Procedure for Mapping
Station
Prepared by Norman Blythe
Reviewed by Mark Klaene and Jon Brinkmann
Maintained by Gretchen Van Doren
Last modified on
The proper method to map a plug plate. More detailed mapping and
inventory
software information is available.
All window shades and the interior elevator bay door needs to be
closed during mapping. Room lights can remain on and the entrance
door can be opened and closed with no problem. However, if a shade is
opened or both inside and outside bay doors are opened, mapping will
automatically stop. The problem will have to be rectified and mapping
will have to start over.
The profilometer is an instrument used to measure the depression
slope of the plate mounted on a cartridge
Cautionary Notes
Equipment Involved in
this Procedure
Procedure
CAUTIONARY
NOTES:
You must be properly trained to
accomplishing this procedure; see training
list.
A CLASS II LASER LIGHT IS PRESENT DURING
OPERATIONS; DO NOT LOOK DIRECTLY INTO THE LASER LIGHT. This is a
low hazard laser,. It is in the visible portion of the spectrum and
requires more than 0.25 s to cause a minimal retinal injury.
Therefore, we can depend on the natural "aversion response" to
prevent inadvertant injury (i.e., people will automatically blink in
less time than it takes to get hurt). This is the same hazard class
as visible wavelength laser pointers and some scanners. Be careful
not to point the beam at people's eyes (or shiny objects) and do not
stare into the direct, or specularly reflected beam. The light is
observed by closed circuit monitor.
TRIPPING HAZARDS are present. Clear the
general work area.
CRUSH HAZARDS are present. Be sure to allow
for enough room to accomplish the task.
THIS OPERATION CAN BE ACCOMPLISHED BY ONE
PERSON; TWO PERSONS ARE REQUIRED IF THERE IS A HEAVY
WORKLOAD.
SAFETY WEAR: (recommended)
rubber-soled shoes, eyewear
EQUIPMENT
INVOLVED IN THIS PROCEDURE
Data cartridge; Location:
stored on racks inside elevator bay, plugging station, or mapping
station; Appearance: approx 4.5' dia X 3' thick., fiber optic
wire interior, plate top, protected sides, top, and bottom; nine
total; 400 lbs. each.
Light defuser; Location:
stowed on top of computer printer, southwest corner;
Appearance: round, white, opaque plastic sheet smae diameter
as plate.
Manipulator arm; Location:
mounted on a rail both inside and outside, adjacent to the elevator
bay; Appearance: grey main body and double-pivot, articulated
fork (arm) with two tines; weight supported
Mapping station; Location:
mounted on the west wall; Appearance: metal table mounted on
swing arms below the window.; two laser heads attaached by expandable
cords; laser camera and mirror mounted in ceiling directly above
table.
Plate inventory computer;
Location: on top of work bench, southwest corner;
Appearance: Gateway, 17" monitor with keyboard.
Plug plate; Location: stored
in racks under work table on south wall ,or recently received freight
shipment; Appearance: flat aluminum, approx 4.5' dia. X 1/8"
thick, guide holes around outer surface, small guide tab on outer
edge, mounting hole in center.
Plugging station; Location:
free-standing; Appearance: large, grey, 2-standard mechanism,
with turn wheel to raise and lower cartridge clamps.
Profilometer; Location:
resting on the pink marble Starrett flat surface, southwest corner;
Appearance: five gauges mounted on a metal bar. Each of the
five gauges is connected by cable to a MIG box which is connected to
a computer.
Small hand tools; Location:
free-standing red tool boxes; Appearance: as described in
procedure.
PROCEDURE: accomplish in numerical
order
Setup
- Pull on the handle of the mapping station and swing it toward
the computer as far as it will go.
- Attach the laser heads to the cartridge.
- Turn on the laser using the toggle switch on the box mounted
on the west wall above the mapping station.
- Place the light defuser on top of the plate mounted in the
cartridge.
Mapping
Using the plate inventory computer, open Netscape and choose the
"Home" icon. Under "Mapper Links," choose "Plate Inventory Database";
choose "Production Database."
At the login prompt, choose your name and enter your password.
Click on "Enter" box; DO NOT press the
"Return/Enter" key. If login procedures are correct, a box
appears showing "Login Verified", click on "OK". If you have problems
contact APO's Computer Systems Manager.
If the same plate is used in the same cartridge, choose "Routine
Commands" and click on "Plan a night's observing". This brings up a
screen listing cartridges and which plates are in them. Click on that
particular cartridge number, then click "Unload". If this step is not
done the computer will recognize the previous map which is not the
correct map.
To map new and previous cartridges, go to the bottom of the
computer screen and click on the area marked "Xterm". This will bring
up a black window with a prompt marked "plate-mapper%" in the upper
left hand corner of the window. Enter in lower case "idtest". This
will return a message showing the cartridge number and the cartridge
number +32. If this is incorrect it needs to be fixed.
If correct, type in lower case letters, "map" and then press
"Enter." You are given a series of prompts:
- your last name; the computer will accept any last name; press
"Enter"
- the plate number; type in the plate number and press "Enter"
- which speed to map in; extreme is the fastest, simply type "e"
and press "Enter"
Mapping takes approximately twenty minutes. Once mapping begins
the next cartridge can be brought out and preparations started on it
or other duties can be performed.
Verification
When mapping is completed, the computer generates and prints a
sheet marked "unplugged hole fiber scan." Listed across the top are
the plate number, modified Julian date (MJD), and the version number
(i.e., 412-51933- 01.par). There is also a replication of the plug
plate including holes. Red stars indicate "dead" holes. If there are
no red stars, this is a perfect map.
If there are red stars, fibers could be dirty, dead, unplugged,
not plugged, or broken. Check each of the red stars for the
following:
- unplugged or not plugged, then replug
- end of fiber is dirty, clean with isopropyl alcohol and a
kimwipe EXL; use a gentle motion and rub the end of the fiber
- dead/broken fibers exist when cleaning and replugging are
unsuccessful
After servicing the red stars, the plate must be
re-mapped.
- Remove the laser slit heads from the cartridge. Cover the slit
heads with the white plastic head covers.
- Click on the database and make it active. Click on "Enter new
plate map". This will open a page which asks for plate number,
MJD, the Version # and the cartridge number. Use the numbers
printed across the top of the "unplugged hole fiber scan"
(example: 412-51933-01.par). Click once on the area for plate
number. Next type in the plate number (example 412). Go to "Enter
New Plate Map," click on "plate id," and enter the plate number.
Once completed, click on "Fetch latest map info," which
automatically enters the MJD, version, and cartridge number.
Compare and insure this information is correct.
- Remove the light defuser and place it back on top of the
printer.
- While the profilometer is setting on the flat stone, turn on
all five indicators, making sure they all read zero. If they
don't, press the "origin" or the "zero" button.
- Facing the mapping table, set the profilometer on the plate
with the gauges pointing to your right. The bar contains one knob
on the end that is to your right and two knobs on the end that is
to your left. Make sure the bar is mounted correctly: all three
knobs should be setting on the plate. The center set screw should
also be visible from the hole in the bar to the left of the first
indicator.
- The readings on the indicator are displayed below each
indicator; this number should be a negative. If the numbers do not
match, first loosen each cap screw in the bending ring 1/2 to 3/4
turn using the red t-handle wrenches. Re-tighten the cap screws in
the approved method. If more adjustment is needed, loosen the
thumb screw on the center pin and rotate the center pin until the
proper adjustment is achieved. Total perfection may not always be
achieved. Try to get the outer three indicators as close as
possible to the standard. This is important as they cover a larger
area of the plate than do the areas closer to the center.
- Once the numbers are as close as possible to the standard,
click on the areas marked "North", "East", "South", and "West" as
the profilometer is moved to those areas. This records the
profiles in the database.
- Once all four profiles are entered into the database and you
are satisfied with the results, scroll down and click on "Commit"
button. A message that the database has been updated or that an
error has occurred is generated. Correct the error before closing
out or click "OK". Once everything in this block is completed,
print a copy of this screen.
- Click on "Plan a night's observing" and make sure the correct
cartridge, plate, and plate map is entered.
- Place the white disk cover over the top of the cartridge and
replace the front cover.
- Swing the mapping station to your right. Using the manipulator
arm, pick up the cartridge and place it on the elevator rack ready
for use.
- Stow the manipulator arm
END OF PROCEDURE