Документ взят из кэша поисковой машины. Адрес оригинального документа : http://jet.sao.ru/hq/lsfvo/devices/scorpio-2/ifp_eng.html
Дата изменения: Unknown
Дата индексирования: Sun Apr 10 04:24:29 2016
Кодировка: Windows-1251

Поисковые слова: equinox
SCORPIO-2 project: Fabry-Perot

РУССКАЯ версия
Main page
Description
Parameters
Filters
Grisms
Modes:
Direct images
Long-slit
Spectropolarimetry
Scanning Fabry-Perot
IFU
Multislit

MANUALS
PUBLICATIONS


Back to SAO web page

Last update: 21 Nov 2015
This page is designed by A.V. Moiseev
moisavsao.ru

The SCORPIO-2 project:
Spectral Camera with Optical Reducer for Photometric and Interferometric Observations - 2

Scanning Fabry-Perot interferometers (FPI)

Observations with the FPI consist of successive acquisition of a few dozen images of interference rings from the object (or from a calibration lamp) when changing the optical path between the parallel plates. The radius of the rings is a function of wavelength and gap between the plates of the interferometer. The complete set of such images which fills the free spectral range of the interferometer is called 'a scanning cycle'. After data reduction images are collected in a data cube. Control of the scanning (change of the gap between flat reflecting plates) is managed by the controller CS100 IC Optical Systems Ltd., connected with a computer. The description of observations and data reduction is presented in the papers:
  • Moiseev A.V. 2002, "Reduction of the CCD-observations with scanning interferometer Fabry-Perot", Bull. SAO, 54, 77 (English PDF, Русский PS)
  • Moiseev A.V., Egorov O.V., "Reduction of the CCD-observations obtained with the Fabry-Perot scanning interferometer. II Additional procedures", 2008, Astrophy. Bull., 63, 181 (English PDF, Русский PDF)

SCORPIO-2 has a set of bandpass filters which cut the desired narrow spectral range around the emission lines H-alpha, [OIII]5007, [NII]6548,6583, [SII]6717,6731. You can use this IDL-programm for chose an optimal filter for desired systemic velocities of target galaxies.

The piezoelectric interferometers were mad by IC Optical Systems Ltd. (early name - Queensgate).

The main parameters if FPIs for observations with SCORPIO-2 in the Halpha spectral range.
  IFP186 IFP501 IFP751
Order 186501751
Interfringe 35 A
(~1600) km/s)
13 A
(~600 km/s)
8.7 A
(~390 km/s)
Sp. resolution 1.7 A 0.8A0.4A
Finesse 21 1620
Number of spectral channels40 36 40


The photo shows IFP751 mounted into SCORPIO-2 during tuning procedures Sorry, this page is under construction