 GHz at 25 C.

(090 dBm for HP 83712A) Display Resolution: 0.01 dB

Minimum Leveled Output Power (without Option 1E1): 0 dBm Minimum Leveled Output Power (with Option 1E1): 0110 dBm

10 MHz to 50 MHz: 61.3 dB 50 MHz to 20 GHz: 61.0 dB Accuracy: (over all power levels) 10 MHz to 50 MHz: 62.3 dB (power 090 50 MHz to 20 GHz: 62.0 dB (power 090 10 MHz to 20 GHz: 62.5 dB (power <090 The use of Type-N RF connectors above 18.0 typically by 0.2 dB.

Accuracy: (0 dBm to speci ed maximum leveled output power)

dBm) dBm) dBm) GHz degrades speci cation

Flatness:

60.5 60.7

dB (power dB (power

090 <090

dBm) dBm) 4-11


Speci cations and Options

HP 83712A/12B Speci cations

The use of Type-N RF connectors above 18.0 GHz degrades speci cation typically by 0.2 dB. Level Switching Time: <17 ms (without step attenuator range change. Attenuator range changes occur at 01 dBm, 011 dBm, 021 dBm, etc.) Output SWR: <2.0 : 1 nominal

Figure 4-8. Typical Output Level Accuracy and Flatness at +10 and 085 dBm

4-12


HP 83712A/12B Speci cations

Speci cations and Options

Spectral Purity

SSB Phase Noise (dBc/Hz):
Table 4-3. Osets (HP 83712A only)
Carrier Freq. 100 Hz 1 kHz 10 kHz 100 kHz

500 MHz 2 GHz 10 GHz 18 GHz

0 0 0 0

70 66 69 69

0 0 0 0

86 74 75 70

0103 091 079 073

0119 0107 0101 099

Table 4-4. Osets (HP 83712B only)
Carrier Freq.

500 MHz to 1 GHz to 2 GHz to 5 GHz to

10 GHz to 20 GHz

<1 GHz <2 GHz <5 GHz <10 GHz

100 Hz

0 0 0 0 0

1 kHz

78 73 70 69 65

0 0 0 0 0

10 kHz

92 83 78 78 73

0103 092 083 082 076

100 kHz

0 0 0 0 0

115 107 100 100 100

Phase noise decreases 6 dB/octave below 500 MHz and reaches a oor of 0140 dBc/Hz.

4-13


Speci cations and Options

HP 83712A/12B Speci cations

Figure 4-9. Typical single-sideband phase noise at 50 MHz, 1 GHz, 10 GHz and 20 GHz, 25 C, CW mode. Osets less than 100 Hz require the high stability timebase, Option 1E5.

Harmonics: <055 dBc at output levels <+6 dBm, 0.01 to 1 GHz Harmonics: <050 dBc at output levels <+6 dBm, 1 to 20 GHz

4-14


HP 83712A/12B Speci cations

Speci cations and Options

Figure 4-10. Typical 2nd Harmonic Levels Measured at Output Power of +6 dBm

dBc (includes power supply and frequency synthesis spurious) Non-Harmonic Spurious (<3 kHz): < 050 dBc Sub-Harmonics: None

<060

Non-Harmonic Spurious (3 kHz):

4-15


Speci cations and Options

HP 83712A/12B Speci cations

Residual FM: At 1 GHz, <15 Hz in 50 Hz015 kHz bandwidth. Residual FM decreases 6 dB per octave below 1 GHz.

Figure 4-11. Typical Residual FM Measured in 50 Hz - 15 kHz Bandwidth, CW mode, with High Stability Timebase (Option 1E5)

<0150 dBm/Hz, 1020 GHz. <0140 dBm/Hz, 0.0101 GHz.

AM Noise Floor: (at 0 dBm and osets greater than 5 MHz from carrier)

4-16


HP 83712A/12B Speci cations

Speci cations and Options

General

Programming
The HP 83711A/12A and HP 83711B/12B are fully compatible with the Standard Commands for Programmable Instruments (SCPI). SCPI programming complies with IEEE 488.2-1987. In addition, these synthesizers will emulate most applicable HP 8673 commands, providing general compatibility with ATE systems which include HP 8673 Series signal generators. Refer to the HP 8673 compatibility chapter in the Programming Guide for speci c information. Please consult your HP sales representative for details.

50% relative humidity at 40 C. EMC: Complies with CISPR Publication 11/1990, Class A, Group 1 and MIL-STD-461C, Part 2, Methods CE03 (NB full limits, BB 10 dB relaxation 150 kHz to 500 kHz, full limits elsewhere); CS01; CS02; RE02 (Curve 2 +10 dB); RS03 (1 V/m, 15 kHz to 1 GHz). This product is designed for use in INSTALLATION CATEGORY II and POLLUTION DEGREE 2, per IEC 1010 and 664 respectively.

Environmental Operating Temperature Range: 0 to 55 C (for indoor use). Altitude: Up to 15,000 feet (4,572 meters). Relative Humidity: 80% for temperatures up to 31 C decreasing linearly to

Acoustic Noise Emission (Geraeuschemission)

LpA <70 dB(A) per ISO 3744 (LpA <70 dB(A) nach DIN 45635 pt. 1) LpA Operator position: 44.6 dB, based upon type test per ISO 6081. (LpA am Arbeitsplatz: 44.6 dB, typpruefungsergebnis nach DIN 45635 pt. 19) LpA Bystander position: 38.4 dB, based upon type test per ISO 6081. (LpA ktiver Arbeitsplatz: 38.4 dB, typpruefungsergebnis nach DIN 45635 pt. 19)

4-17


Speci cations and Options

HP 83712A/12B Speci cations

Power Requirements Power: 90-132V, 48-440 Hz. 198-264V, 48-66 Hz.
260 VA maximum.

Physical Dimensions Net Weight: < 16 kg (35 lb) Shipping: < 23 kg (49 lb) Size: 498 mm D 2 426 mm W 2 133 mm H (19.6"216.8"25.2")
Transit case available by ordering HP Part Number 9211-2655.

Front Panel Connectors

The standard front panel RF OUTPUT connector is a Type-N precision connector. When Option 1E9 is installed, this front panel connector is a 3.5 mm precision connector. The nominal source impedance is 50
. This front panel BNC connector allows the synthesizer to leveled. It is used with external power meter leveling or leveling. The leveling signal at this input must be in the nominal input impedance is 150 k
. The damage level is

RF OUTPUT

ALC IN

be externally external diode 61V range. The +12 V or 012 V.

4-18


HP 83712A/12B Speci cations

Speci cations and Options

Rear Panel Connectors

10 MHz Input

Accepts a 10 MHz 6 100 Hz, 0 to +10 dBm, external reference signal for operation from an external high stability timebase. Nominal input impedance is 50
.

10 MHz Output

Outputs the 10 MHz reference signal, nominally +3 dBm, for use as an external reference signal. Nominal source impedance is 50
.

0.5V/GHz Output

Supplies a voltage proportional to output frequency for use with mm-wave frequency multipliers, including the HP 83550 Series Millimeter Wave Source Modules.

4-19


Options

There are several electrical, mechanical, warranty, and documentation options available for the CW Generator.

Electrical Options

There are four electrical options available for the CW Generator. These options are as follows:
Option 1E1 - Add Output Step Attenuator

(90 dB for 83711A/83712A) in 10 dB steps (0|110 dB for 11B/12B). The correct amount of attenuation is selected automatically by the CW Generator dependent on the output power level selected. If this option is installed, you can select whether or not the step attenuator will automatically switch. This function is useful during certain applications, such as when external automatic level control is used.
RF OUTPUT connector. The step attenuator has a range of 0 to 110 dB

If Option 1E1 is ordered, an internal step attenuator is included before the

Option 1E5 - Add High Stability Timebase

Option 1E8 - 1 Hz Frequency Resolution

4-20


Speci cations and Options

Options

Option 1E9 - 3.5 mm RF Output Connector

If Option 1E9 is ordered, the RF OUTPUT connector is a male APC-3.5 precision connector in place of the standard female Type-N connector.

Mechanical Options

There are three mechanical options available for the synthesizer. If these options were not ordered with the original shipment, a Handle Kit (part number 5062-3989) is included. If a mechanical option is desired later, they can be ordered from the nearest Hewlett-Packard oce using the part numbers included in each of the following paragraphs. These options are as follows:
Option 1CM - Rack Mount Kit Option 1CP - Rack Mount and Handle Kit

The synthesizer can be mounted to an instrument rack using the rack mount kit. The Rack Mount Kit part number is 5062-3977. The synthesizer can be mounted to an instrument rack using the rack ange kit. In addition, ease of handling is increased when the synthesizer is not rack mounted by using the front panel handles. The Rack Mount and Handle Kit part number is 5062-3983. This kit is useful when the synthesizer is rack mounted. Access to internal circuits and components or the rear panel is possible without removing the synthesizer from the rack. The Rack Slide Kit part number is 1494-0059. If a non-HP rack enclosure is used, rack adapters can be ordered to allow the slide kit to be used with the non-HP rack. The part number for metric rack adapters is 1494-0023 and the part number for standard inch rack adapters is 1494-0061.

Option 1CR - Rack Slide Kit

4-21


Speci cations and Options

Options

Warranty Options

There are two warranty options available for the synthesizer. These options are as follows:
Option W30 - Two Additional Years Return-to-HP Service Option W32 - Three Year Return-to-HP Calibration Service

This option extends the bene ts of factory warranty to provide a total of 3 years of customer return repair service. This option provides 3 years of HP calibration service at HP customer service centers.

Documentation Options

Option OB0 - Delete Operating Documentation

There are two documentation options available for the synthesizer. If the documentation was not ordered with the original shipment and is now desired, it can be ordered from the nearest Hewlett-Packard oce using the part numbers included in each of the following paragraphs. These options are as follows: If Option OB0 is ordered, the shipment deletes copies of the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators User's Guide, the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Programming Guide, the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Quick Start Guide, and the HP 83711A/12A, HP 83711B/12B, HP 83731A/32A, and HP 83731B/32B Synthesized CW Generators Calibration Guide.

4-22


Speci cations and Options

Option OB1 - Extra Operating Documentation

If Option OB1 is ordered, the shipment includes an extra copy of the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators User's Guide, the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Programming Guide, HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Quick Start Guide, and the HP 83711A/12A, HP 83711B/12B, HP 83731A/32A, and HP 83731B/32B Synthesized CW Generators Calibration Guide. If Option OBV is ordered, the shipment includes a copy of the HP 83711A/12A, HP 83711B12B, HP 83731A/32A, and HP 83731B/32B Synthesized CW Generators Component-Level Information Packet. If Option OBW is ordered, the shipment includes a copy of the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Service Guide. If Option OBX is ordered, the shipment includes a copy of the HP 83711A/12A, HP 83711B/12B, HP 83731A/32A, and HP 83731B/32B Synthesized CW Generators Component-Level Information Packet and a copy of the HP 83711A/12A and 83711B/12B Synthesized CW Generators Service Guide.

Option OBV Component-Level Information Packet (CLIP) Option OBW - Service Documentation

Option OBX Component-Level Information Packet (CLIP)/Service Documentation

4-23


Speci cations and Options


5

Front/Rear Panel


Front/Rear Panel

This chapter contains detailed information on various aspects of the synthesizer front and rear panel. Information on the synthesizer display, front panel connectors, power cable, etc., can be found in this chapter.

5-2


Annunciators

The HP 83711A/12A and 83711B/12B front panel display contains annunciators that show the status of several of the synthesizer functions and settings.

Figure 5-1. Synthesizer Display Annunciators

All annunciators except the UNLOCK, UNLVL, and MSG annunciators are contained on the uorescent display. A description of each of the annunciators follows. INT (ALC) This annunciator indicates that internal power leveling is selected. DIODE (ALC) This annunciator indicates that external diode power leveling is selected. MTR (ALC) This annunciator indicates that external power meter leveling is selected. MULT This annunciator indicates that the frequency multiplier function is active. EXT REF This annunciator indicates that an external time base is currently being used.

5-3


Front/Rear Panel

Annunciators

SPCL LVL CR RMT LSN TLK SRQ UNLOCK UNLVL MSG

This annunciator indicates that one or more special functions are enabled. This annunciator indicates that the synthesizer output is currently being leveled using stored level correction data. This annunciator indicates that the instrument is in the HP-IB remote state, that is, the instrument is under control of an external HP-IB controller. This annunciator indicates that the instrument is addressed to listen (accept data or commands). This annunciator indicates that the instrument is addressed to talk (output data). This annunciator indicates that the instrument is generating a service request to the external controller. This LED annunciator indicates that one or more of the synthesizer frequency control circuits is not phase locked. This LED annunciator indicates that the synthesizer output power is unleveled. When the UNLVL annunciator is lit, the power level shown in the display might be incorrect. This LED annunciator indicates that uncleared error messages are in the synthesizer front panel error queue.

5-4


Front/Rear Panel

Annunciators

See Also

Connectors Display
4EXT 4EXT 4INT
DIODE METER LEVEL

5 5

5

4MSG5

MULTIPLIER
4SPCL5

5-5


Connectors

Figure 5-2. Synthesizer Connectors - Front Panel

Figure 5-3. Synthesizer Connectors - Rear Panel

5-6


Front/Rear Panel

Connectors

Coaxial Connectors

0.5 V/GHz OUT This rear panel BNC connector can be used as one of the inputs to a recorder. It produces a DC voltage output that varies linearly with the frequency currently at the RF OUTPUT connector. For example, if the current frequency setting is 5.5 GHz, the voltage at this connector would be 2.75 V. The nominal source impedance is 250
. 10 MHz IN This rear panel BNC connector accepts a 10 MHz, 0 to +10 dBm reference signal for operation referenced to an external time base. The nominal input impedance of this input is 50
. This connector detects when a valid reference signal is connected to it and automatically switches from internal to external reference operation. 10 MHz OUT This rear panel BNC connector provides a 3 dBm 63 dB, 10 MHz signal derived from the internal frequency standard of the synthesizer. The nominal source impedance is 50
. ALC IN This front panel BNC connector allows the synthesizer to be externally leveled. It is used with external power meter leveling or external diode leveling. The leveling signal at this input must be in the 61V range. The nominal input impedance is 150 k
. The damage level is +12 V or 012 V. RF OUTPUT The standard front panel RF OUTPUT connector is a Type-N precision connector. When Option 1E9 is installed, this front panel connector is a 3.5 mm precision connector. The nominal source impedance is 50
. When making connections to this connector, carefully align the center conductor elements, then rotate the knurled barrel while mating components remain still. Tighten the barrel until rm contact is made. Take care when working with this connector. If the connector is mechanically degraded in any way, high frequency losses can occur. Refer to application note 326, Coaxial Systems - Principles of Microwave Connector Care (HP part number 5954-1566) for more information. 5-7


Front/Rear Panel

Connectors

HP-IB Connector

This connector allows the synthesizer to be connected to other instruments or devices on the interface bus. Details of this cable are shown in Figure 5-4. HP part numbers for various HP-IB cables that are available are shown in the table following the gure.

Figure 5-4. HP-IB Connector and Cable Table 5-1. HP-IB Interface Cables Available
HP-IB Cable Part Number Length

HP 10833A HP 10833B HP 10833C HP 10833D

1 m (3.3 ft) 2 m (6.6 ft) 4 m (13.2 ft) 0.5 m (1.6 ft)

5-8


Front/Rear Panel

Connectors

As many as 14 HP-IB instruments can be connected to the synthesizer (15 total instruments in the system). The cables can be interconnected in a star pattern (one central instrument with the HP-IB cables emanating from that instrument like spokes on a wheel), or in a linear pattern (like boxcars on a train), or a combination of the two. There are certain restrictions that must be followed when interconnecting instruments. These restrictions are as follows: Each instrument must have a unique HP-IB address, ranging from 0 to 30 (decimal). Refer to \To Set the HP-IB Address" in Chapter 2 for information on setting the synthesizer HP-IB address. In a two-instrument system that uses just one HP-IB cable, the cable length must not exceed 4 meters (13.2 ft). When more than two instruments are connected on the bus, the cable length to each instrument must not exceed two meters (6.6 ft). The total cable length between all instruments must not exceed 20 meters (65 ft). Hewlett-Packard manufactures HP-IB extender instruments (Models HP 37201A and HP 37204A/B) that overcome the range limitations imposed by the cabling rules. These extenders allow twin pair cable operation up to 1 km (3,280 ft), and telephone modem operation over any distance. HP sales and service oces can provide additional information on the HP-IB extenders.

5-9


Front/Rear Panel

Connectors

The codes in the following list describe the HP-IB interface capabilities of the synthesizer using IEEE Std.488.1 compatibility codes (HP-IB, GP-IB, IEEE-488, and IEC-625 are all electrically equivalent). Brie y, the mnemonics translate as follows: SH1 Source Handshake, complete capability. AH1 Acceptor Handshake, complete capability. T5 Talker; capable of basic talker, serial poll, and unaddress if MLA. TE0 Talker, Extended address; no capability. L3 Listener; capable of basic listener and unaddress if MTA. LE0 Listener, Extended address; no capability. SR1 Service Request, complete capability. RL1 Remote Local, complete capability. PP0 Parallel Poll, no capability. DC1 Device Clear, complete capability. DT0 Device Trigger, no capability. C1 Controller capability, system controller. E2 Electrical speci cation indicating tri-state outputs. These codes are described completely in the IEEE Standard 488 (1978), IEEE Standard Digital Interface for Programmable Instrumentation or the identical ANSI Standard MC1.1.

5-10


Display

The HP 83711A/12A and 83711B/12B front panel display contains an area for displaying the current operating parameters of the synthesizer. Front panel annunciators show the status of several of the synthesizer functions and settings.

Figure 5-5. Synthesizer Display

A description of the display follows: Annunciators The front panel annunciators show the status of several of the synthesizer functions and settings. An annunciator that is in uorescent display is not visible if its associated function is not active or selected. For example, if diode detector leveling is currently selected, the DIODE annunciator will be lit, otherwise, it will not be visible. The uorescent display annunciators are INT (ALC), DIODE (ALC), MTR (ALC), MULT, EXT REF, SPCL, LVL CR, RMT, LSN, TLK, and SRQ. The three front panel LED annunciators (UNLOCK, UNLVL, and MSG) will be lit to warn you that an associated condition exists. (For example, if error messages are in the error queue, the MSG annunciator will be lit.) An

5-11


Front/Rear Panel

Display

Display

explanation of each annunciator can be found in the \Annunciators" entry in this section of the manual. The carrier display is the uorescent display that is on the right side of the front panel. The primary purpose of this display is to indicate the current setting of the synthesizer output frequency and power level.When parameter entry is initiated, the information in this display will be temporarily replaced with a parameter entry display. Pressing 4 5 (ENTER) returns the frequency and power level information 5 key is pressed, to the display. As an example, if the 4 the text RECALL STATE FROM REG 0 appears in the display. If a cursor ( 9 ) appears over one of the digits in the display, this digit will be increased or decreased as the knob is rotated. Note that the cursor indicates that particular parameter is active. The cursor can be moved left or right by pressing the 4(5 or 4)5 keys.
Hz RECALL

NOTE
When operating the synthesizer in a secure environment, the display can be blanked so that the synthesizer parameters and status cannot be viewed. For more information, refer to the \Clear Display" entry in Chapter 7 of this manual.

See Also
4 4

Annunciators CLEAR DISPLAY DISPlay[:WINDow][:STATe] Knob 5-12

(5 )5


Knob

The knob is used to increase or decrease parameter values. A cursor ( 9 ) over a digit in the display indicates that digit will be increased or decreased in steps of one as the knob is rotated. Clockwise rotation of the knob increases the display digit under the cursor and counterclockwise rotation decreases it. Pressing the 4(5 key moves the display cursor to the left and pressing the 4)5 key moves the cursor to the right. As the knob is rotated, the display will carry over to the adjacent digit. For example, if the display reads 9.000000000 GHz with the cursor over the \9" digit, clockwise rotation of the knob will cause the displayed frequency to change to 10.000000000 GHz. The cursor will remain over the rst digit before the decimal point (the \0" in \10"). If 4 5 (ENTRY OFF), 4)5) has been enabled, the cursor will disappear from the display and rotating the knob will have no aect on the synthesizer parameters.
SHIFT

Equivalent SCPI Command

There is no equivalent SCPI command for knob rotation, however, the \SYST:KEY 61" command simulates clockwise knob rotation and the \SYST:KEY 62" command simulates counterclockwise knob rotation.

5-13


Front/Rear Panel

Knob

See Also
4 4

ENTRY OFF SYSTem:KEY

(5 )5

5-14


POWER Switch

The POWER switch (LINE on HP 83711A/B) turns power to the synthesizer to either on ( ) or standby ( ). The POWER switch (LINE on HP 83711A/B) is located at the bottom left corner of the front panel and is a rocker-type switch. Pressing the symbol turns the synthesizer on and pressing the symbol turns it to standby. When set to standby, most of the synthesizer circuitry is powered o, however, power to the internal timebase is not disconnected.
WA RN IN G

Before turning the synthesizer on, make sure that it is grounded through the protective conductor of the power cable to a mains power receptacle provided with protective earth contact. Any interruption of the protective grounding conductor inside or outside of the synthesizer or disconnection of the protective earth terminal can result in personal injury.

CAUTION

Before turning the synthesizer on, set the line voltage selector to the voltage of the power source. Failure to do this can cause instrument damage when the power cable is plugged in. For information on setting the line voltage selector, refer to \To Install the Synthesizer" in Chapter 1 of this manual.

See Also

To Install the Synthesizer Power Cables

5-15


Power Cables

The line power cable is supplied in one of several con gurations, depending on the destination of the original shipment. Each instrument is equipped with a three-wire power cable. When connected to an appropriate AC power receptacle, this cable grounds the instrument chassis. The type of power cable shipped with each instrument depends on the country of destination. See Figure 5-6, \Power Cable and Line (Mains) Plug Part Numbers," for the part numbers of these power cables. Cables are available in dierent lengths and some with right-angle plugs to the instrument. Check with your nearest Hewlett-Packard service center for descriptions and part numbers for these cables.

5-16


Front/Rear Panel

Figure 5-6. Power Cable and Line (Mains) Plug Part Numbers

5-17


Front/Rear Panel


6

Keys/Shifted Functions


Keys/Shifted Functions

This chapter contains detailed information on the various keys and shifted functions found on the synthesizer front panel. Key names are printed on the keycap and represent the primary function of the key. The shifted function names appear above certain keys on the synthesizer front panel. Shifted functions are always accessed by pressing the 4 5 key and then the key below the shifted function name. If no text appears above a given key, that key has no shifted function associated with it.
SHIFT

6-2


)

The 4)5 key moves the cursor ( 9 ) that appears over one of the digits in the display to the right one digit each time it is pressed. The digit that is currently under the cursor will be increased or decreased in steps of one as the knob is rotated. Clockwise rotation of the knob increases the digit under the cursor and counterclockwise rotation decreases it. If pressing the 4)5key moves the cursor too far to the right, pressing the 4(5 key will move the cursor back to the left.

NOTE

The digit under the cursor is aected only by the knob, not the 4 5 and 4 5 keys.

*

+

If 4 5 (ENTRY OFF), 4)5) has been enabled, the cursor will disappear from the display and rotating the knob will have no aect on the parameter. Note that the cursor indicates that particular parameter is active. For example, if the cursor appears over any digit in the frequency display, the frequency parameter is the active parameter. If an arrow key is pressed that moves the cursor to a position that is o the display, another cursor will appear next to the original cursor to indicate that the actual cursor position is o the display. For example, if a frequency multiplier of six is entered into the synthesizer, the maximum frequency entry that can be made is 120 GHz (20 GHz26). The display, however, only has room to display two digits to the left of the decimal point. If you were to attempt to position the cursor three digits to the left of the decimal point, the double cursor would appear, indicating that the actual cursor position is o the display.
SHIFT

6-3


Keys/Shifted Functions

)

Equivalent SCPI Command

There is no equivalent SCPI command for the 4)5 key, however, sending the \SYST:KEY 45" command is eectively the same as pressing the 4)5 key.

See Also
4

ENTRY OFF Knob SYSTem:KEY

(5

6-4


(

The 4(5 key moves the cursor ( 9 ) that appears over one of the digits in the display to the left one digit each time it is pressed. The digit that is currently under the cursor will be increased or decreased in steps of one as the knob is rotated. Clockwise rotation of the knob increases the digit under the cursor and counterclockwise rotation decreases it. If pressing the 4(5 key moves the cursor too far to the left, pressing the 4)5 key will move the cursor back to the right.

NOTE

The digit under the cursor is aected only by the knob, not the 4 5 and 4 5 keys.

*

+

If 4 5, 4)5 (ENTRY OFF) has been enabled, the cursor will disappear from the display and rotating the knob will have no aect on the parameter. Note that the cursor indicates that particular parameter is active. For example, if the cursor appears over any digit in the frequency display, the frequency parameter is the active parameter. If an arrow key is pressed that moves the cursor to a position that is o the display, another cursor will appear next to the original cursor to indicate that the actual cursor position is o the display. For example, if a frequency multiplier of six is entered into the synthesizer, the maximum frequency entry that can be made is 120 GHz (20 GHz26). The display, however, only has room to display two digits to the left of the decimal point. If you were to attempt to position the cursor three digits to the left of the decimal point, the double cursor would appear, indicating that the actual cursor position is o the display.
SHIFT

6-5


Keys/Shifted Functions

(

Equivalent SCPI Command

There is no equivalent SCPI command for the 4(5 key, however, sending the \SYST:KEY 37" command is eectively the same as pressing the 4(5 key.

See Also
4

ENTRY OFF Knob SYSTem:KEY

)5

6-6


*

The 4*5 key allows you to incrementally add to the current value of a parameter. Pressing the 4*5 key will increase the currently active parameter by the 5 key. A parameter can be made increment value set with the 4 active by pressing its associated function key. For example, pressing the 4 5 key will make frequency the active parameter, pressing the 5 key will make output power level the active parameter, etc. If 4 5, 4)5 (ENTRY OFF) function no parameter is currently active, or the 4 has been enabled, pressing the 4*5 key will have no eect. If the 4*5 key is pressed and held down, the key will auto-repeat, that is, the active parameter will automatically increase rapidly in increment value steps.
STEP SIZE FREQ POWER LEVEL SHIFT

Equivalent SCPI Commands

There is no equivalent SCPI command for the 4*5 key, however, the various SCPI commands that send numeric parameter data include an \UP" parameter option. Sending the \UP" parameter in place of the numeric parameter is eectively the same as pressing the 4*5 key. For example, if you want to increase the output frequency by its current increment value, send the following command: OUTPUT 719; "FREQ UP"

6-7


Keys/Shifted Functions

*

See Also
45

ENTRY OFF
4STEP
SIZE

+

5

6-8


+

The 4+5 key allows you to incrementally subtract from the current value of a parameter. Pressing the 4+5 key will decrease the currently active parameter by the 5 key. A parameter can be made increment value set with the 4 active by pressing its associated function key. For example, pressing the 4 5 key will make frequency the active parameter, pressing the 5 key will make output power level the active parameter, etc. If 4 5, 4)5 (ENTRY OFF) function no parameter is currently active, or the 4 has been enabled, pressing the 4+5 key will have no eect. If the 4+5 key is pressed and held down, the key will auto-repeat, that is, the active parameter will automatically decrease rapidly in increment value steps.
STEP SIZE FREQ POWER LEVEL SHIFT

Equivalent SCPI Commands

There is no equivalent SCPI command for the 4+5 key, however, the various SCPI commands that send numeric parameter data include a \DOWN" parameter option. Sending the \DOWN" parameter in place of the numeric parameter is eectively the same as pressing the 4+5 key. For example, if you want to decrease the output frequency by its current increment value, send the following command: OUTPUT 719; "FREQ DOWN"

6-9


Keys/Shifted Functions

+

See Also
45

ENTRY OFF
4STEP
SIZE

*

5

6-10


ADDRESS

4LOCAL5

5 key and then the Invoking the ADDRESS function (pressing the 4 key) displays and allows you to change the synthesizer HP-IB address.
SHIFT

5 is pressed, the current HP-IB address will be displayed When 4 5, 4 across the display in the following format: HPIB ADDRESS XX where XX is the current HP-IB address. The range for valid addresses is 00 to 5 30. The HP-IB address is preset at the factory to 19. Pressing the 4 key has no aect on the HP-IB address. The preset up/down arrow increment value is 1.
SHIFT LOCAL PRESET

Equivalent SCPI Command

SYST:COMM:GPIB:ADDR address sets the synthesizer HP-IB address as

de ned by the \address" parameter.

See Also

SYSTem:COMMunicate:GPIB:ADDRess

6-11


BACK SPACE

The 4 entry. The (4

BACK SPACE

5

key allows you to cancel part or all of a parameter during

key has an eect on the display only after a function key 5, etc.) is pressed and before the entry is terminated. If 5 is pressed repeatedly so that the whole parameter is canceled, 4 the display reverts back to what it was before the function key was pressed.
4BACK
FREQ

5, 4POWER

SPACE

5

LEVEL

BACK SPACE

Equivalent SCPI Command
5 key, however, There is no equivalent SCPI command for the 4 sending the \SYST:KEY 54" command is eectively the same as pressing the 5 key. 4
BACK SPACE BACK SPACE

See Also

SYSTem:KEY

6-12


ENTER (Hz)

The ENTER key (actually the secondary function of the 4 5 key) is used to terminate unitless parameter entries. It can also be used to return the display to its default state.
Hz

The 4 5 (ENTER) key is used to terminate unitless parameter entries. The unitless parameters are multiplier values, special function numbers, instrument state register numbers, level correction register numbers, and the synthesizer and power meter HP-IB addresses. The ENTER key can also be used to return the display to its default state. When the display is in the default state, the display shows frequency and power level. In the display default state, no parameters are active. To use the ENTER key to return the display to the default state, press the ENTER key without rst pressing a numeric key (that is, the keys numbered 0 through 9). For example, if a multiplier value of 5 is entered, the display will continue to show FREQUENCY MULTIPLIER 5 after the entry is terminated with the cursor ( 9 ) over the 5. Pressing ENTER again will return the display to the default state mentioned above.
Hz

Equivalent SCPI Command

There is no equivalent SCPI command for the ENTER key, however, sending the \SYST:KEY 58" command is eectively the same as pressing the ENTER key.

6-13


Keys/Shifted Functions

ENTER (Hz)

See Also

ADDRESS MULTIPLIER PWR MTR ADDRS RCL CR
4RECALL5

SAVE SAVE CR
4SPCL5

SYSTem:KEY

6-14


ENTRY OFF

5 key and then the Invoking the ENTRY OFF function (pressing the 4 4)5 key) disables the parameter entry/modi cation keys. The parameter entry/modi cation keys include the numeric keypad as well as the 4*5, 4+5, 4 5, 4(5, and 4)5 keys and the knob.
SHIFT STEP SIZE

Once the ENTRY OFF function has been enabled, it will be cancelled as soon 5 etc.) is pressed. 5, 4 as one of the function keys (4
FREQ POWER LEVEL

Equivalent SCPI Command

There is no equivalent SCPI command for the ENTRY OFF function, however, sending the \SYST:KEY 0;KEY 45" command is eectively the same as pressing 4 5, 4)5.
SHIFT

See Also

SYSTem:KEY

6-15


EXT DIODE

The 4 5 key enables external diode leveling. External diode leveling is a method of automatic level control (ALC) of the output signal using an external diode detector.
EXT DIODE

5 key is pressed, the DIODE annunciator will be lit. If When the 4 the procedure, \To Use External Diode Detector Leveling" in Chapter 3 has been followed correctly, the synthesizer RF output will be externally leveled via the diode detector feedback loop.
EXT DIODE

External Diode Detector Leveling

The purpose of a leveling circuit is to provide constant power, independent of the load, and minimize power variations versus frequency. External diode detector leveling is used in applications where it is desired to level the power at some point outside the synthesizer with an external diode detector. When frequency dependent losses are involved, the RF output power at the end of the signal path will not have a constant amplitude over the synthesizer frequency range. For example, if a cable is used at the output of the synthesizer that has a constant 0.5 dB/GHz loss, 5 dB of attenuation at the output of the cable occurs after a 10 GHz frequency increase, even though the power at the input to the cable is constant. By externally leveling power at the output of the cable, the synthesizer would increase power at the input of the cable to produce a constant power level at the output of the cable. External diode detector leveling requires that external equipment be connected to the synthesizer, as shown in Figure 6-1.

6-16


Keys/Shifted Functions

EXT DIODE

Figure 6-1. External Diode Detector Leveling Setup

When external diode detector leveling is chosen, power is sampled at the external sampling device (either a directional coupler or power splitter) by an external diode detector that is typically operating in the square law region. When the diode detector is operating in the square law region, it will provide a DC voltage that is proportional to the power sampled at the input to the detector. This DC voltage is fed back to the synthesizer via the ALC IN connector. The synthesizer then adjusts its output power level to maintain a constant power level at the input to the external diode detector.

Applications

External leveling can be used when your application requires long cables that will cause frequency-dependent losses. It also enables devices, such as ampli ers, mixers, etc. to be inserted into the RF signal path so that the output of the inserted device is controlled by the synthesizer.

6-17


Keys/Shifted Functions

EXT DIODE

Advantages of Diode Detector Leveling

When diode detector leveling is used, power level correction is continuous. External diode detector leveling has the advantage of faster settling time than power meter leveling. The settling time is variable and is dependent on the devices in the external leveling loop.

Disadvantages of Diode Detector Leveling

The diode detector must be capable of producing between 1 mV and 1 V of ALC voltage for the power level at the sampling point. This typically restricts the lower limit at which external diode detector leveling will function. Diode detector leveling might not provide an accurate power display on the synthesizer if the diode detector is operating outside of the square law region. Diode detector leveling also does not provide temperature compensation. Power level recalibration might be required in environments that are not temperature stabilized. External diode detector leveling requires that external equipment be connected to the synthesizer.

NOTE
Before selecting 4EXT DIODE5, you should adjust the synthesizer power level using internal leveling so that the step attenuator is set to the correct setting.

6-18


Keys/Shifted Functions

EXT DIODE

Equivalent SCPI Command

POW:ALC:SOUR DIOD enables external diode detector leveling.

See Also
4EXT 4INT
METER LEVEL

5

[SOURce[1]:]POWer:ALC:SOURce To Use External Diode Detector Leveling To Use the Level Correct Routine

5

6-19


EXT METER

The 4 5 key enables external power meter leveling. External power meter leveling is a method of automatic level control (ALC) of the output signal using an external power meter and power sensor.
EXT METER

5 key is pressed, you are prompted to put the power When the 4 meter in the range hold mode and then enter the power meter reading into the synthesizer. After the range hold power meter value is entered, the MTR annunciator will be lit. If the procedure, \To Use External Power Meter Leveling" in Chapter 3 has been followed correctly, the RF output will be externally leveled via the power meter feedback loop.
EXT METER

NOTE
The range hold power meter reading is the power level that is used by the synthesizer to calibrate the recorder output voltage of the power meter versus the synthesizer output power.

External Power Meter Leveling

The purpose of a leveling circuit is to provide constant power, independent of the load, and minimize power variations versus frequency. External power meter leveling is used in applications where it is desired to level the power at some point outside the synthesizer with an external power meter. When frequency dependent losses are involved, the RF output power at the end of the signal path will not have a constant amplitude over the synthesizer frequency range. For example, if a cable is used at the output of the synthesizer that has a constant 0.5 dB/GHz loss, 5 dB of attenuation at 6-20


Keys/Shifted Functions

EXT METER

the output of the cable occurs after a 10 GHz frequency increase, even though the power at the input to the cable is constant. By externally leveling power at the output of the cable, the synthesizer would increase power at the input of the cable to produce a constant power level at the output of the cable. External power meter leveling requires that external equipment be connected to the synthesizer, as shown in Figure 6-2.

Figure 6-2. External Power Meter Leveling Setup

When external power meter leveling is chosen, power is sampled at the external sampling device (either a directional coupler or power splitter) by the external power sensor. An automatic level control voltage is then generated by the external power meter and fed back to the synthesizer via the ALC IN connector. The synthesizer then adjusts its output power level to maintain a constant power level at the input of the external power sensor.

6-21


Keys/Shifted Functions

EXT METER

Applications

External leveling can be used when your application requires long cables that will cause frequency-dependent losses. It also enables devices, such as ampli ers, mixers, etc. to be inserted into the RF signal path so that the output of the inserted device is controlled by the synthesizer.

External Equipment Limitations

Power Meter

The power meter must be capable of producing a 0 V to 1 V output voltage linearly proportional to power over each decade of range. The recorder output of most power meters provides this voltage. The power sensor must have a frequency range that is appropriate for the range of frequencies being leveled. The sensor must also have enough dynamic range to measure the level at the output of the directional coupler or power splitter. As an example, to level signals in the 07 dBm to 0 dBm range using a 10 dB coupler, the power sensor must be capable of measuring power in the 017 to 010 dBm range.

Power Sensor

6-22


Keys/Shifted Functions

EXT METER

Advantages of Power Meter Leveling

When power meter leveling is used, power level correction is continuous. External power meter leveling has the advantages of better accuracy and temperature stability, and improved vernier linearity over external diode detector leveling. Using a sensitive power sensor allows ALC at levels as low as the power meter and sensor can measure.

Disadvantages of Power Meter Leveling

One disadvantage of power meter leveling is a longer settling time than diode detector or internal leveling. The settling time is dependent on the power range and sensor used. The synthesizer assumes a settling time of two seconds to allow the sensor to reach the correct power level. Depending on the power sensor and range being used, the power meter might or might not have nished settling. Power is typically settled within two seconds for the two highest ranges of the power meter. External power meter leveling requires that external equipment be connected to the synthesizer.

NOTE
Before selecting 4EXT METER5, you should adjust the power level using internal leveling so that the step attenuator is set to the correct setting.

6-23


Keys/Shifted Functions

EXT METER

Equivalent SCPI Commands

NOTE
The following is only the command that selects external power meter leveling. External power meter leveling, however, is a multi-step process that involves issuing several commands.

POW:ALC:PMET pmeter sets the power meter range hold value as POW:ALC:SOUR PMET

de ned by the \pmeter" parameter. sets the alc source to power meter.

See Also
4EXT 4INT
DIODE LEVEL

5

[SOURce[1]:]POWer:ALC:PMETer [SOURce[1]:]POWer:ALC:SOURce To Use External Power Meter Leveling

5

6-24


FREQ

The 4 5 (frequency) key allows you to set the output frequency of the synthesizer.
FREQ

Frequency is displayed in the left portion of the display. The preset value is 3 GHz. The valid output frequency range diers depending on the synthesizer model: HP 83711A/11B - 1.0 GHz to 20.0 GHz HP 83712A/12B - 0.01 GHz to 20.0 GHz If a frequency entry is made that is outside the allowable range, an error message will be generated and the actual frequency will be set to either its upper or lower limit (whichever is closest to the input frequency). Standard frequency resolution is 1 kHz over the range of 0.01 GHz to 20 GHz.

NOTE
When Option 1E8 is installed, frequency resolution is 1 Hz over the entire frequency range.

The preset up/down arrow increment value is 100 MHz. The increment value will be rounded to the nearest 1 kHz.

6-25


Keys/Shifted Functions

FREQ

Notes
1. When Option 1E8 is installed in either the HP 83711A/12A or HP 83711B/12B, increment values with 1 Hz resolution are accepted over the entire frequency range. 2. If the Multiplier function is being used, the frequency displayed is the frequency at the output of the multiplier, not the output of the synthesizer.

Equivalent SCPI Command

FREQ freq sets the synthesizer output frequency as de ned by the

\freq" parameter.

See Also

MULTIPLIER [SOURce[1]:]FREQuency[:CWj:FIXed]

6-26


INT LEVEL

The 4 5 key enables internal leveling. Internal leveling uses an internal ALC (automatic level control) detector to provide automatic level control of the output power at the RF OUTPUT connector.
INT LEVEL

5 key is pressed, the INT (ALC) annunciator will be lit. When the 4 Internal leveling is used to control the internal RF signal over a speci ed range (the vernier range) of 01 to +13 dBm. Additional dynamic range is provided by an optional 0110 dB step attenuator (Option 1E1) to give an eective dynamic range of 0110 to +8 dBm (090 to +8 dBm on HP 83711A/12A). An ALC unleveled condition occurs when the internal ALC circuitry cannot maintain leveling. This can occur due to an instrument fault or because the instrument is set to level for an RF output level that is beyond its capability. Calibrated output level is only guaranteed when the UNLVL annunciator is not lit. When the synthesizer UNLVL annunciator lights, the knob, arrow keys, or the numeric keypad can still be used to change displayed power up to the maximum value. However, the actual output power will not increase. Only the displayed value changes. The internal ALC circuit maintains a constant RF power level over frequency at the RF OUTPUT connector. The ALC circuit is a feedback control system where output power is measured and compared to the desired power level. When output power does not equal the desired power level, the ALC changes the output until the actual and desired levels are equal. The actual maximum leveled power available is dependent upon the frequency and varies across the range of the synthesizer. 5 key) is The allowable range for power level entries (using the 4 015 dBm to +30 dBm for standard con guration instruments and 0120 dBm (0100 dBm on HP 83711A/12A) to +30 dBm if Option 1E1 is installed.
INT LEVEL POWER LEVEL

6-27


Keys/Shifted Functions

INT LEVEL

NOTE
The actual maximum internally leveled output power for your instrument at a given frequency can be found by increasing the synthesizer output power until the UNLVL annunciator lights.

When the synthesizer is set to the preset state, internal leveling is selected over external diode leveling or external power meter leveling.

Advantages of Internal Leveling

Internal leveling is self-contained; it does not require any external equipment as does external diode leveling or external power meter leveling. Leveled power is speci ed at the RF OUTPUT connector.

Disadvantages of Internal Leveling

Internal leveling does not compensate for losses or gains in the output signal path.

6-28


Keys/Shifted Functions

INT LEVEL

Equivalent SCPI Commands

POW:ALC:SOUR INT enables internal leveling.

See Also
4EXT 4EXT
DIODE

5 5 5

METER

4POWER

[SOURce[1]:]POWer:ALC \To Use the Level Correct Routine" in Chapter 3

LEVEL

6-29


LOCAL

The

4LOCAL5

key removes the synthesizer from the remote state.

When the synthesizer is no longer in the remote state, the RMT annunciator in the display will no longer be lit. 5 key If the instrument is in the local lockout (LLO) state, pressing the 4 will not remove the instrument from the remote state. In this case, the only way to return the synthesizer to local operation is either by setting the REN bus control line false or sending the instrument the go-to-local (GTL) bus command.
LOCAL

Equivalent SCPI Command
5 key, however, sending There is no equivalent SCPI command for the 4 5 the \SYST:KEY 8" command is eectively the same as pressing the 4 key.
LOCAL LOCAL

See Also

Connectors SYSTem:KEY

6-30


LVL CR

The LVL CR (level correct) function allows you to calibrate external path losses in the signal path. The level correct function creates a table in synthesizer memory of external path loss values versus frequency. 5 key and Invoking the LVL CR (level correct) function (pressing the 4 then the 4 5 key) causes the synthesizer to prompt you for the start frequency, stop frequency, and number of points in the level correction table. Once the start frequency, stop frequency, and number of points are entered, you are given the option of running the level correct routine. 5 (LVL CR) is pressed, the synthesizer initially When 4 5, 4 prompts you for the level correct start frequency. When 4 5, 4 5 (LVL CR) is pressed, the text COR START XXXGHZ will be displayed where XXX is the current start frequency. At this point, a new start frequency can be entered or the existing start frequency can be modi ed using the knob or arrow keys. The start frequency is used to determine the beginning frequency for the level correction table. The start frequency is the rst frequency point in the table.
SHIFT INT LEVEL SHIFT INT LEVEL SHIFT INT LEVEL

N OT E
You do not have to choose a new start frequency (or stop frequency or number of points). To keep the same parameter value, press 4Hz5 (ENTER) without entering a new parameter and the next parameter in the sequence will be displayed.

6-31


Keys/Shifted Functions

LVL CR

Once you have chosen the proper start frequency, pressing any terminator key while a numeric parameter entry is not in progress will cause the synthesizer to go to stop frequency entry mode. When in stop frequency entry mode, the synthesizer will display COR STOP XXXGHZ. Stop frequency entry is identical to start frequency entry. The stop frequency is used to determine the ending frequency for the level correction table. The stop frequency is the last frequency point in the table. Once you have chosen the proper stop frequency, pressing any terminator key while a numeric parameter entry is not in progress will cause the synthesizer to go to number of points entry mode. When in number of points entry mode, the synthesizer will display SET NUMBER OF POINTS XXX. The number of points value determines how many frequency points will be in the level correct table. The number of points includes the start and stop frequency points. The synthesizer uses the start and stop frequency values and number of points and places evenly spaced frequency points in the level correction table. Note that if the calculated frequency points are not within the synthesizer frequency resolution, they will be rounded. Once you have chosen the proper number of points, pressing any terminator key while a numeric parameter entry is not in progress will cause the synthesizer to ask if you want to run the level correction routine. When the synthesizer is asking if you want to run the level correction routine, the display indicates RUN CORRECTION?, HIT ENTER. If you press the 4 5 (ENTER)) key, the synthesizer begins collecting new level correction data. For each frequency point in the table, the synthesizer sends the power meter the frequency, waits for the meter to settle, and then reads the power value from the power meter. As the correction runs, the synthesizer calculates the loss data by subtracting the power meter reading from the current power level. (The synthesizer power level is constant during the level correct routine.)
Hz

6-32


Keys/Shifted Functions

LVL CR

Notes
1. If the level correction routine is aborted (by pressing the 4LOCAL5 key), the existing level correction data will be unaected. 2. After level stop that run. the level correction routine has successfully completed, save the level correction data to the correction table chosen by the \SAVE CR" function. If the routine fails, the entered start and frequencies, and number of points will remain at what they were set at. Be aware, however, the data in the level correct table will still re ect the last successful level correction that was

Once the synthesizer has completed the level correct routine and the new level correction table has been generated, it is resident in the synthesizer 5 will erase the table. The table should be volatile memory. Pressing 4 stored in one of four level correction table registers. The start frequency lower limit is equal to the synthesizer lower frequency limit. The start frequency upper limit is equal to the stop frequency minus the minimum frequency resolution at that frequency. For example, the preset value for start frequency increment value is 100 MHz . The stop frequency lower limit is equal to the start frequency plus the minimum frequency resolution at that frequency. The stop frequency upper limit is 20 GHz. The preset value for stop frequency increment value is 100 MHz . The allowable range for number of points is 2 to 401. The preset increment value is 1.
PRESET

6-33


Keys/Shifted Functions

LVL CR

Equivalent SCPI Command

There is no equivalent SCPI command for the LVL CR function, however, the commands under the \Level Correction Commands" tab in Chapter 1 of the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Programming Guide explain how to load data into level correct tables, select tables, turn level correction on, etc.

See Also

LVL CR ON/OFF PWR MTR ADDRS RCL CR SAVE CR \To Use the Level Correct Routine" in Chapter 3

6-34


LVL CR ON/OFF

The LVL CR ON/OFF (level correct on/o) function (4 toggles the level correct function on and o.

SHIFT

5, 4POWER

LEVEL

5

)

When the level correct function is on, the LVL COR annunciator is lit and the data in the current level correction table is used to change the power level of the synthesizer. The current level correction table is either the last level correction that was run or the last table recalled from one of the four level correction register locations. The level correct function compensates for path loss measured during the level correct routine and changes power levels accordingly. Thus, power is at and leveled over the range of frequency points where data was measured during the routine. For example, assume that the level correction factor at 10 GHz is 04 dB (because there is 4 dB of path loss between the RF OUTPUT connector and the test point). Then, if 010 dBm is desired at 10 GHz (and the level correct function is turned on), the synthesizer actually generates 06 dBm so that the power level at the test point is 010 dBm. Notice that path loss correction is independent of power level. The preset condition for this function is o.

Notes
1. When level correction is activated and more power is required at the RF OUTPUT than the synthesizer can deliver, an error message is generated and the UNLVL annunciator lights. This can occur when the external signal path has loss and the power level is set close to maximum leveled power. 2. If the level correct function is on and an output frequency is requested from the synthesizer that is either greater than the stop frequency or less than the start frequency,
an error message will be generated and the level correction factor for the requested frequency will be zero (no correction).

6-35


Keys/Shifted Functions

LVL CR ON/OFF

Equivalent SCPI Commands

CORR:CSET:SEL FDATtableno selects the level correct table as de ned CORR:STATe ON|OFF CORR:CSET:STAT ON|OFF

by the \tableno" parameter. turns user corrections on or o. turns level corrections on or o.

See Also

LVL CR PWR MTR ADDRS RCL CR SAVE CR [SOURce[1]:]CORRection:CSET[:SELect] [SOURce[1]:]CORRection:CSET[:STATe] [SOURce[1]:]CORRection[:STATe] To Use the Level Correct Routine

6-36


MSG

The 4 5 key causes the instrument to display the most recent uncleared error number and a short description of the error to the front panel display.
MSG

When uncleared error messages are in the error queue, the front panel MSG annunciator will be lit. The MSG annunciator will remain lit until all error messages in the queue have been cleared. When an error is read using the 4 5 key, it is cleared as long as the error condition no longer exists. When the 4 5 key is pressed, the most recent error in the error queue will be displayed. Successive presses of the 4 5 key will display any other errors in the queue. Pressing any function key will remove the error from the display. When the synthesizer is set to the preset state, the error queue is cleared.
MSG MSG MSG

Equivalent SCPI Command

SYST:ERR? returns the oldest uncleared error number and message

that is in the error queue.

See Also

Error Messages SYSTem:ERRor?

6-37


MULTIPLIER

5 key and then the Invoking the MULTIPLIER function (pressing the 4 key) allows you to enter a multiplier value so that the frequency display will indicate the frequency at the output of a frequency multiplier.
SHIFT

4FREQ5

NO TE
External equipment is required for frequency multiplication.

When 4 5, 4 5 is pressed, the multiplier value will be displayed across the display in the following format: FREQUENCY MULTIPLIER XXX where XXX is the multiplier value. The allowable range for multiplier values is 1 to 100. The preset value is 1 and the preset up/down arrow increment value is 1. Entering a frequency multiplier value is useful when generating millimeter-wave signals with external multiplier equipment. The display shows the frequency at the output of the external frequency multiplier, not at the synthesizer RF OUTPUT connector. For example, assume a frequency of 30 GHz is required. The synthesizer cannot generate a 30 GHz signal directly, but a frequency doubler can be connected at the RF OUTPUT connector to multiply a 15 GHz signal by two. Setting the multiplier value to two allows you to display the frequency at the output of the multiplier on the synthesizer. Setting the multiplier value to two will display 30 GHz, while the synthesizer is actually generating 15 GHz. Entering a new frequency of 32 GHz will set the synthesizer to 16 GHz.
SHIFT FREQ

6-38


Keys/Shifted Functions

MULTIPLIER

The minimum resolution at the output of the frequency multiplier is the synthesizer minimum resolution multiplied by the frequency multiplier value. As an example, assume a multiplier value of two has been entered and you attempt to enter a frequency of 30,000,001,000 Hz from the numeric keypad. The synthesizer will try to generate 15,000,000,500 Hz. However, the resolution of this signal (assuming Option 1E8 is not installed) is 500 Hz which is ner than the minimum speci ed resolution of 1,000 Hz. The actual output frequency would be rounded to 15,000,001,000 Hz and the display would show 30,000,002,000 Hz.

Equivalent SCPI Command

FREQ:MULT multiplier sets the multiplier value as de ned by the

\multiplier" parameter.

See Also
4FREQ5

[SOURce[1]:]FREQuency:MULTiplier To Generate Millimeter Signals

6-39


Numeric Keypad

The numeric keypad is used to enter a value for the current active function. The number is entered (recognized by the synthesizer) when you press the appropriate units terminator key or the 4 5 (ENTER) key.
Hz

The numeric keypad consists of the numeric keys 4 5 through 4 5, 4 5, 405, and the units terminator keys 4 5, 4 5, 4 5, and 4 5. When making an entry, the parameter change will not take eect until a units terminator key is pressed. The 4 5 key also functions as the units terminator key for dBm and dB. The 4 5 key functions as the terminator key for entries that are unitless (for example, multiplier value entries). When the numeric keypad is used, data is entered for the active parameter. A cursor ( 9 ) will always appear over one of the digits in the active parameter display. If the cursor is absent from the display, the entry hold function is active and the desired function key must be pressed to re-enable the numeric keypad. The knob and 4*5, 4+5 keys can also be used to change the currently active parameter.
0 9 . GHz MHz kHz Hz GHz Hz

See Also
45 45

ENTER ENTRY OFF Knob \To Enter Data with the Numeric Keypad" in Chapter 2

* +

6-40


POWER LEVEL

The 4 synthesizer.

POWER LEVEL

5

key allows you to set the output power level of the

Power level is displayed in the right-hand position of the display. The allowable range for power level entries is 015 dBm to +30 dBm for standard con guration instruments and 0120 dBm to +30 dBm when Option 1E1 is installed (0100 dBm to +30 dBm on HP 83711A/12A).
NOTE
The actual maximum internally leveled output power for your instrument at a given frequency can be found by increasing the synthesizer output power until the UNLVL annunciator lights.

Power level resolution is 0.01 dB. The preset power level value is 0 dBm for standard con guration instruments. For instruments with Option 1E1 installed, the preset power level value is 0110.00 dBm (090.0 dBm on HP 83711A/12A). The preset up/down arrow increment value is 1.00 dBm. The minimum increment value is 0.01 dBm. Three options are available for leveling of the output power. These are 5), 5), external diode detector leveling (4 internal leveling (4 5). Refer to the respective and external power meter leveling (4 reference entries in this chapter for information on the dierent leveling options.
INT LEVEL EXT DIODE EXT METER

6-41


Keys/Shifted Functions

POWER LEVEL

Equivalent SCPI Command

POW level sets the synthesizer output power level as de ned by the

\level" parameter.

See Also
4EXT 4EXT 4INT
DIODE

5 5

METER LEVEL

[SOURce[1]:]POWer[:LEVel] [SOURce[1]:]POWer:UNIT

5

6-42


PRESET

The

4PRESET5

key sets the synthesizer to a known state.

The preset conditions are shown in the following table:
Table 6-1. PRESET Conditions
Parameter Condition Parameter Condition

:CORRection:STATe :CORRection:CSET:SELect :CORR:CSET:STATe :DISPlay:WINDow:STATe :FREQuency:CW :FREQuency:MULTiplier :MEMory:TABLe:SELect :OUTPut:STATe :OUTPut:PROTection:STATe :POWer:LEVel
1 2

OFF FDAT1 OFF ON 3 GHz 1 FDAT1 ON ON 0 dBm
1

:POWer:ALC:SOURce :POWer:ALC:PMETer:LEVel :POWer:ATTenuation:AUTO :POWer:PROTection:STATe :SYSTem:COMMunicate:GPIB:ADDress :SYSTem:COMMunicate:PMETer:ADDress :SYSTem:LANGuage :UNIT:FREQuency :UNIT:POWer :UNIT:TIMe

INT 0 dBm ON OFF 19 13
2 2

\SCPI" HZ DBM S

When Option 1E1 is installed, the preset value for power level is

0

90 dBm for the HP 837311A/12A, and

0

110 dBm for the HP 83711B/12B.

Do not change with preset.

These are default values when RAM memory is lost.

6-43


Keys/Shifted Functions

PRESET

Equivalent SCPI Commands

*RST

sets the synthesizer to a known state.

or
SYST:PRES

See Also

*RST SYSTem:PRESet

6-44


RCL CR

Invoking the RCL CR (recall level correction table) function (pressing the 4 5 key and then the 4 5 key) allows you to recall a previously stored table of level correction data from one of four level correction register locations.
SHIFT EXT METER

5, The allowable range for register locations is 1 through 4. When 4 5 is pressed, the text RECALL FRM TABLE XXX will be shown on 4 the synthesizer display where XXX is the last level correction register number entered. Once you press a valid numeric key (1 through 4) and terminate the entry by pressing 4 5 (ENTER), the table of level correction data will be recalled from the location indicated by the numeric key pressed. If level correction data has not been previously stored to a level correction register, an error message is generated and no change to the current data takes place if you attempt to recall a level correction table from that register.
SHIFT EXT METER Hz

NOTE
The four level correction registers are separate from the instrument state registers. Level correction data is not cleared from memory when the synthesizer is set to the preset state and is unaected if an instrument state is recalled from one of the instrument state registers.

6-45


Keys/Shifted Functions

RCL CR

Equivalent SCPI Command

There is no equivalent SCPI command for the RCL CR function, however, the \SOUR1:CORR:CSET:SEL" command is used to select which level correct table is used to correct power at the synthesizer RF OUTPUT connector.

See Also

LVL CR LVL CR ON/OFF SAVE CR \To Use the Level Correct Routine" in Chapter 3

6-46


RECALL

The 4 5 key allows you to recall a previously stored instrument state from one of ten register locations.
RECALL

5 The allowable range for register locations is 0 through 9. When 4 is pressed, the text RECALL STATE FROM REG XXX will be shown on the instrument display where XXX is the last register number entered. Once you press a valid numeric key (0 through 9) and terminate the entry, the instrument state will be recalled from the location indicated by the numeric key pressed. If the instrument state has not been previously stored to an instrument state register, the synthesizer will be set to the preset state if you attempt to recall the instrument state from that register.
RECALL

Equivalent SCPI Command

*RCL register recalls a previously stored instrument state from the

register de ned by the \register" parameter.

See Also
4PRESET5

*RCL *SAV SAVE

6-47


RF ON/OFF

The o.

4RF

ON/OFF

5

key toggles the signal at the RF OUTPUT connector on and

When the RF OUTPUT is turned o, the text OFF will be indicated in the level display in place of dBm. When the RF output is turned o, the internal oscillators are turned o and the internal RF power shutdown circuit is turned on. When the synthesizer is set to the preset state, the signal at the RF OUTPUT connector is turned on.

Equivalent SCPI Command

OUTP:STAT ON|OFF turns the signal at the RF Output connector on or

o.

See Also

OUTPut[:STATe]

6-48


SAVE

5 5 key and then the 4 Invoking the SAVE function (pressing the 4 key) allows you to save the instrument state in one of ten register locations.
SHIFT RECALL

5, The allowable range for register locations is 0 through 9. When 4 4 5 is pressed, the text SAVE STATE IN REG XXX will be shown on the instrument display where XXX is the last register number entered. Once you press a valid numeric key (0 through 9) and terminate the entry, the instrument state will be saved to the location indicated by the numeric key pressed. Saving the instrument state to a given register location will write over any instrument state previously stored at that location. 5 key is pressed will be All user settings that are aected when the 4 saved. Level correction tables, however, will not be saved. For information on saving level correction tables, refer to \SAVE CR" in this chapter.
SHIFT RECALL PRESET

Equivalent SCPI Command

*SAV register saves the instrument state to the register de ned by

the \register" parameter.

See Also
4RECALL5

*SAV SAVE CR

6-49


SAVE CR

Invoking the SAVE CR (save level correction table) function (pressing the 4 5 key and then the 4 5 key) allows you to save the current table of level correction data in one of four level correction register locations.
SHIFT EXT DIODE

5, The allowable range for register locations is 1 through 4. When 4 4 5 is pressed, the text SAVE IN TABLE XXX will be shown on the synthesizer display where XXX is the last level correction register number entered. Once you press a valid numeric key (1 through 4) and terminate the entry by pressing 4 5 (ENTER), the table of level correction data will be saved to the location indicated by the numeric key pressed. Saving the table of level correction data to a given level correction register location will write over any level correction data previously stored at that location.
SHIFT EXT DIODE Hz

NOTE
The four level correction registers are separate from the instrument state registers. Level correction data is not cleared from memory when the synthesizer is set to the preset state and is unaected if an instrument state is recalled from one of the instrument state registers.

6-50


Keys/Shifted Functions

SAVE CR

Equivalent SCPI Command

There is no equivalent SCPI command for the SAVE CR function, however, the \MEM:TABL:FREQ" and \MEM:TABL:LOSS:MAGN" commands can be used to load frequency and correction factor points into a selected level correct table.

See Also

LVL CR LVL CR ON/OFF RCL CR \To Use the Level Correct Routine" in Chapter 3

6-51


SHIFT

The 4 5 key causes the alternate function of any key pressed directly after it to be executed.
SHIFT

5 key is pressed, the word SHIFT will be displayed in the When the 4 5 key will execute the display. The next key that is pressed after the 4 function that is indicated by the text that is above the key. As an example, 5 key. MULTIPLIER is note that MULTIPLIER appears directly above the 4 5 will enable 5, 4 5 key. Pressing 4 the alternate function of the 4 the frequency multiplier entry mode. 5 key is pressed prior to pressing a key that has no alternate If the 4 function, SHIFT will be canceled from the display and no action will be taken by the instrument. 5 key accidentally, pressing it again will cancel it If you press the 4 without altering synthesizer operation.
SHIFT SHIFT FREQ FREQ SHIFT FREQ SHIFT SHIFT

Equivalent SCPI Command
5 key, however, sending There is no equivalent SCPI command for the 4 5 the \SYST:KEY 0" command is eectively the same as pressing the 4 key.
SHIFT SHIFT

See Also

SYSTem:KEY

6-52


SPCL

The 4 5 key initiates activation of special functions. Once a special function has been activated, it can be turned on or o, or its parameter value can be changed.
SPCL

5 key is pressed the left-most display shows the last special When the 4 function used. Press 4*5 and 4+5 keys to scroll through the list until the 5 key and a number desired special function is reached, or press the 4 on the keyboard between 1 to 52, that corresponds to the desired special function. When the desired special function is reached, it can be changed by one of two methods, depending on whether it is an ON/OFF function, or has an associated value. If the function has only ON and OFF values, it can be changed by pressing 5 key and using the 4 5 and 4 5 keys, or by pressing the 4 the 4*5 and 4+5 keys. If the function has an associated parameter, the parameter can be modi ed by rst pressing 4 5, and then adjusting the parameter by using either the 4*5 and 4+5 keys, the RPG, or by entering a value from the keyboard. When entering from the keyboard, the parameter entry is terminated the same way as any parameter entry. When the synthesizer is set to the preset state, the currently active special function is set to 1 (SHOW INFO).
SPCL SPCL SPCL ON SPCL OFF ENTER ENTER

Equivalent SCPI Command
5 key, however, sending There is no equivalent SCPI command for the 4 the \SYST:KEY 9" command is eectively the same as pressing the 4 5 key.
SPCL SPCL

6-53


SPCL OFF (kHz)

The SPCL OFF key (actually the secondary function of the to turn on/o-type special functions o.
kHz

4kHz5

key) is used

The 4 5 (SPCL OFF) key is used to turn on/o-type special functions o when they are displayed in the display. The on/o-type special functions are ATTEN HOLD ON/OFF and CLEAR DISPLAY. The SPCL OFF key will turn the special function o whether it has been selected via the numeric keypad or by using the knob or 4*5, 4+5 keys. When the special function has been turned o, the word \OFF" will be shown in the right side of the display and the SPCL annunciator will turn o if no other special functions are selected.

Equivalent SCPI Command

There is no equivalent SCPI command for the SPCL OFF key, however, sending the \SYST:KEY 50" command is eectively the same as pressing the SPCL OFF key.

6-54


Keys/Shifted Functions

SPCL OFF (kHz)

See Also

ATTEN HOLD ON/OFF CLEAR DISPLAY
4SPCL5

SPCL ON SYSTem:KEY

6-55


SPCL ON (MHz)

The SPCL ON key (actually the secondary function of the to turn on/o-type special functions on.
MHz

4MHz5

key) is used

The 4 5 (SPCL ON) key is used to turn on/o-type special functions on when they are displayed in the display. The on/o-type special functions are ATTEN HOLD ON/OFF and CLEAR DISPLAY. The SPCL ON key will turn the special function on whether it has been selected via the numeric keypad or by using the knob or 4*5, 4+5 keys. When the special function has been turned on, the word \ON" will be shown in the right side of the display and the SPCL annunciator will be lit.

Equivalent SCPI Command

There is no equivalent SCPI command for the SPCL ON key, however, sending the \SYST:KEY 42" command is eectively the same as pressing the SPCL ON key.

See Also

ATTEN HOLD ON/OFF CLEAR DISPLAY
4SPCL5

SPCL OFF SYSTem:KEY

6-56


STEP SIZE

The 4 5 key enables you to change the increment value for the current active parameter. The increment value is the value that the current parameter will be increased or decreased by when the 4*5or 4+5 keys are pressed.
STEP SIZE

The preset increment value for each parameter can be found under the pertinent key/function entries in this section. 5 (ENTRY OFF), 4)5) function has been enabled, you will not be If the 4 5 key and pressing able to change any increment values using the 4 the 4*5 or 4+5 keys will have no eect on the active parameter.
SHIFT STEP SIZE

Equivalent SCPI Commands

sets frequency increment value as de ned by the \increment" parameter. FREQ:MULT:STEP increment sets multiplier increment value as de ned by the \increment" parameter. POW:STEP increment sets power level increment value as de ned by the \increment" parameter. POW:ALC:PMET:STEP increment sets external power meter leveling increment value as de ned by the \increment" parameter.
FREQ:STEP increment

6-57


Keys/Shifted Functions

See Also
45 45

ENTRY OFF

* +

6-58


7

Special Functions


Special Functions

This chapter contains detailed information on the various special functions available for the synthesizer. Special functions are hidden during normal 5 key synthesizer operation and can only be invoked by pressing the 4 and then entering the special function number or by scrolling through the list with the 4*5 and 4+5 keys. Note that special functions are organized in this chapter alphabetically by special function name, not by special function number.
SPCL

7-2


ATTEN HOLD ON/OFF

5, 4 5, 4 5 Invoking the ATTEN HOLD ON/OFF function (pressing 4 (ENTER)) selects the attenuator hold function. This function toggles between locking and unlocking the 10 dB step attenuator in its current setting.
SPCL 2 Hz

NO TE
The attenuator hold function is only available if Option 1E1 is installed.

5, 4 5, 4 5 (ENTER) is pressed, the display shows the following: When 4 ATTEN HOLD XXX where XXX is the current state of the step attenuator (ON=Locked and OFF=Unlocked). Once the attenuator hold function is enabled, pressing 4 5 (SPCL ON) locks the synthesizer step attenuator at its current setting. Pressing 4 5 (SPCL OFF) unlocks the step attenuator. When the synthesizer is set to the preset state, the attenuator hold function is turned o.
SPCL 2 Hz MHz kHz

7-3


Special Functions

ATTEN HOLD ON/OFF

Applications

The attenuator hold function can be used to extend the vernier range to prevent the step attenuator from switching between two attenuator settings. Locking the step attenuator keeps the attenuator from switching between the two levels as leveled power is varied above and below the threshold level, thus saving wear on the attenuator. Refer to the speci cation table in Chapter 4 for the level at which the attenuator switches.

Advantages

Locking the step attenuator prevents switching between two levels when the leveled output power is set near an attenuator switching threshold.

7-4


ATTEN HOLD ON/OFF

Special Functions

Disadvantages

When the step attenuator is locked, the output power dynamic range is limited to the vernier range at the current output frequency. Locking the step attenuator typically extends the lower limit of the vernier range by 5 dB. The upper limit of the vernier range is the synthesizer maximum output power which changes with frequency. The minimum dynamic range when the attenuator hold function is on is typically 19 dB.

NOTE
The attenuator hold function on or o will not activate when the synthesizer is in the external diode detector leveling or external power meter leveling mode.

Equivalent SCPI Command

POW:ATT:AUTO ON|OFF turns the attenuator hold function on or o.

When the parameter is set to \on", the attenuator hold function is o and when the parameter is set to \o", the attenuator hold function is on.

See Also
4POWER
LEVEL

[SOURce[1]:]POWer:ATTenuation:AUTO 7-5

5


CLEAR DISPLAY

5, 4 5, 4 5 (ENTER) enables Invoking the clear display function (pressing 4 you to turn o the synthesizer uorescent display and LED annunciators or turn them on if they are currently turned o.
SPCL 3 Hz

When 4 5, 4 5, 4 5 (ENTER) is pressed, the display will show the following: CLEAR DISPLAY XXX where XXX is \ON" if the clear display function is currently on and \OFF" if the clear display function is currently o. Once the clear display function has been enabled, pressing SPCL ON blanks the display and LED annunciators and DISPLAY BLANKED is shown across the display. The display state is stored in the instrument state registers along with other instrument state data, so if sensitive instrument settings are stored to a register, the settings are not revealed when the register is recalled. When the synthesizer is set to the preset state, the display is restored if it had been previously cleared.
SPCL 3 Hz

NO TE
Cycling the POWER switch o and then on will not restore the display.

7-6


CLEAR DISPLAY

Special Functions

Equivalent SCPI Command

DISP:STAT ON|OFF turns the display on or o.

See Also

Display DISPlay[:WINDow][:STATe]

7-7


ERASE MEMORY

5, 4 5, 4 5 (ENTER) Invoking the ERASE MEMORY function (pressing 4 clears all application-speci c information from synthesizer memory.
SPCL 4 Hz

When 4 5, 4 5, 4 5 (ENTER) is pressed, all user settings are set to the preset state, save/recall registers are erased, and level correction tables are cleared. The ERASE MEMORY function does not clear factory calibration data stored in the EEPROM. When the ERASE MEMORY function is invoked, the display will momentarily display the following: INITIALIZING MEMORY
SPCL 4 Hz

Application

The ERASE MEMORY function is useful when removing the synthesizer from a secure area as the setup history of the synthesizer will be erased.

Equivalent SCPI Command

MEM:RAM:INIT clears synthesizer memory.

7-8


ERASE MEMORY

Special Functions

See Also
4PRESET5

RCL CR

4RECALL5

SAVE SAVE CR MEMory:RAM:INITialize[:ALL]

7-9


PWR MTR ADDRS

Invoking the PWR MTR ADDRS (power meter address) function (pressing 4 5, 4 5, 4 5 (ENTER)) changes the HP-IB address that the synthesizer uses when communicating with an external power meter during the level correct routine.
SPCL 20 Hz

When 4 5, 4 5, 4 5 is pressed, you will be prompted to enter the address that the synthesizer will use when communicating with the external power meter or when receiving data from the external power meter during the level correct routine. The valid power meter address range is 00 to 30 (decimal). The external power meter HP-IB address set at the factory is 13. Setting the synthesizer to the preset state will not modify the address.
SPCL 20 Hz

NOTE
This function set the address at the power meter. You should refer to the power meter manual for information on how to change the power meter address.
does not

7-10


PWR MTR ADDRS

Special Functions

Equivalent SCPI Command

SYST:COMM:PMET:ADDR address changes the HP-IB address that the

synthesizer uses when communicating with an external power meter as de ned by the \address" parameter.

See Also

LVL CR LVL CR ON/OFF PWR MTR SELECT RCL CR SAVE CR SYSTem:COMMunicate:PMETer:ADDRess \To Use the Level Correct Routine" in Chapter 3

7-11


PWR MTR SELECT

5, 4 5, 4 5 (ENTER)) Invoking the power meter select function (pressing 4 allows you to set the programming language that the synthesizer will use when communicating with the power meter during the level correct routine.
SPCL 21 Hz

When 4 5, 4 5, 4 5 (ENTER) is pressed, the display shows the following: POWER METER IS XXX where XXX is the currently selected power meter language (HP70100A, HP437B, HP438A, or SCPI). Once the power meter select function is enabled, you can use the 4*5 or 4+5 keys to choose one of the four power meter language options. When \SCPI" is chosen, the synthesizer will communicate with any SCPI-compatible power meter. When the synthesizer is set to the preset state, the power meter language chosen with this command is not changed.
SPCL 21 Hz

Equivalent SCPI Command

There is no equivalent SCPI command for the power meter select function.

7-12


PWR MTR SELECT

Special Functions

See Also

LVL CR LVL CR ON/OFF PWR MTR ADDRS RCL CR SAVE CR \To Use the Level Correct Routine" in Chapter 3

7-13


REMOTE LANGUAGE

5, 4 5, 4 5 (ENTER)) Invoking the remote language function (pressing 4 allows you to manually choose either SCPI (Standard Commands for Programmable Instruments) or HP 8673 compatible language.
SPCL 25 HZ

When 4 5, 4 5, 4 5 (ENTER) is pressed, the left-most display shows the following: REMOTE LANG IS XXX where XXX is the currently selected language (SCPI or HP 8673). Once the remote language function is enabled, you can use the 4*5 or 4+5 keys to toggle between SCPI or HP 8673 language options. When the synthesizer is set to the preset state, the language chosen with this command is not changed. The language will default to SCPI when the ERASE MEMORY function is executed.
SPCL 25 HZ

Equivalent SCPI Command

SYSTem:LANGuage "SCPI"|"COMP=8673"

selects either SCPI or HP 8673 compatible language.

See Also

ERASE MEMORY

7-14


SELF TEST

5, 4 5, 4 5 (ENTER)) enables you Invoking the self test function (pressing 4 to cause the synthesizer to run a functional veri cation (self test) on itself.
SPCL 5 Hz

When 4 5, 4 5, 4 5 (ENTER) 4!5 (ENTRY OFF) is pressed, the text SELF TEST?, PRESS ENTER will be shown on the synthesizer display. Pressing 4 5 (ENTER) will then cause all self test segments to be run. Pressing any key other than 4 5 will cause the self test not to run. If any of the self test segments fail, error messages will be placed in the error queue to explain the failures. The error messages can then be read either via the front panel or using programming commands. If a particular self test segment failure makes running subsequent self test segments impossible, the self tests will abort.
SPCL 5 Hz Hz Hz

Notes
1. Running the self test will result in the synthesizer being set to the preset state. 2. It is recommended that you clear the error queue before running the self test. 3. For more information on reading the contents of the error queue, refer to \To Read the Contents of the Error Queue" in Chapter 2 of this manual. 4. Refer to the listing of the test segments that are run during the self test.
HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Service Guide

for a

7-15


Special Functions

SELF TEST

Equivalent SCPI Command

*TST? causes the synthesizer to run a functional veri cation (self test)

on itself.

See Also
4MSG5

SYSTem:ERRor? \To Read the Contents of the Error Queue" in Chapter 2 *TST?

7-16


Service-Related Special Functions

The synthesizer rmware contains several service related routines that can be used when diagnosing faults in its circuitry. Use of these special functions is detailed in the HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Service Guide (Option 0B3; HP part number 83711-90134). The service-related special functions are listed in the following table. They 5, f-number, 4 5 (ENTER) where f-number can be accessed by pressing 4 is a number from the rst column of the table. Details on how to use each special function can be found in the service manual.
SPCL Hz

Table 7-1. Service-Related Special Functions
Special Function Number Special Function Name Description

60 61 62 63 70 71

DIRECT CONTROL METER NODE LO FREQ OFFSET FREQ HOLD POWER ON/OFF CAL YIG OSC

Allows direct manipulation of certain data bits within the synthesizer circuitry. Allows entry of a measurement node where the internal volt/ammeter will take a measurement. Displays the synthesizer internal LO phase locked loop frequency. Displays the synthesizer internal Oset phase locked loop frequency. Holds the power loop at its current setting. Recalibrates the YIG oscillator and stores the new tuning curve in EEPROM.

See Also

HP 83711A/12A and 83711B/12B Synthesized CW Generators Service Guide

7-17


SHOW INFO

5, 4 5, Invoking the SHOW INFO (show information) function (pressing 4 4 5 (ENTER)) displays the software version number and instrument serial number on the synthesizer display.
SPCL 1 Hz

When 4 5, 4 5, 4 5 is pressed, the following text will be shown on the synthesizer display. SW: X.X SERIAL: YYYYYYYYYY where X.X is the version number of the software currently installed in the synthesizer and YYYYYYYYYY is the synthesizer serial number.
SPCL 1 Hz

Equivalent SCPI Command

There is no equivalent SCPI command for the SHOW INFO function, however, the \*IDN?" query returns the synthesizer model number, serial number, and rmware revision number.

See Also

*IDN?

7-18


8

Error Messages


Error Messages

If an error condition occurs in the synthesizer, it will always be reported to both the front panel and HP-IB error queues. These two queues are viewed and managed separately. The 4 5 key is used to view the contents of the front panel error queue. The HP-IB query \SYSTem:ERRor?" is used to view the contents of the HP-IB error queue. If there are any error messages in the front panel error queue, the front panel MSG annunciator will be lit. Pressing the 4 5 key repeatedly until the MSG annunciator turns o will empty the front panel error queue. The 4 5 key has no aect on the HP-IB error queue. Emptying the HP-IB error queue has no aect on the front panel queue, therefore, it will not aect the MSG annunciator. There are some special error types that are called permanent errors. Permanent errors remain in the error queues until the error condition is cleared. Pressing the 4 5 key will empty the front panel error queue, but the permanent errors will be re-reported if the error conditions still exist. In the HP-IB error queue, the permanent errors are re-reported after the message, 0,\No error" is read using the \SYSTem:ERRor?" query or after the \*CLS" command is executed.
MSG MSG MSG MSG

8-2


Error Messages List

The list of error messages in this chapter lists all of the error messages associated with synthesizer operation. An example of the error format found in the list of error messages is as follows: 2003

0222,"Data out of range;CW FREQ(2003)"
Select a CW frequency that is within range of the installed options. If other modules or options are installed that extend the CW frequency range of the signal generator, this frequency range will be extended also.

The following explains each element of an error message listing. Manual Error Number 0 The number 2003 to the left and in the parenthesis is called the Manual Error Number. The error message list is organized in ascending order o the manual error number. The manual error number will always be found in the parenthesis contained in the message. Error Message 0 The bold text 0222,\Data out of range;CW FREQ(2003)" is the error message. When the 4 5 key is pressed, the error message is displayed in the left-most display. The entire message is returned by the HP-IB query \SYSTem:ERRor?". The error message contains the following parts: SCPI Error Number 0 The standard SCPI error number (0222 in the example) usually diers from the manual error number because the manual error number is unique for every possible message. Standard SCPI error numbers are always negative (except for 0, \No error"). If there is no standard SCPI error number for a message, the manual error number replaces it in the error message. SCPI Error Message 0 The SCPI error message is Data out of range in the example. Detailed Description 0 All information after the semicolon is a detailed description of what exactly caused the error. In the example, CW FREQ tells you that CW frequency was out of range. If no detailed description exists, it will be omitted from the message.
MSG

8-3


Error Messages

Error Messages List

Action Required 0

The text that appears below each error message listing contains corrective actions that should be followed in order to correct the error condition.

Notes
1. For more information related to error messages, refer to \To Read the Contents of the Error Queue" in Chapter 2 or the \MSG" reference entry in Chapter 6 of this manual, and the \SYSTem:ERRor?" reference entry in Chapter 1 of this manual. 2. Error messages related to hardware failures are listed in the (HP part number 83711-90134).
HP 83711A/12A and HP 83711B/12B Synthesized CW Generators Service Guide

8-4


Messages

The following pages list all error messages in ascending manual error number order: 0440 0440,"Query UNTERMINATED after inde nite response;(0440)" Correct the HP-IB controller program so that the query that returns inde nite length block data is the last item on the program line. 0430 0430,"Query DEADLOCKED;(0430)" Correct the HP-IB controller program so that no more than eight queries are executed within the same line of the program. 0420 0420,"Query UNTERMINATED;(0420)" Correct the HP-IB controller program so that the controller terminates commands with the newline character (NL) before the controller attempts to read query response data. 0410 0410,"Query INTERRUPTED;(0410)" Check the HP-IB controller program to see if the controller is programmed to read the entire query response data before issuing a subsequent command. 0400 0400,"Query error;(0400)" Some problem occurred while parsing an HP-IB query. Insure that your programming is correct and try the query again. Look at 0440 through 0400 for types of problems to look for. 0350 0350,"Queue over ow" The error queue over owed at this point and this message replaced the 16th error message. No action is required. Note: To clear the HP-IB error queue, use *CLS. 0330 0330,"Self test failed;(0330)" See the explanation for error number 4000. 0315 0315,"Con guration memory lost;(0315)" See error 1803. 0314 0314,"Save/recall memory lost;(0314)" 8-5


Error Messages

Messages

0311 0311,"Memory error;(0311)" 0310 0310,"System error;(0310)" 0300 0300,"Device speci c error;(0300)"
Some problem occurred while parsing an HP-IB command or query. Insure that your programming is correct and try the command again. See error 1803.

See error 1803.

0278 0278,"Macro header not found;(0278)"

A remote command or query could not be executed because an error occurred in the synthesizer.

0277 0277,"Macro rede nition not allowed;(0277)" 0276 0276,"Macro recursion error;(0276)"

A *GMC? or *RMC macro label could not be found in the list of de ned macro labels. Use *LMC? to get a list of all the currently de ned macro labels.

Indicates that a macro label in the *DMC command could no be de ned because the macro label was already de ned.

0275 0275,"Macro de nition too long;(0275)" 0274 0274,"Macro parameter error;(0274)" 0273 0273,"Illegal macro label;(0273)"

The nesting/recursion of macros is deeper than 4 levels. Don't use more than 4 levels when de ning macros of macros.

The macro de nition must be 255 characters or less. A macro parameter placeholder was improperly used.

0272 0272,"Macro execution error;(0272)"

Indicates that a macro label de ned in the *DMC command has a legal string syntax; but, it is too long. It is the same as a common command header, or contain invalid header syntax. Indicates that a syntactically legal macro program data sequence could not be executed due to some error in the macro de nition.

8-6


Error Messages

Messages

0271 0271,"Macro syntax error;(0271)" 0270 0270,"Macro error;(0270)"
Indicates that a syntax error exists in the macro de nition. An error occurred while attempting to de ne, query or use a macro. Check that the macros are correct using *LMC? and *GMC?.

0261 0261,"Math error in expression;(0261)" 0260 0260,"Expression error;(0260)"

An expression could not be evaluated due to a math error; for example, a divide-by-zero was attempted. An expression could not be evaluated because it contains an error. The requested hardware does not exist in the synthesizer. Use *OPT? to check which options are installed. The remote command or query could not be executed because of a hardware error. See error 731.

0241 0241,"Hardware missing;(0241)" 0240 0240,"Hardware error;(0240)"

0226 0226,"Tables not same length;(0226)" 0225 0225,"Out of memory;(0225)"

0224 0224,"Illegal parameter value;(0224)" 0223 0223,"Too much data;(0223)"

The synthesizer has run out of memory. The memory requested has not been allocated. Correct the HP-IB controller program so that the data included with the HP-IB command is an acceptable parameter for the command. Correct the HP-IB controller program so that there is less data on a single command line. The synthesizer does not have enough memory to buer it all.

0222 0222,"Data out of range;(0222)"

8-7


Error Messages

Messages

0221 0221,"Settings con ict;(0221)"

The parameter data was out of range. Unlike other 0222 errors, details are not known about the command or query which caused this error.

0220 0220,"Parameter error;(0220)" 0213 0213,"Init ignored;(0213)"

The current synthesizer state does not allow the remote command or query to be executed.

The parameter included with the remote command or query is incorrect.

0212 0212,"Arm ignored;(0212)"

Indicates that an initiate was ignored because a trigger was already in progress. An arming signal was received and recognized but was ignored.

0211 0211,"Trigger ignored;(0211)"

0210 0210,"Trigger error;(0210)"

A GET, *TRG or triggering signal was received and recognized but was ignored. Currently, there is no bus trigger capability in the synthesizer. A trigger error occurred in the synthesizer. The remote command or query cannot be executed when the synthesizer is in local mode.

0201 0201,"Invalid while in local;(0201)" 0200 0200,"Execution error;(0200)"

0184 0184,"Macro parameter error;(0184)"

Some problem occurred while executing an HP-IB command or query. Insure that your programming is correct and try the command again. Indicates that a command inside the macro de nition had the wrong number or type of parameters. Indicates that the program message sequence sent with *DMC or *DDT command is syntactically invalid.

0183 0183,"Invalid inside macro de nition;(0183)"

8-8


Error Messages

Messages

0181 0181,"Invalid outside macro de nition;(0181)" 0180 0180,"Macro error;(0180)"
Indicates that a macro parameter placeholder was encountered outside of the macro de nition. An error occurred while attempting to de ne, query or use a macro. Check that the macros are correct using *LMC? and *GMC?. Correct the HP-IB controller program so that the data included with the HP-IB command does not contain parentheses. The expression contained a syntax error like unmatched parenthesis or an illegal character. The expression contains a syntax error.

0178 0178,"Expression data not allowed;(0178)" 0171 0171,"Invalid expression;(0171)" 0170 0170,"Expression error;(0170)"

0168 0168,"Block data not allowed;(0168)" 0161 0161,"Invalid block data;(0161)"
Correct the HP-IB controller program so that the data included with the HP-IB command does not contain block data (no # character).

0160 0160,"Block data error;(0160)"

Correct the HP-IB controller program so that it contains a correct block data type. A block data type should begin with \#" followed by a number. The block data contains a syntax error.

0158 0158,"String data not allowed;(0158)"
Correct the HP-IB controller program so that the data included with the HP-IB command does not contain string data (no single or double quote characters). Correct the HP-IB controller program so that the string data included with the HP-IB command is terminated with a single or double quote. The terminating quote must be the same as the leading quote of the 8-9

0151 0151,"Invalid string data;(0151)"


Error Messages

Messages

0150 0150,"String data error;(0150)"

string. A string can also be valid if invalid characters are contained in it. The string data was too long to be buered in the synthesizer string data area.

0148 0148,"Character data not allowed;(0148)" 0144 0144,"Character data too long;(0144)" 0141 0141,"Invalid character data;(0141)" 0140 0140,"Character data error;(0140)" 0138 0138,"Sux not allowed;(0138)"

Correct the HP-IB controller program so that the data included with the HP-IB command is not character data. The character data element contains more than 12 characters. Either the character data element contains an invalid character or the particular element is not valid for the command or query. The character data contains a syntax error. Correct the HP-IB controller program so that the decimal data included with the HP-IB command does not use a sux. Use exponential notation instead. The sux contained more than 12 characters.

0134 0134,"Sux too long;(0134)" 0131 0131,"Invalid sux;(0131)"

0130 0130,"Sux error;(0130)"

Correct the HP-IB controller program so that the decimal data included with the HP-IB command contains a valid sux for that command or query. The sux contains a syntax error.

0128 0128,"Numeric data not allowed;(0128)"
Correct the HP-IB controller program so that the data included with the HP-IB command is not numeric data.

8-10


Error Messages

Messages

0124 0124,"Too many digits;(0124)" 0123 0123,"Exponent too large;(0123)"
The mantissa of a decimal numeric data element contained more than 255 digits excluding leading zeros. The magnitude of the exponent was larger than 32000.

0121 0121,"Invalid character in number;(0121)"

0120 0120,"Numeric data error;(0120)"

Correct the HP-IB controller program so that the decimal data or non-decimal numeric included with the HP-IB command contains the correct numeric characters. An invalid numeric or non-decimal numeric was parsed but it was syntactically invalid. Indicates that a header sux was too large.

0114 0114,"Header sux out of range;(0114)" 0113 0113,"Unde ned header;(0113)"

0112 0112,"Program mnemonic too long;(0112)" 0111 0111,"Header separator error;(0111)"
An error was detected in the header.

The header is syntactically correct, but it is unde ned for the synthesizer. The header contains more than 12 characters. An illegal header separator was encountered while parsing the header.

0110 0110,"Command header error;(0110)" 0109 0109,"Missing parameter;(0109)"

0108 0108,"Parameter not allowed;(0108)"

This error indicates that an HP-IB command or query has too few parameters. Correct the HP-IB controller program so that the HP-IB command or query contains the correct number of parameters.

8-11


Error Messages

Messages

0105 0105,"GET not allowed;(0105)" 0104 0104,"Data type error;(0104)"

This error indicates that an HP-IB command or query has too many parameters. Correct the HP-IB controller program so that the HP-IB command or query contains the correct number of parameters.

Correct the HP-IB controller program so that the group execute trigger does not occur within a line of HP-IB program code. The parser recognized a data element dierent than one allowed. For example, numeric or string data was expected but block data was encountered.

0103 0103,"Invalid separator;(0103)" 0102 0102,"Syntax error;(0102)"

A separator was expected but an illegal character was encountered. For example, the space is missing from the following: FREQ.01GHz.

0101 0101,"Invalid character;(0101)" 0100 0100,"Command error;(0100)"
0 110

An unrecognized command or data type was encountered.

A syntactic element contains a character which is invalid for that type. For example, a header containing an ampersand would give this error.

Some problem occurred while parsing an HP-IB command or query. Insure that your programming is correct and try the command again.

0,"No error"

The error queue contains no errors.

110,"EEPROM unprotected;(110)"
The PG switch is set to 0 which leaves the EEPROM unprotected. Open up the synthesizer and switch the PG switch to 1. This error message is only a warning.

511

511,"YTO cal data init error;(511)"
The YIG oscillator factory calibration data checksum was incorrect. A new YIG calibration should be performed or else the instrument may be unable to attain lock at some frequencies.

8-12


Error Messages

Messages

600

600,"ALC loop went unleveled;(600)"
Power is set to a level that is higher than the instrument can supply. This is usually due to attenuator hold and the power is set to a value that requires the vernier to be operating out of its speci ed range. Change the power level or turn o attenuator hold. This is a \permanent" error.

601

601,"Hardware driver Power limit;(601)"
Due to instrument specials such as attenuator hold, the circuits cannot supply the speci ed power. Change the power level or turn o attenuator hold. This is a \permanent" error.

602

602,"Vernier has been set to the limit;(602)"
Due to instrument options such as attenuator hold, the circuits cannot supply the speci ed power. The vernier has been limited to a valid value. Change the power level or turn o attenuator hold. This is a \permanent" error.

603

603,"RF on/o command not valid;(603)"
An invalid request to turn o RF power was ignored by the instrument.

604

604,"Atten driver error while setting level;(604)"
The attenuators could not be set to the range requested. Change output power to a valid setting.

605

605,"Vernier driver error while setting level;(605)" 606,"Level is not in guaranteed range.;(606)"

The vernier value requested was not possible. Change output power to a valid setting. 606 The power level requested is beyond speci cations and may be invalid. This could be due to a very low vernier setting required when attenuator hold is active. This is a \permanent" error. 608

608,"Attenuator not set before Ext Meter mode;(608)"
The attenuator range must match that of the meter range desired for external meter ALC mode. Turn o attenuator hold mode and make sure the power meter is in range hold before entering external power meter mode. 8-13


Error Messages

Messages

610

610,"Track and hold failed, level is invalid;(610)"
Power level was too high to do a power level setting in pulse or scan AM mode. Try setting power to a lower value.

611

611,"Track and hold failed, level is invalid;(611)"
Power level was too high to do a power level setting in pulse or scan AM mode. Try setting power to a lower value.

650

650,"PG switch not set to 0;(650)"

ALC calibration data was not saved in EEPROM because the PG switch was protecting the EEPROM from \writes". Open up the synthesizer and switch the PG switch to 0. 651

651,"Invalid vernier cal data for 1-20 GHz;(651)"

Valid vernier calibration data is not available for the 1-20 GHz band. If you need to use this frequency range, see the explanation for error number 4000. 652

652,"ALC term veri cation after EEPROM write;(652)" 653,"Invalid vernier cal data for 0.01-1 GHz;(653)"

ALC vernier calibration data was not written into EEPROM correctly. Try writing the data into the synthesizer again. 653 Valid vernier calibration data is not available for the 0.01-1 GHz band. If you need to use this frequency range, see the explanation for error number 4000. 655

655,"PG switch not set to 0;(655)"
Factory frequency correction data was not saved in EEPROM because the PG switch was protecting the EEPROM from \writes". Open up the synthesizer and switch the PG switch to 0.

656

656,"Factory atness cal data veri cation;(656)"
Factory frequency level calibration data was not written into EEPROM correctly. Try writing the data into the synthesizer again.

657

657,"Factory atness cal data is invalid;(657)"
A valid factory frequency level calibration is not available for one or more of the frequency bands and/or attenuator settings. See the explanation for error number 4000.

8-14


Error Messages

Messages

670

670,"Meter power input is out of range;(670)"
The ALC input is not a valid level. The power meter range may be wrong. This is a \permanent" error.

700

700,"Hardware driver Frequency limit;(700)" 701,"Lo synthesizer set error;(701)"

The frequency entered cannot be generated by the synthesizer with the set of options available. 701 The LO synthesizer cannot be set to the level requested. Enter a new frequency. 702

702,"Oset synthesizer set error;(702)"
The oset synthesizer cannot be set to the level requested. Enter a new frequency.

704

704,"YTO driver set error;(704)"
The YIG oscillator cannot be set to the level requested. Enter a new frequency.

706

706,"Low pass lter set error;(706)"

The low pass lter cannot be set to the requested setting. Enter a new frequency. 709

709,"Hardware driver Frequency limit;(709)"
The frequency entered cannot be generated by the Frequency Extension module.

710

710,"LO synthesizer went out of lock;(710)"
The LO synthesizer went out of lock. This may be due to hookup or disconnection of an external time base. Enter a dierent RF frequency and then set the frequency back to the desired value to re-lock. This is a \permanent" error.

711

711,"Oset synthesizer went out of lock;(711)"
The oset synthesizer board was unable to attain lock. Enter a dierent RF frequency and then set the frequency back to the desired value to re-lock. This is a \permanent" error.

712

712,"Frequency loop went out of lock;(712)"
8-15


Error Messages

Messages

Enter a dierent RF frequency and then set the frequency back to the desired value to re-lock. This is a \permanent" error. 730

730,"Invalid data in level correct table;(730)"
The active level correction table has no data in it. Select a level correction table with valid data, perform an automatic level correction to get valid data into the active table, or use HP-IB to load the active table. This is a \permanent" error.

731

0226,"Tables not same length;Level correct(731)"

The active level correction table has a mismatch between the number of frequencies stored and the number of losses stored. Select a level correction table with valid data, perform an automatic level correction to get valid data into the active table, or use HP-IB to load tables with the same length. This is a \permanent" error. 732

732,"Same frequencies with dierent losses;(732)"
The active level correction table has duplicate dierent losses. Select a level correction table an automatic level correction to get valid data or use HP-IB to load tables with non-duplicate \permanent" error. frequencies with with valid data, perform into the active table, frequencies. This is a

733

733,"Frequency table not in ascending order;(733)"
The MEM:TABL:FREQ command did not contain frequencies in ascending order. The whole MEM:TABL:FREQ command was rejected, leaving the old selected table unaltered.

734

734,"Frequency table not in ascending order;(734)"

The active level correction table does not contain frequencies in ascending order. Select a level correction table with valid data, perform an automatic level correction to get valid data into the active table, or use HP-IB to load a table with ascending ordered frequencies. This is a \permanent" error. 735

735,"Level correct points less than 2;(735)"
The number of points in a level correction table must be from 2 to 401. Either too few points were entered or duplicate frequencies caused the number of \real" points to shrink below 2.

736 8-16

736,"Factory level corr 1-20 GHz, 1-9 table;(736)"


Error Messages

Messages

Factory frequency level correction data for 1-20 GHz band, non-thru paths are not valid. If you need to use this frequency range and you are using a step attenuator, see the explanation for error number 4000. 737

737,"Factory level corr 0.01-1 GHz, 1-9 table;(737)"
Factory frequency level correction data for 0.01-1 GHz band, non-thru paths are not valid. If you need to use this frequency range and you are using a step attenuator, see the explanation for error number 4000.

738

738,"Factory level corr 0.01-1 GHz, 0 dB table;(738)"
Factory frequency level correction data for 0.01-1 GHz band, thru path is not valid. If you need to use this frequency range and you work with output levels that don't use the step attenuator, see the explanation for error number 4000.

739

739,"Invalid data in table, not recalled;(739)"
This error indicates that a level correction table recall failed. Try selecting the same level correction table again. If this error message persists you will have to recreated the saved table.

740

740,"Another controller is on the HP-IB bus;(740)"
An automatic level correction was attempted but failed because there is a controller on the HP-IB bus. Remove all controllers from the HP-IB bus and try again.

741

741,"No HP-IB devices found;(741)"
An automatic level correction was attempted but failed to nd any other devices on the HP-IB bus. Connect the desired power meter to the HP-IB bus and try again. Check the HP-IB cable(s) for loose connections.

742

742,"Errors in cleanup of HP-IB;(742)"
When exiting the automatic level correction, the synthesizer failed to nish resetting the HP-IB bus and presetting the power meter. Make sure the power meter address matches the power meter address setting on the synthesizer. Check the HP-IB cable(s) for loose connections.

743

743,"No HP-IB devices found;(743)"

An automatic level correction was attempted but failed to nd any other devices on the HP-IB bus. Connect the desired power meter to 8-17


Error Messages

Messages

the HP-IB bus and try again. Make sure the power meter address matches the power meter address setting on the synthesizer. Check the HP-IB cable(s) for loose connections. 744

744,"Cannot nd power meter on HP-IB bus;(744)"
An automatic level correction was attempted but failed to nd a power meter on the HP-IB bus. Connect the desired power meter to the HP-IB bus and try again. Make sure the power meter address matches the power meter address setting on the signal generator. Check the HP-IB cable(s) for loose connections.

745

745,"Meter returns error msg;(745)"

While running the automatic level correction, the power meter returned +9.0000E+40 as the power reading. This number indicates an error within the power meter. 746

746,"Data measured is invalid or out of range;(746)"
While running the automatic level correction, the power meter returned an out of range power reading or the power meter returned a non-number as its power reading. Check that the power meter is reading an appropriate value by looking at it.

747

747,"Unable to receive msg from meter;(747)"

An automatic level correction was attempted but failed to power readings back from the power meter. Make sure the power meter address matches the power meter address setting on the synthesizer. Check the HP-IB cable(s) for loose connections. 748 749

748,"Erasing corrupted level correct table;(748)"

A level correction table was corrupt and was erased to x it.

749,"Frequency not within level correct data;(749)"
This message is a warning that the current CW frequency is not contained within the frequencies in the active level correction table. Therefore, the correction applied to the output will be 0 dB. This is a \permanent" error.

751

751,"Parameters cause points to be too large;(751)"
The level correction parameters cannot de ne a new table because they create too many level correction points.

8-18


Error Messages

Messages

752 753 754

754,"Total points reduced from that requested;(754)"
The number of points requested for an automatic level correction has been reduced to avoid duplicate frequencies.

0222,"Data out of range;Data set to minimum(752)" Loss data must be in the range of 040 to +40 dB. 0222,"Data out of range;Data set to maximum(753)" Loss data must be in the range of 040 to +40 dB.

755

755,"Invalid data in active table, not saved;(755)"
This error indicates that a level correction table save failed. Try selecting the same level correction table again. If this error message persists you will have to recreate the table.

756

756,"Factory level corr 1-20 GHz, 0 dB table;(756)"
Factory frequency level correction data for 1-20 GHz band, thru path is not valid. If you need to use this frequency range and you work with output levels that don't use the step attenuator, see the explanation for error number 4000.

757

757,"Bad attenuator setting parameter;(757)"
The attenuator range for looking up factory frequency level correction data, is 0 through 120 (resolution is 10).

758

0222,"Data out of range;Data set to minimum(758)"

Loss data for factory frequency level correction was less than minimum. 759

0222,"Data out of range;Data set to maximum(759)"
760,"Bad index into data table;(760)"

Loss data for factory frequency level correction was more than maximum. 760 A data lookup from a calibration table found that the index data is out of range. Try setting the same synthesizer function again. If this error message persists, run the instrument self test. 761

761,"Bad index into oset table;(761)"

8-19


Error Messages

Messages

A data lookup from the factory level correction oset table found that the index data is out of range. Try setting the same synthesizer function again. If this error message persists, run the instrument self test. 763

763,"Unable to write to EEPROM;(763)"
A calibration table was not loaded into EEPROM because the EEPROM was protected or the EEPROM load did not verify. Open up the synthesizer and switch the PG switch to 0.

764

764,"Unable to write to RAM;(764)"
A write to RAM failed to verify. Run the self test routine to check RAM for problems.

765

765,"Attempt to write to ROM;(765)"
There was an attempt to write calibration data to ROM. This should not occur, but if it does, try setting the same synthesizer function again. If this error message persists, run the instrument self test.

766

766,"Number of writes to EEPROM exceeds max;(766)"
The number of EEPROM writes has exceeded the maximum allowed. However, the data was written to the EEPROM anyway. This is only a warning; but, you should check to make sure your data was correctly stored in EEPROM.

770

770,"YTO cal data invalid;(770)"
The YIG oscillator factory calibration data checksum was incorrect. Select the CAL YIG OSC feature to perform a YIG oscillator calibration. If you do not re-calibrate, the synthesizer may be unable to attain lock at some frequencies.

771

771,"Invalid YIG DAC value in cal table.;(771)"
The YIG oscillator factory calibration data checksum was incorrect. Select the CAL YIG OSC feature to perform a YIG oscillator calibration. If you do not re-calibrate, the synthesizer may be unable to attain lock at some frequencies.

772

772,"YTO cal values for Up/Down search vary;(772)"
The YIG oscillator factory calibration data checksum was incorrect. Select the CAL YIG OSC feature to perform a YIG oscillator calibration.

8-20


Error Messages

Messages

If you do not re-calibrate, the synthesizer may be unable to attain lock at some frequencies. 774

774,"EEPROM protected, YTO cal aborted;(774)"
An automatic YIG oscillator calibration was not performed because the PG switch was protecting the EEPROM from \writes". Open up the synthesizer and switch the PG switch to 0.

775

775,"Low band yto cal failed;(775)"
The low band calibration failed and the data for the calibration was not saved. Re-try the YIG calibration and watch for the default cal points indicated when the DAC value shown for a given point says 'd t = ' instead of 'DAC = '.

776

776,"High band yto cal failed;(776)"
The low band calibration failed and the data for the calibration was not saved. Re-try the YIG calibration and watch for the default cal points indicated when the DAC value shown for a given point says 'd t = ' instead of 'DAC = '.

777

777,"Low band yto cal could not write EEPROM;(777)"
The data for the low band YIG calibration could not be written to EEPROM. Make sure the PG switch on the processor board was closed during the cal and re-try the YIG calibration.

778

778,"High band yto cal could not write EEPROM;(778)"
The data for the low band YIG calibration could not be written to EEPROM. Make sure the PG switch on the processor board was closed during the cal and re-try the calibration.

779

779,"Only 2GHz or 10GHz allowed for YTO Cal;(779)"
The start frequency speci ed for the CAL:YIG:FREQ:START command was not valid. Re-enter the start frequency and start the YIG calibration again.

785

785,"Cal Data not saved, PG switch is not 0;(785)"
The calibration data could not be saved because the PG switch on the microprocessor board was not closed. Close the PG switch and do the calibration again.

786

786,"Cal Data veri cation after EEPROM write;(786)"
8-21


Error Messages

Messages

The calibration data was not written correctly after the calibration. Close the PG switch and do the calibration again. 787

787,"FM cal and Pincho cal not initialized;(787)"
The YIG calibration has not been done for both YIG bands or else the pincho cal values have not been entered yet.

790

790,"Scan-mod 0.01-1 GHz gain tables bad;(790)"
Checksum was invalid for the AM gain tables. If you need to use this frequency range and scan AM modulation, see the explanation for error number 4000.

793

793,"Scan AM cal not valid, defaults used;(793)"

Scan AM level may be in error due to invalid calibration data. If you need to use scan AM modulation, see the explanation for error number 4000. 794

794,"Scan-mod 1-20 GHz gain tables bad;(794)"

Checksum was invalid for the AM scan gain tables. If you need to use this frequency range and scan AM modulation, see the explanation for error number 4000. 795

795,"Scan-mod 1-20 GHz linear 1 tables bad;(795)"

Checksum was invalid for the AM scan linear 1 tables. If you need to use this frequency range and scan AM modulation, see the explanation for error number 4000. 796

796,"Scan-mod 1-20 GHz linear 2 tables bad;(796)"

Checksum was invalid for the AM scan linear 2 tables. If you need to use this frequency range and scan AM modulation, see the explanation for error number 4000. 800

800,"Options not saved, PG switch is not 0;(800)"
The instrument option bit- elds were not saved to EEPROM. Open up the synthesizer, switch the PG switch to 0, and try setting the option bit- elds again.

801

801,"Serial num not saved, PG switch is not 0;(801)"
The instrument serial number was not saved to EEPROM. Open up the synthesizer, switch the PG switch to 0, and try setting the serial number again.

8-22


Error Messages

Messages

802

802,"Assembly Revs not saved, PG switch is not 0 (802)"
The protection switch on the CPU board is in an incorrect position to change EEPROM data.

803

803,"EEPROM not cleared, PG switch is not 0 (803)"
The protection switch on the CPU board is in an incorrect position to clear EEPROM data.

900

900,"PRI increased to t pulse width;(900)"
The current pulse width is too large for the current PRI. The PRI is increased to allow for the pulse width. This is a \permanent" error.

901

901,"Delay and width decreased to t max PRI;(901)"

The current pulse width plus the current pulse delay is too large because they are greater than the maximum PRI. The delay and/or the width were reduced to t. This is a \permanent" error. 940

940,"Oven is cold;(940)"

The high stability time base oven is cold. The oven must be allowed to warm up before proper instrument operation will occur. This is a \permanent" error. 944

944,"Reference synthesizer went out of lock;(944)"

The reference synthesizer went out of lock. The out of lock condition may have been due to an external time base being connected or disconnected. Enter a dierent RF frequency and then set the frequency back to the desired value to re-lock. This is a \permanent" error. 1101 1101,"Loop number is invalid.;(1101)" If you are using the direct hardware control service feature, you have entered an out of range value. If this error occurs while not using the direct hardware control service feature, low level hardware drivers could not set the requested synthesizer setting. Try setting the synthesizer again. If this error message persists, run the instrument self test. 1102 1102,"Start bit is negative.;(1102)" See the explanation for error number 1101. 1103 1103,"Length less than 0 or more than 32;(1103)" 8-23


Error Messages

Messages

See the explanation for error number 1101. 1104 1104,"Start bit is invalid for given loop.;(1104)" See the explanation for error number 1101. 1105 1105,"Length is invalid for given loop;(1105)" See the explanation for error number 1101. 1106 1106,"Data is too large for given length;(1106)" See the explanation for error number 1101. 1107 0222,"Data out of range;Bit eld number(1107)" See the explanation for error number 1101. 1108 0222,"Data out of range;Query port eld(1108)" If you are using the direct hardware control service feature, you have entered an out of range query address. If this error occurs while not using the direct hardware control service feature, low level hardware drivers could not complete a query. Try setting the synthesizer again. If this error message persists, run the instrument self test. 1109 1109,"Query Port mode;(1109)" See the explanation for error number 1108. 1462 1462,"Keybd processor reports status error;(1462)" A key press was not successful. Try pressing the same key again. If the error message persists, see the explanation for error number 4000. 1463 1463,"Keybd processor data lines incorrect;(1463)" A key press was not successful. Try pressing the same key again. If the error message persists, see the explanation for error number 4000. 1501 0302,"8673 command not recognized (1501)" While in HP 8673 emulation mode, an incoming command string was not recognized as being valid. Refer to the front-panel error que for the command sequence. 1511 0301,"8673 command K0 not emulated (1511)" The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus

8-24


Error Messages

Messages

discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1512

0301,"8673 command K1 not emulated (1512)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1513

0301,"8673 command Mn not emulated (1513)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1514

0301,"8673 command Xn not emulated (1514)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1515

0301,"8673 command Wn not emulated (1515)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1516

0301,"8673 command BS not emulated (1516)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1517

0301,"8673 command L2 not emulated (1517)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1518

0301,"8673 command 9 not emulated (1518)"

8-25


Error Messages

Messages

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1519

0301,"8673 command OC not emulated (1519)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1520

0301,"8673 command CT not emulated (1520)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1521

0301,"8673 command Nx not emulated (1521)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1522

0301,"8673 command SU not emulated (1522)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1523

0301,"8673 command SD not emulated (1523)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1524

0301,"8673 command Tx not emulated (1524)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information.

8-26


Error Messages

Messages

1525

0301,"8673 command TR not emulated (1525)"
The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information.

1526

0301,"8673 command RS not emulated (1526)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1527

0301,"8673 command SM not emulated (1527)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1528

0301,"8673 command 2 not emulated (1528)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1529

0301,"8673 command 3 not emulated (1529)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1530

0301,"8673 command A not emulated (1530)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1531

0301,"8673 command SV not emulated (1531)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus

8-27


Error Messages

Messages

discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1532

0301,"8673 command OL not emulated (1532)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1534

0301,"8673 command FA not emulated (1534)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1535

0301,"8673 command FB not emulated (1535)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1536

0301,"8673 command FS not emulated (1536)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1537

0301,"8673 command DF not emulated (1537)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1538

0301,"8673 command DW not emulated (1538)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1539

0301,"8673 command SF not emulated (1539)"

8-28


Error Messages

Messages

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1540

0301,"8673 command SP not emulated (1540)"

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information. 1541

0301,"8673 command IF not emulated (1541)"

1802 1802,"Low Battery Voltage;(1802)" The synthesizer battery voltage is low. This could cause loss of RAM data if signal generator power is turned o. Note: Calibration data will never be lost. 1803 1803,"RAM data lost at power on;(1803)" All RAM data was lost. This includes all front panel settings, save/recall registers, level corrections, and other user settable values. This error message can occur when the battery voltage is low, or options change in the synthesizer. Note: Calibration data will never be lost. 1804 1804,"Self test failure, run the self test;(1804)" The power-on self test detected an error or warning. See the explanation for error number 4000. 1805 1805,"Processor Board or IBUS test Failure;(1805)" The power-on self test detected an error or warning for the microprocessor board circuits or power supply monitors. See the explanation for error number 4000. 1806 1806,"ROM checksum test failure;(1806)" The synthesizer ROM check sum does not match the data in ROM. See the explanation for error number 4000. 8-29

The above command is recognized by the synthesizer in HP 8673 emulation mode, but it is not emulated. The command is thus discarded and no action is taken. Refer to Chapter 4 of the HP 8673 Compatibility Guide for more information.


Error Messages

Messages

2003

0222,"Data out of range;CW FREQ(2003)"
Select a CW frequency that is within range of the installed options. If other modules or options are installed that extend the CW frequency range of the synthesizer, this frequency range will be extended also.

2006

0222,"Data out of range;POWER LEVEL(2006)"

Select a power level within the following ranges: No attenuator options, 015 dBm to +30 dBm. Option 1E1, 0100 dBm to +30 dBm. 2012

0224,"Illegal parameter value;ALC SOURCE(2012)" 0222,"Data out of range;SPECIAL(2015)"

The requested ALC source is not available in the synthesizer. Use *OPT? to check which options are installed. 2015 2018 2024 Select a special function number that is available in the synthesizer.

0222,"Data out of range;FREQ MULTIPLIER INCR(2018)" 0222,"Data out of range;CW FREQ INCR(2024)"
Select a CW frequency modules or options are of the synthesizer, this Option 1E8 is installed resolution.

increment from 1 kHz to 19.99 GHz. If other installed that extend the CW frequency range frequency range will be extended also. If the limits will also change to allow for 1 Hertz

2030

0222,"Data out of range;DIRECT HW CONTROL(2030)"
Select synthesizer direct hardware control values within range. See the service manual for more details on this feature.

2033

0222,"Data out of range;POWER LEVEL INCR(2033)"

Select a power level within the following ranges: No attenuator options, 0.01 dBm to +45 dBm. Option 1E1, 0.01 dBm to +130 dBm. 2036

0222,"Data out of range;EXT METER LEVEL(2036)"

Select an external power meter reading within the following ranges: No attenuator options, 015 dBm to +30 dBm. Option 1E1, 0100 dBm to +30 dBm. 2042 8-30

0222,"Data out of range;DIAG:IBUS:DIR(2042)"


Error Messages

Messages

Correct the HP-IB command DIAG:IBUS:DIR or DIAG:IBUS:DIR? so that its parameters are within their appropriate ranges. See the service manual for more details on this HP-IB only feature. 2045 2048 2051

0224,"Illegal parameter value; *EMC (2045)" 0222,"Data out of range;SYST:KEY(2048)"

This command requires a 1 or 0 as an argument. Select a key code available on the synthesizer's front panel.

0161,"Invalid block data;SYST:SET bad size(2051)"

The \learn string" sent to the synthesizer is corrupt (incorrect number of bytes). Check that the HP-IB controller is sending the string correctly. In addition, insure that the controller loaded the learn string correctly in the rst place. Note: The *LRN? query always returns the same length string regardless of the state of the synthesizer; but, the *LRN? response can change if the rmware version changes. 2054

0222,"Data out of range;CAL:ALC:CURV(2054)"

Correct the HP-IB command CAL:ALC:CURV so that its parameters are within the following ranges: 1st parameter: 0.0 to 4.0 2nd parameter: 02.0 to 2.0 3rd-6th parameter: 01.0 to 1.0 7th parameter: 0.0 to 25.0 2057 2060 2066 2075 2078

0222,"Data out of range;HPIB ADDRESS(2057)" 0222,"Data out of range;SAVE(2060)"
Select a save state register number from 0 to 9.

Select an HP-IB address for the synthesizer from 0 to 30.

0222,"Data out of range;RECALL(2066)"

Select a recall state register number from 0 to 9.

0222,"Data out of range;LO FREQ(2075)"

Select an LO frequency from 300 MHz to 359.5 MHz.

0222,"Data out of range;OFFSET FREQ(2078)"

Select an oset frequency from 5 MHz to 40 MHz.

8-31


Error Messages

Messages

2081

0222,"Data out of range;DIAG:FREQ:CYCL(2081)"
Correct the HP-IB command DIAG:FREQ:CYCL so that its parameters are within range. See the service manual for more details on this feature.

2087

0222,"Data out of range;YIG OSC CAL FREQ(2087)"

Correct the HP-IB command CAL:YIG:FREQ:STARt so that its parameter is 2 GHz or 10 GHz. 2090

0222,"Data out of range;CAL:YIG(2090)"

Correct the HP-IB command CAL:YIG[:DATA] so that all of its parameters are from 0 to 65535. 2099 2102

0222,"Data out of range;FREQ MULTIPLIER(2099)"
Select a frequency multiplier from 1 to 100.

0222,"Data out of range;EXT METER INCR(2102)"
Select an external power meter reading increment within the following ranges: No attenuator options, 0.01 dBm to +45 dBm. Option 1E1, 0.01 dBm to +130 dBm.

2105

0222,"Data out of range;CAL:ALC:CURV:FREQ(2105)"

Correct the HP-IB command CAL:ALC:CURVe:FREQuency:STARt so that its parameter is from 10 MHz to 40 GHz. See the service manual for more details on this feature. 2111 2114

0224,"Illegal parameter value; *PSC (2111)"

This command requires a 1 or 0 as an argument.

0224,"Illegal parameter value;ATTEN LOCK(2114)"
The HP-IB command \POWer:ATTenuation:AUTO OFF" can only be used if Option 1E1 is installed.

2165

0224,"Illegal parameter value;REMOTE LANGUAGE(2165)" 0222,"Data out of range;CAL:FLAT(2177)"

Select an HP-IB remote language which is available in the synthesizer. \SCPI" is the default but others are available as options. 2177 Loss data for factory frequency level correction was out of range. 8-32


Error Messages

Messages

2189 2192 2198

0224,"Illegal parameter value;DISPLAY STATE(2189)"
The display state could not be set to the state requested. This command requires a boolean argument.

0224,"Illegal parameter value; LEV CORR STATE (2192)" 0224,"Illegal parameter value;EXT REF(2198)"
The reference oscillator could not be set to INTernal because an internal oscillator does not exist or the reference oscillator could not be set to EXTernal because an external oscillator does not exist.

2210 2216

0222,"Data out of range; ASSembly REVision (2210)"
The value must remain between 0 and +254.

0222,"Data out of range;NODE MEASURE(2216)"
Select a meter node number within range. See the service manual for more details on this feature.

2219

0222,"Data out of range;OPTION WRITE(2219)"

Select an option bit- eld number within range. See the service manual for more details on this feature. 2237

0222,"Data out of range;OFFSET FREQ INCR(2237)"

Select an oset frequency increment from 1 kHz to 35 MHz. If the Option 1E8 is installed the lower limit will change to 1Hz to allow for 1 Hertz resolution. 2240 2243

0222,"Data out of range;LO FREQ INCR(2240)"

0222,"Data out of range;DIRECT HW CONTROL INC(2243)"
Select synthesizer direct hardware control increment values within range. See the service manual for more details on this feature.

Select an LO frequency increment from 1 kHz to 359.5 MHz.

2249

0222,"Data out of range;DIAG:ABUS?(2249)"

Correct the HP-IB query DIAG:ABUS? so that its parameter is within their appropriate range. See the service manual for more details on this HP-IB only feature. 2276

0222,"Data out of range;CORR:FLAT(2276)"

8-33


Error Messages

Messages

2277 2277,"CORR:FLAT cannot query empty table;(2277)" The selected level correction table data cannot be queried because it is invalid or it does not exist. Check that MEMory:TABLe:SELect is set to a level correction table that has data. 2291 0224,"Illegal parameter value;SERIAL NUM(2291)" Correct the HP-IB command SYSTem:SNUMber so that its string parameter is 10 characters or less. 2292 0151,"Invalid string data;SERIAL NUM bad char(2292)" A serial number can only contain characters from ASCII 32 (space) through ASCII 126 (~). However, ASCII 44 (,) and ASCII 59 (;) cannot be used. 2294 0222,"Data out of range;POW METER ADDRESS(2294)" Select a power meter address for automatic level correction from 0 to 30. Secondary addresses may be allowed in future rmware revisions. 2300 0222,"Data out of range;CAL:FLAT:FREQ:START(2300)" Correct the HP-IB command CAL:FLATness:FREQuency:STARt so that its parameter is from 10 MHz to 40 GHz. See the service manual for more details on this feature. 2303 0222,"Data out of range;CAL:FLAT:ATT(2303)" Correct the HP-IB command CAL:FLATness:ATTenuation so that its parameter is from 0 dB to 120 dB with a resolution of 10 dB. See the service manual for more details on this feature. 2444 0222,"Data out of range;LEVEL CORR START FREQ(2444)" Select an automatic level correction start frequency from 1GHz to 20GHz. If other modules or options are installed that extend the CW frequency range of the synthesizer, this frequency range will be extended also. 2447 0222,"Data out of range;LEVEL CORR STOP FREQ(2447)" 8-34

Correct the HP-IB command CORRection:FLATness[:DATA] so that all of its frequency parameters are from 1 GHz to 20 GHz and all of its loss parameters are from 040 dB to +40 dB. If other modules or options are installed that extend the frequency range of the synthesizer, this frequency range will be extended also.


Error Messages

Messages

Select an automatic level correction start frequency from 1GHz to 20GHz. If other modules or options are installed that extend the CW frequency range of the synthesizer, this frequency range will be extended also. 2457 2457,"RF on before running level correct;(2457)" The RF must be turned on before running an automatic level correction. Turn RF on and try running the automatic level correction again. 2462 0222,"Data out of range;LEVEL CORR START INC(2462)" Select an automatic level correction start frequency increment from 1 kHz to 19.99 GHz. If other modules or options are installed that extend the CW frequency range of the synthesizer, this frequency range will be extended also. If Option 1E8 is installed the limits will also change to allow for 1 Hertz resolution. 2465 0222,"Data out of range;LEVEL CORR STOP INC(2465)" Select an automatic level correction stop frequency increment from 1 kHz to 19.99 GHz. If other modules or options are installed that extend the CW frequency range of the synthesizer, this frequency range will be extended also. If the Option 1E8 is installed the limits will also change to allow for 1 Hertz resolution. 2471 0222,"Data out of range;HPIB ADDRESS INCR(2471)" Select an HP-IB address increment from 1 to 29. 2474 0222,"Data out of range;YIG OSC CAL FREQ INC(2474)" Correct the HP-IB command CAL:YIG:FREQuency:STARt:STEP so that its parameter is from 1 GHz to 10 GHz. 2522 0222,"Data out of range;POW METER ADDRESS INC(2522)" Select a power meter address increment from 1 to 29. 2525 0222,"Data out of range;NODE MEASURE INC(2525)" Select a meter node increment within range. 2531 0222,"Data out of range;LEVEL CORR POINTS(2531)" Select automatic level correction number of points from 2 to 401. 2534 0222,"Data out of range;LEVEL CORR POINTS INC(2534)" 8-35


Error Messages

Messages

Select automatic level correction number of points increment from 1 to 401. 2537 2540 2570 2588 2591

0222,"Data out of range;LEVEL CORRECT SAVE(2537)" 0222,"Data out of range;LEVEL CORR SELECT(2540)"
Select an automatic level correction register from 1 to 4. Select an automatic level correction register from 1 to 4.

0224,"Illegal parameter value;POWERMETER TYPE(2570)" 0224,"Illegal parameter value;MEM:TABL:SEL(2588)" 0222,"Data out of range;MEM:TABL:FREQ(2591)"
The selected power type is not supported by the synthesizer. Select an automatic level correction register from 1 to 4.

