Calibration Scans next up previous contents
Next: CORRELATOR BOARDS Up: OBSERVING PROCEDURES Previous: Frequency Switching

Calibration Scans

As with any type of observing, it is always a good idea to run a few calibration scans at the beginning of, and during, your observing run to make sure that the telescope and receiver are behaving properly. At Arecibo this is done by performing a ``spider scan" (a pattern of four scans across an object, each separated in position angle by 45 tex2html_wrap_inline895 ) on a standard calibration source. The output of this scan is automatically displayed on the screen, allowing the observer to asses the current state of the telescope and chosen receiver.

Continuum sources in the list used for flux density calibration were selected in an ad-hoc fashion from a variety of places, but primarily from Kuehr et al (A&AS, 45, 367, 1981), the compendium of linearly polarized sources by Tabara & Inoue (A&AS, 39, 379, 1980), and an unpublished list of unconfused calibrator sources prepared from the NVSS survey by Jim Condon and Qi Feng Yin. For each source, all reliable flux density measurements from both the literature and web available catalogs have been fitted against frequency using the following spectral form:

displaymath977

The values for the constants are given in the source list after the word ``flux:". All sources in the catalog listed as 'good' sources are believed to be non-variable. Additionally, if the source size is not given, it is believed to be less than about 15 tex2html_wrap_inline933 . It should be noted, however, that: (a) some of the sources in this catalog are significantly linearly polarized. If you wish to calibrate on a polarized source, you should determine in advance which source(s) are appropriate polarization sources; (b) We cannot guarantee that all sources are excellent calibrators at all frequencies. For example, you should bear confusion and source-size considerations in mind when selecting a calibrator source at any frequency. It is therefore highly recommended that you observe a few different calibrators. (And if you do find a source to be a poor calibrator at a given frequency, please tell us about it, via the ``gui comment" button.)

The frequencies at which the standard spider-scan calibration is done is set by the receiver chosen. Note that when running the calibration button you cannot change the board center frequencies, bandwidth, or scan length. The scan length is set to be six times the half-power beam width, and the center frequencies and bandwidth are listed as Table 2.

  table203
Table 2: Calibration Frequencies for Standard Spider Scans  

A typical output from a standard spider-scan calibration is shown in Figure 6. This information is obtained by running Carl Heiles' Mueller Matrix routines on the data (described fully in Heiles, et.al. 2001 PASP, preprint). The output is as follows: