Instrument Science Report WFC3 2009-03

WFC3 TV3 Testing: UVIS-1' Crosstalk
S.Baggett May 6, 2009

ABSTRACT This report summarizes the behavior of the crosstalk in the Wide Field Camera 3 UVIS-1' flight detector based on thermal-vacuum data taken with the integrated instrument. A nonlinear type of crosstalk seen in previous tests, where a target in one quadrant gives rise to a significant mirror image in each of the other three quadrants, is confirmed as gone (<0.1 e- ), thanks to a hardware fix to the CCD electronics box and a new timing pattern. A small amount of crosstalk remains, in which point sources, extended targets, hot pixels, and cosmic rays generate a low-level mirror image in the quadrant adjoining the target quadrant, on the same chip. This residual crosstalk is linear, negative, and appears at the level of ~10-4 to 10-5 in unbinned, nominal gain (1.5 e-/DN), full-frame 4-amp readouts; the higher level occurs for targets imaged in quadrants A or C and the lower level for targets placed in quadrants B or D. In 3x3 binned 4-amp readouts, the crosstalk from point source targets placed in D is about 20% higher than that seen in unbinned readout. However, the crosstalk in binned frames is negligible for targets placed in quadrants A and B and only a small amount of crosstalk (~5x10-5) occurs due targets placed in C quadrant.

Introduction
Images taken with the integrated WFC3 instrument in early 2004 under ambient conditions and in late 2004 under thermal vacuum conditions revealed the presence of crosstalk (CT): a light source in one quadrant of the field of view generating low level electronic mirror images in one or more other quadrants. This phenomenon was not completely unexpected, as it frequently occurs whenever two or more channels are read out

Copyright© 2007 The Association of Universities for Research in Astronomy, Inc. All Rights Reserved.


Instrument Science Report WFC3 2009-03 simultaneously (Janesick, 2001). The CT in WFC3 as characterized during those early tests has been discussed in previous reports (Baggett et al., 2004; Baggett et al., 2005) and can be summarized briefly as follows. Two types of crosstalk were initially identified: 1) sources at any exposure level in any quadrant caused highly non-linear CT in the three other quadrants and 2) pixels in any quadrant caused low-level CT in the adjoining quadrant, i.e., on the same CCD chip. The first type of CT produced mirror-image features at relatively low levels (~15 e- at gain 1.5) regardless of the source exposure level. Changing bias levels and gain settings caused changes to the sign, and to some extent, the magnitude of the CT, implying that the crosstalk-inducing signal was somewhere within or following the gain stage. In addition, the CT varied as a function of the binning and number of amps used during the image readout: full-frame images that were binned (either 2x2 or 3x3) and read out with only 2 amps exhibited no crosstalk at all; full-frame images that were binned and read out with 4 amps exhibited only the second type of CT (same chip, i.e., on amp adjacent to the target amp). The second type of CT was initially attributed to saturated pixels only but has since been found to be present at a low level at all exposure levels and is the subject of this report. Following the observations of the two types of crosstalk, modelling of the electronics and testing with the non-flight CCD electronics box (CEB) at the GSFC Detector Characterization Lab revealed that there was a common-mode susceptibility in the CEB and that adding a balancing capacitor would partially reduce the CT levels. In addition, the timing pattern in use at the time was found to be a contributing factor: the digitization of pixels by the A/D converter could interfere with the analog sampling of the subsequent pixels, thereby injecting CT. This hypothesis was consistent with the lack of CT in 2-amp binned images (Baggett et al., 2004) and confirmed via images taken with the A/D conversion speed doubled: the CT could be alleviated by compressing and shifting the sample time away from the A/D timing. As a consequence, a new timing pattern was developed and is now in use. With these hardware and timing pattern changes in place, the UVIS-1' detector package integrated into WFC3 underwent further thermal vacuum testing in the spring of 2008 (thermal vacuum ground test 3, TV3). This report summarizes the behavior of the residual WFC3 UVIS crosstalk in light of the modifications.

Data
The results discussed here are based on a variety of data taken during the last thermal vacuum test. First, two programs specifically aimed at investigating crosstalk were executed. The first test was performed on Feb 28, 2008, with the detector at the off-nominal temperature of about -50C. A set of 3 images were taken in each of the 4 quadrants: a HeNe point source with peak level ~30K e- , an extended point source using the 200 VISIR target with peak level ~50K e- , and finally, another HeNe point source that was

2


Instrument Science Report WFC3 2009-03 deliberately saturated to ~100x the level of the first image. The second program was run on March 11, 2008, with the detector at the nominal operating temperature of ~-83 C. In this case, a set of 6 images were acquired on each quadrant with the extended 200 VISIR point source. The first four images in each quadrant were taken at the nominal focus setting, tuned to achieve peak levels of ~9K, 23K, 37K, and 50K e- . The last two images of each set were taken with the optical stimulus at -15mm defocus setting in order to spread the light out even farther; peak exposure levels were ~8K e- and ~45K e- . In all cases, images were full-frame, unbinned, four-amp read outs taken at the nominal gain setting (1.5 e-/DN); an example of one of the full-frame images is shown in Figure 1. All the images from the two crosstalk tests are tabulated in Appendix A; a mosaic of the targets and the corresponding crosstalk are include in Appendix B (Figure 4 and Figure 5). Finally, data from two other proposals, while not obtained specifically for the purpose of investigating crosstalk, have been included in the analysis as well: the long exposuretime, full-frame UVIS dark frames (iu01* series) and the UVIS "glint" frames (iu26* series). The former are used here to evaluate the behavior of CT due to hot pixels and cosmic rays while the latter, taken in 3x3 binned mode as part of a program to search for residual scattered light, are used to quantify CT behavior in binned frames. All the images from these two tests are tabulated in Appendix A as well; mosaics of some of the binned frames are in Appendix B (Figure 6). Figure 1: At left, a typical crosstalk test image, shown with a hard inverted greyscale stretch. In this case, the extended 200 point source target was placed in quadrant C, using -15mm defocus, and the crosstalk appears in quadrant D. The faint large ring in C offset from the primary target is an optical ghost due to reflection between the detector window and filter. The right plot shows a 20 line average cut through the target and through the crosstalk (after each quadrant is aligned so the readouts are in the same direction); the logarithmic yaxis is in units of electrons. A B
10
5

Average of lines 865 to 885, iu31060sr_08071120554_flt.fits

10

4

10

3

10

2

10

1

0

-101

-102

800

1000

C

D

1200 Column (pixels)

1400

3


Instrument Science Report WFC3 2009-03

Analysis
Each full-frame, four-amp image containing a target was searched for any evidence of CT, both visually and quantitatively. The images were first fully calibrated using calwf3 and the best available reference files. The full-frame images were then split into four separate images, one for each quadrant, with the images rotated/flipped as necessary such that the read out directions all aligned with that of quadrant A. Figure 5 in Appendix B presents an image mosaic of 100x100 pixels subsections of the data taken as part of the first crosstalk proposal, with the detector at ambient. Pairs of columns show sections extracted from the full-frame images: columns 1 and 2 are from images where the target was placed in quadrant A (col 1) and crosstalk appeared in quadrant B (col 2). Similarly, columns 3 and 4 are from images where the target was placed in quadant B and crosstalk appeared in quadrant A (col 4), and so on. The HeNe laser was used to generate an unsaturated point source target (peak level ~30K e-) in the first row images and 100x that exposure level in the third row images. Images in the middle row contain a more extended source (200 fiber), with exposure level just at saturation. The CT is apparent only for the more heavily-exposed targets and is clearly worse when the target is in quadrant A or C than when the target is in B or D. Figure 4 in Appendix B contains a mosaic of image subsections from part of the second crosstalk proposal, with the detector at its nominal operating temperature. The odd columns show the target in quadrants A, B, C, and D while the even columns show the respective mirror images in the adjacent quadrant (i.e., B, A, D, and C, respectively). The rows contain data at different exposure and/or focus settings; the exposure level increases from row 1 through 4; the last two rows show defocussed images at two different exposure levels. Again, the CT appears worst when the target is placed in quadrants A or C. To quantify the effect, linear fits were made to the pixel values from the target amp as a function of their value in the victim amp. Figure 2 illustrates two of the resulting plots and fits; individual points represent each pixel while the large triangles show the averages across 1000e- wide bins. At left is the case where a target is placed in quadrant D and CT arises in C; at right, the target is in C and CT in D. Fits for point source or extended targets were performed on image subsections, typically 100x100 or 200x200 pixels in size, while the hot pixel/cosmic ray fits were done on all pixels with level higher than 20e- within the target quadrant. To minimize degradation of the fits from cosmic rays or hot pixels within the victim quadrant, the victim pixel values were limited to +/-20 e-.

4


Instrument Science Report WFC3 2009-03

Figure 2: Comparison of point source pixel values in the target quadrant versus victim quadrant along with the resultant linear fit. The large triangles are averages over bins 1000e- in width. The point source target was placed in quadrant D and C in the left and right plots, respectively, with crosstalk in C and D.

This fitting procedure was performed on all crosstalk images and the resulting linear fits were used to compute the level of crosstalk at a fiducial point of 50,000 e-. The results from all fits have been summarized in Table 1 for the various datasets outlined earlier: the crosstalk data with the detector at ambient and nominal operating temperatures as well as the glint data (binned 3x3) and the darks taken at the nominal detector temperature. Each table row lists the amp in which the target had been placed, whether all pixels or only unsaturated pixels were used in the fit, the number of images in the set, the mean CT level at 50K e-, the error in the mean CT level, as well as the max, min, and median CT levels. Visually, there did not appear to be any CT in each of the quadrants on the chip other than the target chip; fits to the target pixels vs each of the other quadrants, tabulated in Table 2, confirmed that conclusion. The results of the fitting and image evaluations can be summarized as follows. 1. The CT levels are about 2x10-4 when the target is in quadrants A or C and about 8x10-5 when the target is in B or D. 2. The quadrants on the chip other than the chip containing the target show no evidence for CT. 3. To within the errors, the CT due to hot pixels and cosmic rays is the same as that due to point or extended sources. 4. The CT levels are somewhat higher at the nominal operating temperature than when the detector is warmer though the error bars on the latter are also higher, attributed to the increased noise from operating at a higher temperature.

5


Instrument Science Report WFC3 2009-03 5. The binned data exhibited mixed CT results. While targets placed in quadrants A or B show no evidence of CT, targets placed in C contained considerably less CT (~5x10-5) than the unbinned frames (~2x10-4) while the CT due to targets placed in D is a bit worse in binned data (~1.2x10-4) than in unbinned data (~9.2x10-5). The reason for this is unclear as the same type of point source was used in all quadrants. More data acquired onorbit may help to shed more light on this issue. The resulting fits were used to correct two images from the crosstalk proposals, one extended and one point source target taken at the nominal and ambient operating temperatures, respectively. The appropriate linear fit was applied to the target quadrant and subtracted from the victim quadrant. The average fit sufficed for correcting the nominal temperature data but due to the noisier CT results from the ambient temperature images, it was necessary to use the median fit to obtain an adequate removal of the CT. The results are shown in Figure 3, where the image subsections and average slices through the corrected image show that the CT removal was quite effective: there is no trace of CT after the correction and no apparent change in noise characteristics. Table 1. Crosstalk levels at 50Ke- in the quadrant adjacent to the quadrant containing the target.
target quadrant pixels num images mean CT at 50K eCT error CT max CT min CT median

Ambient operating temperature, unbinned a a b b c c d d all nosat all nosat all nosat all nosat 3 3 3 3 3 3 3 3 -8.16 -7.93 -0.68 -0.77 -10.21 -10.31 -3.69 -3.96 1.16 1.16 1.21 1.22 1.24 1.25 1.26 1.28 -5.30 -5.30 1.82 1.82 -10.02 -10.18 -3.21 -3.36 -9.71 -9.47 -2.19 -2.46 -10.44 -10.44 -4.50 -4.50 -9.47 -9.01 -1.67 -1.67 -10.18 -10.31 -3.36 -4.00

Nominal operating temperature, unbinned a a b b c all nosat all nosat all 6 6 6 6 6 -10.24 -10.24 -3.08 -3.08 -11.04 0.27 0.27 0.25 0.25 0.26 -9.37 -9.37 -1.87 -1.87 -10.18 -10.92 -10.92 -4.59 -4.59 -11.53 -10.39 -10.39 -3.03 -3.03 -11.05

6


Instrument Science Report WFC3 2009-03

target quadrant c d d

pixels nosat all nosat

num images 6 6 6

mean CT at 50K e-11.04 -4.63 -4.63

CT error 0.26 0.26 0.26

CT max -10.18 -4.04 -4.04

CT min -11.53 -5.99 -5.99

CT median -11.05 -4.42 -4.42

Nominal operating temperature, binned a a b b c c d d all nosat all nosat all nosat all nosat 16 16 16 16 16 16 16 16 0.04 0.04 0.02 0.02 -2.11 -3.48 -5.81 -5.89 0.08 0.08 0.09 0.09 0.09 0.36 0.08 0.39 0.50 0.50 1.05 1.05 -1.42 -1.13 -5.32 -1.99 -0.52 -0.52 -0.82 -0.82 -3.28 -9.30 -6.47 -8.62 0.11 0.11 -0.09 -0.09 -2.08 -2.79 -5.86 -5.99

Nominal operating temperature, unbinned, hotpixels a a b b c c d d all nosat all nosat all nosat all nosat 6 6 6 6 6 6 6 6 -10.34 -10.65 -3.23 -3.52 -12.13 -12.84 -4.03 -3.91 0.20 0.39 0.18 0.41 0.89 0.94 1.14 1.16 -9.51 -10.09 -2.21 -1.27 -10.22 -11.67 1.08 1.08 -10.79 -11.26 -4.07 -5.36 -15.22 -15.22 -7.31 -7.31 -10.55 -10.69 -3.39 -3.38 -11.71 -12.50 -4.56 -4.65

Table 2. Mean crosstalk level at 50K e- in quadrants on the chip other than target chip.
target quadrant CT quad pixels num images mean CT at 50K eCT error CT max CT min CT median

Ambient operating temperature, unbinned a a b b c quad1 quad2 quad1 quad2 quad1 all all all all all 3 3 3 3 3 -0.87 0.08 -0.30 1.48 0.21 1.18 1.15 1.21 1.19 1.23 -0.01 0.95 0.07 4.55 0.76 -2.26 -0.39 -0.55 -0.17 -0.08 -0.35 -0.32 -0.41 0.06 -0.04

7


Instrument Science Report WFC3 2009-03

target quadrant c d d

CT quad quad2 quad1 quad2

pixels all all all

num images 3 3 3

mean CT at 50K e1.79 -0.08 -0.09

CT error 5.80 1.20 1.26

CT max -0.43 0.15 0.01

CT min -0.01 -0.39 -0.25

CT median 3.48 0.01 -0.04

Nominal operating temperature, unbinned a a b b c c d d quad1 quad2 quad1 quad2 quad1 quad2 quad1 quad2 all all all all all all all all 6 6 6 6 6 6 6 6 -0.05 0.20 -0.12 0.28 -0.34 0.07 -0.44 0.02 0.27 0.27 0.26 0.26 0.25 0.26 0.25 0.26 0.37 1.00 0.12 0.80 0.20 1.59 0.08 0.18 -0.55 -0.28 -0.50 0.13 -1.21 -1.23 -0.96 -0.08 0.05 0.13 -0.05 0.19 -0.32 0.10 -0.48 -0.01

Nominal operating temperature, binned a a b b c c d d quad1 quad2 quad1 quad2 quad1 quad2 quad1 quad2 all all all all all all all all 16 16 16 16 16 16 16 16 0.04 -0.07 0.03 0.07 0.14 0.17 0.18 -0.00 0.08 0.09 0.09 0.09 0.09 0.09 0.08 0.08 0.60 0.81 0.46 0.74 0.93 0.96 0.70 0.94 -0.42 -1.08 -0.61 -0.63 -0.56 -0.57 -0.28 -0.70 -0.01 -0.06 0.04 0.09 0.17 0.18 0.22 -0.02

8


Instrument Science Report WFC3 2009-03

Figure 3: Correction of crosstalk in a nominal-temperature, extended target image and an ambient-temperature point source target. Images are shown with a hard, inverted stretch; with the target, CT, and corrected images in columns 1,2, and 3, respectively. Plots at right show image slices of the CT before and after the correction (solid and dashed lines).
5 Average of columns 1130 to 1150 of d.amp.iu31060sr_08071120554_flt.fits

0

-5

-10

-15 5

Average of lines 840 to 911 of b.amp.iu31050kr_08059161914_flt.fits 700 800 900 1000 Line (pixels)

1100

0

-5

-10

-15 1050

1100

1150 Column (pixels)

1200

1250

Conclusions
While the extended, highly non-linear crosstalk has been eliminated from the UVIS-1' detector and there is no more crosstalk (CT) between the two chips, a low level of crosstalk within a given chip remains. The CT appears only in the amp adjacent to the amp containing the target, at ~10-4 level for targets placed in quadrants A or C and somewhat less for targets placed in quadrants B or D (~10-5). Hot pixels and cosmic rays exhibit CT at the same level as the point and extended sources; binned data appear to have less CT overall with the exception of quadrant D. The CT in the relatively simple ground test data, with a single source in one quadrant, was easily eliminated via the use of an appropriate linear fit. Application of this technique to on-orbit data, with targets present simultaneously in all four quadrants, will likely require some refinements to the procedure, such as performing the corrections iteratively or applying the correction only to target pixel values exceeding a certain threshold.

9


Instrument Science Report WFC3 2009-03

Acknowledgements
Thanks are due to the extended WFC3 team who supported the ground tests and thanks to Peter McCullough, George Hartig, John MacKenty, Randy Kimble, and Howard Bushouse for helpful discussions.

References
S.Baggett, G.Hartig, E.Cheung, "WFC3 UVIS Crosstalk Images," WFC3 Instrument Science Report 2004-11, July 2004. S. Baggett, R. Hill, G. Hartig, A. Waczynski, Y. Wen, "WFC3 Thermal Vacuum Testing: UVIS Crosstalk," WFC3 Instrument Science Report 2005-05, Feb 2005. Janesick, James R., "Scientific Charge-Coupled Devices," SPIE Press, Bellingham, WA, 2001.

10


Instrument Science Report WFC3 2009-03

Appendix A
Summary of ground test observations examined for crosstalk. Listed are the image identfication number from the ground test database, image name, exposure time, observation date and time, stimulus source (HeNe laser or 200um point source), field point and quadrant location of target. The comment field records either the approximate peak pixel level in e- or indication of whether the image was saturated or taken at a defocus setting. All images are four-amp readouts, taken with the F625W filter, at the nominal gain 1.5 setting, default bias offset level, and on side 1 (MEB1) of the instrument. The iu3105* series were taken with the detector warm, at about -50C, while the iu3106*, iu01*, and iu26* series were taken with the detector at the nominal cold temperature, about -83 C. Due to the large number of images in the latter set, only the first and last images of the series have been listed.
tvnum 49158 49159 49160 49161 49162 49163 49164 49165 49166 49167 49168 49169 49170 50817 50818 50819 50820 50821 50822 50823 50824 image iu310501r_08059142437 iu310503r_08059143705 iu310505r_08059144852 iu310507r_08059145934 iu310509r_08059151705 iu31050ar_08059151705 iu31050cr_08059153040 iu31050er_08059154811 iu31050fr_08059154811 iu31050hr_08059160146 iu31050jr_08059161914 iu31050kr_08059161914 iu31050mr_08059163336 iu310601r_08071092452 iu310602r_08071092452 iu310604r_08071094051 iu310605r_08071094051 iu310607r_08071095828 iu310608r_08071095828 iu31060ar_08071101833 iu31060br_08071101833 exptim e 10.0 12.5 10.0 10.0 12.5 10.0 10.0 12.5 10.0 10.0 12.5 10.0 0. 0 0.0 2.6 6.9 11.0 15.4 18.7 112.0 2. 6 date-obs 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-02-28 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 tim e-obs 14:22:03.04 14:34:28.06 14:46:18.05 14:57:00.06 15:08:40.06 15:14:25.06 15:28:06.05 15:39:46.05 15:45:31.04 15:59:12.06 16:10:53.06 16:16:38.06 16:30:24.06 09:19:24.04 09:22:25.06 09:35:54.06 09:38:14.06 09:51:52.04 09:55:43.05 10:09:29.06 10:16:06.06 source He Ne VISIR200 He Ne He Ne VISIR200 He Ne He Ne VISIR200 He Ne He Ne VISIR200 He Ne --VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 field point UV1 3 UV1 3 UV1 3 UV1 4 UV1 4 UV1 4 UV1 5 UV1 5 UV1 5 UV1 6 UV1 6 UV1 6 --UV1 3 UV1 3 UV1 3 UV1 3 UV1 3 UV1 3 UV1 4 quadrant D D D B B B C C C A A A --D D D D D D B comment 30K 50K saturated 27K 50K saturated 29K 51K saturated 28K 50 K saturated bias bias 9K 23K 37K 52K -15mm defocus -15mm defocus 9K

11


Instrument Science Report WFC3 2009-03

tvnum 50825 50826 50827 50828 50829 50830 50831 50832 50833 50834 50835 50836 50837 50838 50839 50840 50841 50842 50845 50847 50849 52288 52290 52292 55315 55316 thru 55436

image iu31060dr_08071103432 iu31060er_08071103432 iu31060gr_08071105209 iu31060hr_08071105209 iu31060jr_08071111214 iu31060kr_08071111214 iu31060mr_08071112813 iu31060nr_08071112813 iu31060pr_08071114550 iu31060qr_08071114550 iu31060sr_08071120554 iu31060tr_08071120554 iu31060vr_08071122153 iu31060wr_08071122153 iu31060yr_08071123930 iu31060zr_08071123930 iu310611r_08071125757 iu310612r_08071125757 iu013a03r_08071142206 iu013a06r_08071151615 iu013a09r_08071161107 iu013a03r_08077163604 iu013a06r_08077173013 iu013a09r_08077182505 iu261301r_08091215817 iu26133kr_08092043804 thru iu261302r_08091215817

exptim e 6. 9 11.0 15.4 18.7 112.0 2.6 6.9 11.0 15.4 18.7 112.0 2.6 6. 9 11.0 15.4 18.7 112.0 0. 0 3000.0 3000.0 3000.0 3000.0 3000.0 3000.0 1.0 25.0

date-obs 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-11 2008-03-17 2008-03-17 2008-03-17 2008-03-31 2008-03-31

tim e-obs 10:29:35.05 10:31:55.05 10:45:33.06 10:49:24.04 11:03:10.06 11:09:47.05 11:23:16.04 11:25:36.04 11:39:14.06 11:43:05.06 11:56:50.05 12:03:27.04 12:16:56.06 12:19:16.06 12:32:54.05 12:36:45.06 12:50:31.04 12:54:45.06 13:27:32.04 14:21:41.06 15:15:50.06 15:41:30.04 16:35:39.06 17:29:48.06 21:01:02 04:35:31 thru 21:03:56

source VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 VISIR200 -------He Ne He Ne

field point UV1 4 UV1 4 UV1 4 UV1 4 UV1 4 UV1 5 UV1 5 UV1 5 UV1 5 UV1 5 UV1 5 UV1 6 UV1 6 UV1 6 UV1 6 UV1 6 UV1 6 -------UV0 1 various

quadrant B B B B B C C C C C C A A A A A A -------straddle A/B various

comment 23K 37K 52K -15mm defocus -15mm defocus 9K 24K 38K 53K -15mm defocus -15mm defocus 9K 23K 37K 51K -15mm defocus -15mm defocus bias dark dark dark dark dark dark unsaturated, binned 3x3 saturated, binned 3x3

12


Instrument Science Report WFC3 2009-03

Appendix B
Figure 4: Mosaic of 200x200 pixel subsections from images taken with the detector at nominal operating temperature, shown with an inverted greyscale stretch. Columns 1,3,5, and 7 show the target as it appeared in amps A, B, C, and D, respectively; the adjoining columns (2,4,6, and 8) show the resulting crosstalk in amps B, A, D, and C, respectively. The first four rows show the PSF at increasingly higher exposure levels, from about 8K e(row 1) up to >50K e- (row4); the last two rows show the defocussed images taken at ~8K and ~45K e- peak levels.

13


Instrument Science Report WFC3 2009-03 Figure 5: Mosaic of 100x100 pixel subsections from images taken with the detector at ambient temperature, shown with an inverted greyscale stretch. Columns 1,3,5, and 7 show the target as it appeared in amps A, B, C, and D, respectively; the adjoining columns (2,4,6, and 8) show the resulting crosstalk in amps B, A, D, and C, respectively. Images in the first row show an unsaturated point source target, as generated by the HeNe laser (about 30K e- in the peak pixel). Images in the second row show a more extended target (200um point source, peak levels just at saturated, about 48K e-), and the third row is again a point source target at 100x the exposure level of the target in row 1.

Figure 6: Mosaic of 100x100 pixel extracts from a subset of binned images taken with the detector at nominal operating temperature, shown with an inverted greyscale stretch. Targets in odd-numbered columns are in quadrants A, B, C, and D while images in even-numbered columns are the associated CT for each (i.e., quadrants B, A, D, and C).

14