Correct the HP-IB command MEMory:TABLe:FREQuency so that all of its parameters are from 1 GHz to 20 GHz. If other modules or options are installed that extend the frequency range of the synthesizer, this frequency range will be extended also. 2592 2592,"MEM:TABL:FREQ cannot query empty table;(2592)" The selected level correction table data cannot be queried because it is invalid or it does not exist. Check that MEMory:TABLe:SELect is set to a level correction table that has data. 2597 0222,"Data out of range;MEM:TABL:LOSS(2597)" Correct the HP-IB command MEMory:TABLe:LOSS[:MAGnitude] so that all of its parameters are from 040 dB to +40 dB. 2598 2598,"MEM:TABL:LOSS cannot query empty table;(2598)" The selected level correction table data cannot be queried because it is invalid or it does not exist. Check that MEMory:TABLe:SELect is set to a level correction table that has data. 2702 0222,"Data out of range;ADD OPTION(2702)" Select an option bit number within range. See the service manual for more details on this feature. 2705 0222,"Data out of range;DELETE OPTION(2705)"

8-36


Error Messages

Messages

Select an option bit number within range. See the service manual for more details on this feature. 4000

0330,"Self test failed;(4000)"

Run the instrument self test a couple times, checking the error queue each time the self test is run. If the error message persists, use the ERASE MEMORY feature, press the preset key and cycle the power; try the self test again. If the error message persists, an instrument failure may have occurred and servicing may be required. If the synthesizer is functioning to your satisfaction, you may wish to ignore the error message. 4001 4002 4003 4004 4005 4006 4007 4008 4009 4010

0330,"Self test failed;(4001)" 0330,"Self test failed;(4002)" 0330,"Self test failed;(4003)" 0330,"Self test failed;(4004)" 0330,"Self test failed;(4005)" 0330,"Self test failed;(4006)" 0330,"Self test failed;(4007)" 0330,"Self test failed;(4008)"

See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000.

0330,"Self test failed;(4009)" 0330,"Self test failed;(4010)"

See the explanation for error number 4000. See the explanation for error number 4000. 8-37


Error Messages

Messages

4011 4012 4013 4014 4015 4016 4017 4018 4019 4020 4021 4022 4023 4024 8-38

0330,"Self test failed;(4011)" 0330,"Self test failed;(4012)" 0330,"Self test failed;(4013)" 0330,"Self test failed;(4014)" 0330,"Self test failed;(4015)" 0330,"Self test failed;(4016)" 0330,"Self test failed;(4017)" 0330,"Self test failed;(4018)" 0330,"Self test failed;(4019)" 0330,"Self test failed;(4020)" 0330,"Self test failed;(4021)" 0330,"Self test failed;(4022)"
See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000.

0330,"Self test failed;(4023)" 0330,"Self test failed;(4024)"

See the explanation for error number 4000.


Error Messages

Messages

See the explanation for error number 4000. 4025 4026 4027 4028 4029 4030 4031 4032 4033 4034 4035 4036 4037

0330,"Self test failed;(4025)" 0330,"Self test failed;(4026)" 0330,"Self test failed;(4027)" 0330,"Self test failed;(4028)" 0330,"Self test failed;(4029)" 0330,"Self test failed;(4030)" 0330,"Self test failed;(4031)" 0330,"Self test failed;(4032)" 0330,"Self test failed;(4033)" 0330,"Self test failed;(4034)"

See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000.

0330,"Self test failed;(4035)"

See the explanation for error number 4000.

0330,"Self test failed;(4036)" 0330,"Self test failed;(4037)"

See the explanation for error number 4000. See the explanation for error number 4000.

8-39


Error Messages

4038 4039 4040 4041 4042 4043 4044 4045 9000

0330,"Self test failed;(4038)" 0330,"Self test failed;(4039)" 0330,"Self test failed;(4040)" 0330,"Self test failed;(4041)" 0330,"Self test failed;(4042)" 0330,"Self test failed;(4043)" 0330,"Self test failed;(4044)" 0330,"Self test failed;(4045)" 0330,"Self test failed;(9000)"
See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000. See the explanation for error number 4000.

95009999 Many dierent error messages can occur here. Errors with a number from 9500 to 9999 are MMS MSIB errors. These errors occur when the synthesizer has some problem with communication over the MSIB bus. See the explanation for error number 4000.

8-40


9

Legal and Regulatory Information


Legal and Regulatory Information

This chapter contains information pertaining to safety and the warranty. The Declaration of Conformity is located at the end of this chapter.

9-2


Safety Symbols

WARNING

The WARNING sign denotes a hazard. It calls attention to a procedure which, if not correctly performed or adhered to, could result in personal injury or loss of life. Do not proceed beyond a WARNING note until the indicated conditions are fully understood and met.
The CAUTION sign denotes a hazard. It calls attention to a procedure that, if not correctly performed or adhered to, could result in damage to or destruction of part of the product. Do not proceed beyond a CAUTION note until the indicated conditions are fully understood and met.

L M K

Instruction documentation symbol: The product is marked with this symbol when it is necessary for the user to refer to the instruction in the documentation. Indicates hazardous voltages. Indicates earth (ground) terminal.

C A U TI ON

WARNING

No operator serviceable parts inside. Refer servicing to quali ed personnel. To prevent electrical shock do not remove covers.

9-3


Legal and Regulatory Information

Safety Symbols

Miscellaneous Symbols

CE

The CE symbol is a registered trademark of the European Community (if accompanied by a year, it is when the design was proven). ISM 1-A This is a symbol of an Industrial Scienti c and Medical Group 1 Class A product. CSA The CSA symbol is a registered trademark of the Canadian Standards Association.

9-4


Safety Considerations

This product and related documentation must be reviewed for familiarization with safety markings and instructions before operation. This product is a Safety Class I system (provided with a protective earth terminal).

Before Applying Power

CAUTION C AU T I O N

Verify that the product is set to match the available line voltage and the correct fuses are installed. This product is designed for use in Installation Category II and Pollution Degree 2 per IEC 1010 and 664 respectively. VENTILATION REQUIREMENTS: When installing the product in a cabinet, the convection into and out of the product must not be restricted. The ambient temperature (outside the cabinet) must be less than the maximum operating temperature of the product by 4 C for every 100 watts dissipated in the cabinet. If the total power dissipated in the cabinet is greater than 800 watts, then forced convection must be used.

Safety Earth Ground

An uninterruptable safety earth ground must be provided from the main power source to the product input wiring terminals through the power cable or supplied power cable set.

9-5


Legal and Regulatory Information

Safety Considerations

WARNING

This is a Safety Class I product provided with a protective earthing ground incorporated in the power cord. The mains plug shall only be inserted in a socket outlet provided with a protective earth contact. Any interruption of the protective conductor inside or outside of the product is likely to make the product dangerous. Intentional interruption is prohibited. Whenever it is likely that the protection has been impaired, the system must be made inoperative and be secured against any unintended operation. Servicing instructions are for use by service-trained personnel only. To avoid dangerous electric shock, do not perform any servicing unless quali ed to do so. Adjustments described in the manual are performed with power supplied to the system's instruments while protective covers are removed. Energy available at many points may, if contacted, result in personal injury. Capacitors inside the system's instruments might still be charged even if the system has been disconnected from its source of supply. For continued protection against re hazard, replace the line fuses only with 250 V fuses of the same current rating and type (for example, normal blow, time delay, etc.). The use of other fuses or materials is prohibited.

Cleaning the Product

Clean the synthesizer cabinet using a damp cloth only.

9-6


Certi cation

Hewlett-Packard Company certi es that this product met its published speci cations at the time of shipment from the factory. Hewlett-Packard further certi es that its calibration measurements are traceable to the United States National Institute of Standards and Technology, (NIST), to the extent allowed by the Institute's calibration facility, and to the calibration facilities of other International Standards Organization members.

9-7


Warranty

This Hewlett-Packard instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Hewlett-Packard Company will, at its option, either repair or replace products which prove to be defective. For warranty service or repair, this product must be returned to a service facility designated by HP. Buyer shall prepay shipping charges to HP and HP shall pay shipping charges to return the product to Buyer. However, Buyer shall pay all shipping charges, duties, and taxes for products returned to HP from another country. HP warrants that its software and rmware designated by HP for use with an instrument will execute its programming instructions when properly installed on that instrument. HP does not warrant that the operation of the instrument, or software, or rmware will be uninterrupted or error free.

Limitation of Warranty

The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modi cation or misuse, operation outside of the environmental speci cations for the product, or improper site preparation or maintenance.

NO OTHER WARRANTY IS EXPRESSED OR IMPLIED. HP SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE.

9-8


Legal and Regulatory Information

Warranty

Exclusive Remedies

THE REMEDIES PROVIDED HEREIN ARE BUYER'S SOLE EXCLUSIVE REMEDIES. HP SHALL NOT BE LIABLE FOR INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL WHETHER BASED ON CONTRACT, TORT, OR ANY OTHER THEORY.

AND ANY DIRECT, DAMAGES, LEGAL

9-9


Assistance

Product maintenance agreements and other customer assistance agreements are available for Hewlett-Packard products. For any assistance, contact your nearest Hewlett-Packard Sales and Service Oce shown in Table 9-1.

9-10


Legal and Regulatory Information

Assistance

9-12


Legal and Regulatory Information

Notice for Germany: Noise Declaration LpA < 70 dB am Arbeitsplatz (operator position) normaler Betrieb (normal position) nach DIN 45635 T. 19 (per ISO 7779)

9-13


Legal and Regulatory Information


Index


Index

0 1 3 4 A

0.5 V/GHz OUT connector, 5-7 10 MHz IN connector, 5-7 10 MHz OUT connector, 5-7 1 Hz frequency resolution - option 1E8, 4-20 3.5 mm RF output connector - option 1E9, 4-20 488.1 compatibility codes, 5-9 active parameter, 6-40 add output step attenuator - option 1E1, 4-20 address, 6-11 setting the HP-IB, 2-12 advantages attenuator hold function, 7-4 ALC IN connector, v, 5-7 ALC unleveled condition, 6-27 annunciators, v, 5-3, 5-11 uorescent display, 5-11 LED, 5-11 ANSI Standard MC1.1, 5-10 application attenuator hold function, 7-4 erase memory function, 7-8 arrow keys modifying data with, 2-6 ATTEN HOLD ON/OFF, 7-3 attenuator hold function, 7-3 automatic level control keys, v

Index-2


B

BACK SPACE, v, 6-12 +5 key, v, 6-9 (5 key, v, 6-5 )5 key, v, 6-3 *5 key, v, 6-7 book introduction, vi
4 4 4 4

C

carrier keys, v CLEAR DISPLAY, 7-6 coaxial connectors, 5-6{8 component-level information packet - option OBV, 4-23 component-level information packet/service documentation - option OBX, 4-23 connector care, 5-7 connectors, 5-6 coaxial, 5-6{8 damage levels, 5-6{8 front panel, 5-6 HP-IB, 5-8{10 impedances, 5-6{8 rear panel, 5-6 cursor, v cursor, double, 2-3, 6-3, 6-5 CW signal generating, 3-3 damage electrical, 1-8 mechanical, 1-3, 1-8 data entering with numeric keypad, 2-3 modifying with arrow keys, 2-6 modifying with knob, 2-4 data entry keys, v data entry problems, 2-13{14 default state display, 6-13 delete operating documentation - option OB0, 4-22 DIODE (ALC) annunciator, 5-3 diode detector leveling advantages, 6-18 applications, 6-17 disadvantages, 6-18 using, 3-7 diode detector square law region, 6-17, 6-18 disadvantages attenuator hold function, 7-5 display, 5-11

D

Index-3


default state, 6-13 documentation options, 4-22 double cursor, 6-3, 6-5

E

earth ground for safety, 1-6, 9-5 electrical options, 4-20 ENTER, 6-13 ENTRY OFF function, 2-13, 6-15 ERASE MEMORY, 7-8 erase memory function, 7-8 error message action required, 8-3 detailed description, 8-3 hardware failures, 8-4 manual error number, 8-3 SCPI error message, 8-3 SCPI error number, 8-3 error message format, 8-3 error messages user-oriented, v error messages list, 8-5{40 error queue reading, 2-10 errors permanent, 8-2 EXT DIODE, 6-16 EXT METER, 6-20 extra operating documentation - option OB1, 4-22 EXT REF annunciator, 5-3 factory calibration data, 7-8 failures self test, 1-11 fan checking, 1-10 FREQ, 6-25 fuse checking, 1-9

F

Index-4


H

HP-IB address setting, 2-12 HP-IB connector, 5-8{10 HP-IB extender instruments, 5-9 HP-IB interface capabilities, 5-9 HP-IB restrictions, 5-8 IEEE Standard 488 (1978), 5-10 increment value, 2-6, 6-57 installing the synthesizer, 1-3 instrument state recall, 6-47 instrument state save, 6-49 INT (ALC) annunciator, 5-3 interface capabilities HP-IB, 5-9 internal leveling advantages, 6-28 disadvantages, 6-28 in this book, vi INT LEVEL, 6-27 introduction book, vi keypad numeric, 6-40 knob, v, 5-13 modifying data with, 2-4 knob resolution decreasing, 2-4 increasing, 2-4 legal and regulatory information, 9-2 level correct data using previously stored, 3-20 level correction register, 6-45, 6-50 level correct routine, 6-31 using, 3-14 leveling using diode detector, 3-7 using power meter, 3-11 line voltage selector, 1-5 LLO, 6-30 LOCAL, 6-30 LOCAL key, 2-13 local lockout, 6-30 LSN annunciator, 5-4 LVL CR, 6-31

I

K

L

Index-5


LVL CR annunciator, 5-4 LVL CR ON/OFF, 6-35

M

mating connectors RF OUPUT, v mechanical damage, 1-3 mechanical options, 4-21 millimeter signals generating, 3-4 MSG, v, 6-37 MSG annunciator, 1-11, 3-23, 5-4 MTR (ALC) annunciator, 5-3 MULT annunciator, 5-3 multiplier, 6-38 multiplier value, 6-38 multiplier values for HP source modules, 3-4 number of points, 6-31 numeric keypad, 6-40 entering data with, 2-3 option 1CM - rack mount kit, 4-21 option 1CP - rack mount and handle kit, 4-21 option 1CR - rack slide kit, 4-21 option 1E1 - add output step attenuator, 4-20 option 1E5 - add high stability timebase, 4-20 option 1E8 - 1 Hz frequency resolution, 4-20 option 1E9 - 3.5 mm RF output connector, 4-20 option OB0 - delete operating documentation, 4-22 option OB1 - extra operating documentation, 4-22 option OBV - component-level information packet, 4-23 option OBW - service documentation, 4-23 option OBX - component-level information packet/service documentation, 4-23 options, 4-17 documentation, 4-22 electrical, 4-20 mechanical, 4-21 warranty, 4-21 option W30 - two years return-to-HP service, 4-22 option W32 - three year return-to-HP calibration service, 4-22 option W34 - three year Mil-std calibration service, 4-22

N O

Index-6


P

parameter active, 6-40 unitless, 6-13 permanent errors, 8-2 points number of, 6-31 power cable con gurations, 5-16 power cables, 5-16 POWER LEVEL, 6-41 power meter address function, 7-10 power meter leveling advantages, 6-23 applications, 6-22 disadvantages, 6-23 equipment limitations, 6-22 using, 3-11 power meter select function, 7-12 POWER switch, v, 5-15 power-up problems, 1-8{11 PRESET, 6-43 problems data entry, 2-13{14 power-up, 1-8{11 programming, 2-15 RF output, 3-23 programming problems, 2-15 PWR MTR ADDRS, 7-10 PWR MTR SELECT, 7-12 rack mount and handle kit - option 1CP, 4-21 rack mount kit - option 1CM, 4-21 rack slide kit - option 1CR, 4-21 RCL CR, 6-45 reading the error queue, 2-10 RECALL, v, 6-46 recalling synthesizer states, 2-8 remote language, 7-14 restrictions HP-IB, 5-8 RF ON/OFF, 6-48 RF OUTPUT connector, v, 5-7 RF output problems, 3-23 RMT annunciator, 2-13, 5-4

R

Index-7


S

sales and service oce table, 9-10 SA VE, v, 6-49 SA CR, 6-50 VE saving synthesizer states, 2-8 self test running, 1-7 SELF TEST, 7-15 self test failures, 1-11 self test function, 7-15 serial number, 7-18 service documentation - option OBW, 4-23 service-related special functions, 7-17 SHIFT, v, 6-52 SHOW INFO, 7-18 show information function, 7-18 software version number, 7-18 source modules multiplier values for, 3-4 SPCL, v, 6-53 SPCL annunciator, 5-3 SPCL OFF, 6-54 SPCL ON, 6-56 special functions, v, 6-53, 7-2 service-related, 7-17 speci cations, 4-3, 4-17 square law region diode detector, 6-17, 6-18 SRQ annunciator, 5-4 standby, 5-15 start frequency, 6-31 STEP SIZE, 6-57 STEP SIZE , v STEP SIZE key, 2-6 stop frequency, 6-31 supplemental characteristics, 4-3, 4-9 synthesizer installing, 1-3 synthesizer at a glance, iv, v synthesizer states saving and recalling, 2-8

Index-8


T

terminator keys, v, 2-3 three year Mil-std calibration service - option W34, 4-22 three year return-to-HP calibration service - option W32, 4-22 TLK annunciator, 5-4 to generate a CW signal, 3-3 to generate millimeter signals, 3-4 to use external diode detector leveling, 3-7 to use external power meter leveling, 3-11 to use previously stored level correct data, 3-20 to use the level correct routine, 3-14 turn-on sequence, 1-6 two years return-to-HP service - option W30, 4-22 unitless parameters, 6-13 units terminator keys, 2-3 unleveled condition ALC, 6-27 UNLOCK annunciator, 5-4 UNLVL annunciator, 5-4 veri cation procedure, 1-7 warranty options, 4-21

U

V W

Index-9